Table 4.

Results and performance parameters by sectional outcomes interpretation: fluorescent microplate reader vs. MGIT960 system.

Drug RFR No. of isolates MGIT960 DST outcomes
R S
INH <45% 51 1 50
45–60% 23 10 13
>60% 54 50 4
RIF <45% 56 1 55
45–60% 25 9 16
>60% 47 46 1
STR <45% 64 8 56
45–60% 21 7 14
>60% 43 41 2

R, resistant; S, sensitive; RFR, remaining fluorescence rate.