Table 3. Photons and time (s) required for a 256x256 pixel image of a two-component fluorescence decay where τ1 = 0.5, τ2 = 1.2 and α1 = 0.8, with and without noise reduction techniques. Calculations assume a 80MHz laser source and 10% photon efficiency. Minimum photons required for total error < 10%.

Analysis Technique Minimum Photons Required Imaging Time for 256x256 Image (s)
None 30000 246
Temporal^ Binning 4000 32.8
Spatial* Binning 3333 27.3
Temporal and Spatial* Binning 444 3.64
Fixed τ1 & τ2 100 0.819
Fixed τ1 & τ2 with Spatial* Binning 11 0.090

^ Binning to 42 time bins.

* 9 pixel moving spatial bin.