Table 6.
Mean HU values and objective image noise in region of interest placed in liver for standard of care filtered back projection and reduced dose iterative reconstruction techniques
| HU values | Objective image noise | ||
| Vendor A | SD FBP | 104 ± 30 | 17 ± 3 |
| RD A-1 | 93 ± 35 | 16 ± 4 | |
| RD A-2 | 97 ± 29 | 19 ± 4 | |
| RD A-3 | 98 ± 29 (P = 0.9) | 24 ± 5 (P < 0.001) | |
| Vendor B | SD FBP | 101 ± 28 | 26 ± 5 |
| RD B-1 | 100 ± 25 | 29 ± 5 | |
| RD B-2 | 98 ± 24 | 14 ± 3 | |
| RD B-3 | 107 ± 23 (P = 0.3) | 33 ± 11 (P < 0.001) | |
| Vendor C | SD FBP | 103 ± 28 | 23 ± 7 |
| RD C-1 | 97 ± 26 | 36 ± 18 | |
| RD C-2 | 97 ± 26 | 28 ± 14 | |
| RD C-3 | 98 ± 25 (P = 0.9) | 23 ± 11 (P = 0.016) |
FBP: Filtered back projection; SD: Standard of care; RD: Reduced dose.