Skip to main content
. Author manuscript; available in PMC: 2017 Mar 1.
Published in final edited form as: IEEE Trans Microw Theory Tech. 2016 Jan 29;64(3):915–923. doi: 10.1109/TMTT.2016.2519027

TABLE III.

Penetration depths (mm) for the UT-047-M17, RG-405, RG-402, and RG-401 coaxial probes for 2 and 4GHz when tested in water and 0.9% saline.

Probe Water Saline
2 GHz 4 GHz 2 GHz 4 GHz
UT-047-M17 0.157 0.162 0.158 0.158
RG-405 0.286 0.286 0.283 0.283
RG-402 0.486 0.494 0.483 0.487
RG-401 0.874 0.896 0.866 0.894