Skip to main content
. 2016 Jun 24;54(7):1842–1850. doi: 10.1128/JCM.00613-16

TABLE 3.

Performance of TaqMan array card serotyping on isolates versus the Quellung standard

Isolate condition and Quellung serotype or sample type No. of isolates tested No. of concordant results No. of discordant results % Accuracy
Unblinded
    All serotypes from Table 2 54 54 0 100
Blindeda
    1 4 4 0 100
    2 2 2 0 100
    3 3 3 0 100
    4 4 4 0 100
    5 2 2 0 100
    6A 3 3 0 100
    6B 5 5 0 100
    7F 4 4 0 100
    8 3 3 0 100
    9N 4 4 0 100
    9V 4 4 0 100
    10A 3 3 0 100
    11A 3 3 0 100
    12F 1 1 0 100
    14 4 4 0 100
    15B 2 2 0 100
    17F 3 3 0 100
    18C 4 4 0 100
    19A 3 3 0 100
    19F 5 5 0 100
    20 3 3 0 100
    22A 2 2 0 100
    22F 4 0 4b 0
    23F 4 4 0 100
    33F 3 3 0 100
    Serotypes not included in Table 2 10 10b 0 100
    Nonpneumococcal bacteria 13 13c 0 100
Total 159 155 4 97
a

Serotypes in bold are those included in PCV13.

b

Positive with lytA but negative with any serotype-specific probe on the TAC.

c

Negative with any probe on the TAC, including lytA.