Skip to main content
. 2016 Jun 18;9(6):876–880. doi: 10.18240/ijo.2016.06.15

Table 3. Mean interdevice differences in IOL power calculation based on SRK II formula between each pair of devices.

Pairs of devices IOL power (D)
Normal (n=84) P High myopia (n=49) P
Ultrasound-Lenstar -0.50±0.32 0.000 -0.63±0.33 0.000
Ultrasound-IOLMaster -0.51±0.33 0.000 -0.60±0.34 0.000
IOLMaster-Lenstar 0.01±0.13 0.474 -0.02±0.10 0.103

P-values from paired t-tests.