. 1993 Jan-Feb;98(1):71–87. doi: 10.6028/jres.098.006

Table 1.

Specifications and anticipated performance of the Fermi Chopper Spectrometer. Additional details are given in the text

Specifications
 Guide cross section (height × width): 150 × 60 mm
 Maximum beam size (height × width): 100 × 25 mm
 Monochromators (double): Pyrolytic graphite PG(002)
 1st Monochromator (height × width/type): 160 × 100 mm/flat
 2nd Monochromator (height × width/type): 184 × 100 mm/vertically focussing
 Monochromator mosaic (vertical/horizontal): 25750′
 Intermonochromator collimation: 60′
 Beam filters (thickness/type/temperature): 152 mm/Be/77 K for λ > 0.4 nm
38 mm/PG/296 K for select λ <0.4 nm
 Fermi chopper details (2 slit packages) Package 1 Package 2
  Blade curvature radius: 389.0 mm 581.4 mm
  Rotor radius: 26.9 mm 25.6 mm
  Slot width: 0.635 mm 0.635 mm
  Beam window (height × width): 100 × 25 mm 100 × 25 mm
  Optimum chopper speed: 80.93/λ(nm)Hz 54.15/λ(nm) Hz
  Sample — detector distance: 2.286 m
  Scattering angle range: 5° < ϕ < 140°
 Post-Sample radial collimation (oscillating)
  Blade separation:
  Inner/Outer radius: 203/305 mm
 Detectors (cylindrical, 3He-filled)
  Fill pressure: 0.4 MPa
  Diameter: 25 mm
  Angular separation: 1.273°
  Quantity/Active height: 12/229 mm for 5° <2θ <20°
88/457 mm for 20° < 2θ < 140°
 Incident wavelength range: 0.23 < λ0 0.61 nm (÷2 for PG(004))
 Incident energy range: 15.5 >E0 > 2.2 meV (×4 for PG(004))
 Elastic momentum transfer range: 1< Q < 50 nm−1 (×2 for PG(004))
Anticipated performance
 Fluence rate on sample: ∼1 × 108 n/m2/s at λ =0.24 nm
∼1.2 × 108 n/m2/s at λ = 0.40 nm
 Elastic energy resolution: 40 μeV at λ = 0.60 nm
150 μeV at λ = 0.40 nm
600μeV at λ=0.24 nm
Q Resolution: ΔQ/Q < 3%