TABLE III.
Penetration depths (mm) for the UT-047-M17, RG-405, RG-402, and RG-401 coaxial probes for 2 and 4GHz when tested in water and 0.9% saline.
| Probe | Water | Saline | ||
|---|---|---|---|---|
| 2 GHz | 4 GHz | 2 GHz | 4 GHz | |
| UT-047-M17 | 0.157 | 0.162 | 0.158 | 0.158 |
| RG-405 | 0.286 | 0.286 | 0.283 | 0.283 |
| RG-402 | 0.486 | 0.494 | 0.483 | 0.487 |
| RG-401 | 0.874 | 0.896 | 0.866 | 0.894 |