Validation results (mean sensitivity/specificity) for each system version.

Class I water
Class II water
Class III water
Sensitivity Specificity Sensitivity Specificity Sensitivity Specificity
Version 1 85% 88% NP NP NP NP
Version 2 83% 99% 80% 98% NP NP
Version 3 90% 99% 80% 99% 15% 1.0%

NP: Not performed.