TABLE 4.

Five representative IB/Nt-positive samples (among the 182 best positives) giving discrepant interpretation of results, depending on the commercial immunoassay used

Sample no. Titer, IU/ml (interpretation)a
Architect (Abbott Diagnostics) Cobas 6000 (Roche Diagnostics) Vidas (bioMérieux) DxI (Beckman Coulter) Centaur (Siemens Healthcare) Enzygnost (Siemens Healthcare) LXL (DiaSorin) Serion
1 5 (E) 60.4 (P) 10 (E) 9.4 (N) 10.7 (P) 6 (E) 3.5 (N) 8.11 (N)
2 7 (E) >500 (P) 10 (E) 12.2 (E) 14.1 (P) 13 (P) 4.9 (N) 10.8 (E)
3 4.5 (N) 61.8 (P) 7 (N) 11.4 (E) 4.8 (N) 7 (P) 3.2 (N) 5.5 (N)
4 8.9 (E) 6.5 (N) 20 (P) 11.2 (E) 33.3 (P) 14 (P) 16.9 (P) 21.4 (P)
5 10.7 (P) 9.5 (N) 22 (P) 14.4 (E) 58.5 (P) 14 (P) 23.2 (P) 36.9 (P)
a

P, positive; E, equivocal; N, negative.