Table 2. Mean interdevice differences in axial length measurements between each pair of devices.

Pairs of devices Axial length (mm)
Normal (n=84) P High myopia (n=49) P
Ultrasound-Lenstar -0.23±0.09 0.000 -0.25±0.13 0.000
Ultrasound-IOLMaster -0.24±0.09 0.000 -0.24±0.14 0.000
IOLMaster-Lenstar 0.01±0.04 0.031 -0.01±0.04 0.100

P-values from paired t-tests.