Table 1.
Average ratio of MIR parameter values obtained from 60 to 20 micron thick tissues imaged in transmittance on silicon wafers (N = 6).
MIR data Parameters | Parameter ratio ± SD (value at 60 μm/value at 20 μm) |
---|---|
1336 cm−1 peak area | 2.03 ± 0.200 |
1336 cm−1 second derivative peak height | 1.98 ± 0.200 |
856 cm−1 peak area | 2.25 ± 0.116 |
856 cm−1 second derivative peak height | 2.23 ± 0.131 |