Table 2.

The stability and precision of fluorescent microplate reader assay.

Stains Time (h) The counts of Fluorescent microplate reader (mean±SD)
Blank control Positive control INH (RFR*) RIF (RFR*) STR (RFR*)
H37Rv 0 5476.00 ± 342.39 9902.00 ± 503.95 6926.00 ± 84.44 (32.76)* 6296.00 ± 267.82 (18.53) 6540.00 ± 256.12 (24.04)
4 5336.00 ± 359.35 9626.00 ± 515.10 6792.00 ± 181.30 (33.94) 6974.00 ± 323.77 (26.53) 6766.00 ± 253.24 (33.33)
24 5218.00 ± 299.03 10088.00 ± 624.40 6414.00 ± 283.16 (24.60) 6364.00 ± 335.01 (23.53) 6476.00 ± 302.62 (25.83)
14031 0 5144.00 ± 184.07 8650.00 ± 347.20 8058.00 ± 297.19 (83.11) 8688.00 ± 413.55 (108.08) 8314.00 ± 533.32 (90.42)
4 5140.00 ± 252.59 8410.00 ± 312.17 7992.00 ± 342.15 (87.22) 8302.00 ± 452.46 (97.70) 8352.00 ± 448.74 (98.23)
24 5318.00 ± 332.60 8558.00 ± 257.04 8230.00 ± 278.57 (89.88) 8470.00 ± 527.4 (97.28) 8516.00 ± 310.85 (98.70)
14161 0 4568.00 ± 273.17 8354.00 ± 417.17 7694.00 ± 327.46 (82.57) 5380.00 ± 185.07 (21.45) 7612.00 ± 353.09 (80.40)
4 4576.00 ± 298.63 8338.00 ± 363.00 7566.00 ± 458.34 (79.48) 5768.00 ± 206.57 (31.69) 7650.00 ± 401.25 (81.71)
24 4688.00 ± 80.44 8502.00 ± 378.31 7948.00 ± 382.84 (85.47) 5320.00 ± 257.97 (16.57) 7884.00 ± 477.94 (83.80)
*

Remaining fluorescence rate (RFR) = (reaction -blank)/(positive control–blank) × 100% (“reaction” indicates the well with drug exposure; “blank” indicates the well without drug and phage; “positive control” indicates the reaction without drug).