. 1992 Mar-Apr;97(2):253–265. doi: 10.6028/jres.097.007

Table 4.

Error budget for an average diameter measurement of 3 μm spheres, using MEMa

Category Error source Error contribution (μm)
Systematic Random
Microsphere sensing Imperfect scan conditions: scan direction unequal-size spheres 0.0011
0.0002
E-beam exposure 0.0009
SEM resolution and E-beam wander 0.0008
Row length measurements Stage travel sampling interferometer least count 0.0009
0.0005
Microsphere grouping Sphere flattening on contact air gaps between spheres 0.0020
0.0020
Sample size (N=120) Small sample with non-zero σD 0.0055
Totals 0.0030 0.0057
a

Uncertainty in D¯:0.0030+0.0057=0.009 μm. Measured D¯ (after corrections, see Sec. 4.2.5): 2.990 μm.