Table 4. PTR-ToF-MS-8000 measurement parameters for our study.

Transfer line Ion source PTR drift tube ToF-MS
Inlet Flow Temp. Inlet Flow Current E/N Ratio Temp. Pressure Voltage Mass resolution
20 sccm 75 °C 6–7 sccm 4–5 mA 137–139 Td 75 °C 2.3 mbar 610 V 4000–5000 FWHM

Optimised operating values for transfer line, ion source, PTR drift tube and time of flight-mass spectrometry (ToF-MS) are presented.