Table 4. PTR-ToF-MS-8000 measurement parameters for our study.
Transfer line | Ion source | PTR drift tube | ToF-MS | |||||
---|---|---|---|---|---|---|---|---|
Inlet Flow | Temp. | Inlet Flow | Current | E/N Ratio | Temp. | Pressure | Voltage | Mass resolution |
20 sccm | 75 °C | 6–7 sccm | 4–5 mA | 137–139 Td | 75 °C | 2.3 mbar | 610 V | 4000–5000 FWHM |
Optimised operating values for transfer line, ion source, PTR drift tube and time of flight-mass spectrometry (ToF-MS) are presented.