Table 4.
Clinical sensitivities of HM-ELISA and HM-ELISA+PSAM for plasma samples in various ranges of HIV concentrations.
| Range No. | RNA viral load (copies/ml) | Method | Total No. positives | No. Positives by test | Clinical sensitivity, % | Lower - Upper confidence interval |
|---|---|---|---|---|---|---|
| 1 | ≤ 3,000 | HM-ELISA | 14 | 0 | 0 | 0 – 0 |
| HM-ELISA+PSAM | 19 | 10 | 52.63 | 30.2 – 75.1 | ||
|
| ||||||
| 2 | 3,000 – 30,000 | HM-ELISA | 46 | 15 | 32.61 | 19.1 – 46.2 |
| HM-ELISA+PSAM | 46 | 42 | 91.30 | 83.2 – 99.4 | ||
|
| ||||||
| 3 | ≥ 30,000 | HM-ELISA | 42 | 22 | 52.3 | 37.2 – 67.4 |
| HM-ELISA+PSAM | 41 | 41 | 100 | 100 – 100 | ||
|
| ||||||
| 1 & 2 | ≤ 30,000 | HM-ELISA | 60 | 15 | 25.0 | 14.0–35.96 |
| HM-ELISA+PSAM | 65 | 52 | 80.0 | 70.3–89.7 | ||
|
| ||||||
| 2 & 3 | ≥ 3,000 | HM-ELISA | 88 | 37 | 42.04 | 31.7 – 52.4 |
| HM-ELISA+PSAM | 87 | 83 | 95.40 | 91.0 – 99.8 | ||
|
| ||||||
| 1, 2&3 | All positive | HM-ELISA | 102 | 37 | 36.27 | 26.9 – 45.6 |
| HM-ELISA+PSAM | 106 | 93 | 87.74 | 81.5 – 94.0 | ||