TABLE 3.
Comparison of CT values obtained with traditional and ESwab collection devices for each of the four analytes in the SA Complete assay
| Target | No. of specimensa | Traditional swab |
ESwab |
t test P valueb | ||
|---|---|---|---|---|---|---|
| Avg CT | SD of CT | Avg CT | SD of CT | |||
| SPC (internal control) | 147 | 34.64 | 1.90 | 33.64 | 1.20 | 1.42 × 10−7 |
| S. aureus protein A | 60 | 25.43 | 4.91 | 25.40 | 5.10 | 0.98 |
| mecA gene | 140 | 29.93 | 4.20 | 29.73 | 4.55 | 0.71 |
| SCC cassette | 13 | 25.86 | 4.90 | 26.12 | 6.41 | 0.91 |
These values only included specimens that had an actual CT value with both the traditional and ESwab collection devices (the SPC does not need to amplify in specimens that are positive for S. aureus, mecA, or the SCC cassette.
Significant at a value of <0.05.