. 2016 Jun 24;54(7):1904–1906. doi: 10.1128/JCM.00449-16

TABLE 3.

Comparison of CT values obtained with traditional and ESwab collection devices for each of the four analytes in the SA Complete assay

Target No. of specimensa Traditional swab
ESwab
t test P valueb
Avg CT SD of CT Avg CT SD of CT
SPC (internal control) 147 34.64 1.90 33.64 1.20 1.42 × 10−7
S. aureus protein A 60 25.43 4.91 25.40 5.10 0.98
mecA gene 140 29.93 4.20 29.73 4.55 0.71
SCC cassette 13 25.86 4.90 26.12 6.41 0.91
a

These values only included specimens that had an actual CT value with both the traditional and ESwab collection devices (the SPC does not need to amplify in specimens that are positive for S. aureus, mecA, or the SCC cassette.

b

Significant at a value of <0.05.