. 2016 Jun 24;54(7):1766–1773. doi: 10.1128/JCM.00299-16

TABLE 5.

Diagnostic sensitivity and specificity of the multiplex RT-PCR assay compared to virus isolation for detection of viruses in clinical serum samples

Sample No. of samples
Total Positive by:
RT-PCRa Virus isolation
DENV-1 positive 4 4 4
DENV-2 positive 5 5 5
DENV-3 positive 5 4 5
DENV-4 positive 5 5 5
DENV negative 20 0 0
CHIKV positive 40 40 40
CHIKV negative 10 0 0
a

In comparison with the results of the reference method (virus isolation), the multiplex RT-PCR assay had 95% sensitivity and 100% specificity for DENV and 100% agreement for CHIKV.