Table 3. Mean interdevice differences in IOL power calculation based on SRK II formula between each pair of devices.
Pairs of devices | IOL power (D) |
|||
Normal (n=84) | P | High myopia (n=49) | P | |
Ultrasound-Lenstar | -0.50±0.32 | 0.000 | -0.63±0.33 | 0.000 |
Ultrasound-IOLMaster | -0.51±0.33 | 0.000 | -0.60±0.34 | 0.000 |
IOLMaster-Lenstar | 0.01±0.13 | 0.474 | -0.02±0.10 | 0.103 |
P-values from paired t-tests.