Skip to main content
. 2016 Jun 24;54(7):1877–1882. doi: 10.1128/JCM.00128-16

TABLE 5.

Correlation of TaqMan HIV-1 Qualitative Test, v2.0, to the Abbott RealTime HIV-1 Qual assay in confirmed HIV-1-negative EDTA plasma and DBS samples

Sample category Result for TaqMan HIV-1 Qual Test, v2.0 No. of samples with indicated Abbott RealTime HIV-1 Qual assay result
Positive Negative
Confirmed HIV-1-negative EDTA plasma (n = 100) Positive 0 0
Negative 0 100
Confirmed HIV-1-negative adult and baby DBS (n = 300) Positive 0 0
Negative 0 300