Skip to main content
Shoulder & Elbow logoLink to Shoulder & Elbow
. 2014 May 6;6(3):191–199. doi: 10.1177/1758573214532794

Biomechanical testing of a novel osteosynthesis plate for the ulnar coronoid process

Johannes Kiene 1,, Jorn Bogun 1, Nina Brockhaus 1, Klaus Waizner 1, Arndt-Peter Schulz 1, Robert Wendlandt 1
PMCID: PMC4935064  PMID: 27582936

Abstract

Background

The present study aimed to biomechanically evaluate a novel locking plate intended for osteosynthesis of coronoid fracture compared to mini L-plates and cannulated screws.

Methods

Biomechanical tests were performed on a fracture model in synthetic bones. Three groups, each with eight implant-bone-constructs, were analyzed in quasi-static and dynamic tests. Finally, samples were tested destructively for maximum strength.

Results

The mean (SD) highest stiffness was measured for the novel plate [693 (18) N/mm], followed by the mini L-plate [646 (37) N/mm] and the cannulated screws [249 (113) N/mm]. During the cycling testing of the novel plate and the mini L-plate, no failures occurred, although three of the eight samples of cannulated screws failed during the test. The mean (SD) maximum strength during the destructive testing was 1333 (234) N for the novel plate, 1338 (227) N for the mini-L-plate and 459 (56) N for the cannulated screws. No statistical differences were found during the destructive testing between the two plates (p = 0.999), although statistical differences were found between both plates and the cannulated screws (p = 0.000 each).

Conclusions

Osteosynthesis of the coronoid process using the novel plate is mechanically similar to the mini L-plate. Both plates were superior to osteosynthesis with cannulated screws.

Keywords: Anatomic plate, coronoid process, fracture treatment, locking plate; ulna

Introduction

The ulnar coronoid process is an important stabilizer of the elbow in the sagittal plane and against varus stress.14 Biomechanical research shows instability of the elbow joint, which leads to dorsal luxation despite an intact joint capsule and ligaments related to a fracture of the coronoid process of more than 50% of the height.4,5 In the case of a missing radial head, a fracture of the coronoid process of only 25% will already lead to instability.5 Fractures of the coronoid process are rare. They occur typically in 2% to 11% of elbow luxations,6,7 which corresponds to an incidence of 5.21 per 100,000 persons/year.8 Coronoid process fractures are treated operatively in 23% to 61% of cases.9,10 However, only a few implants designed specifically for this indication are currently available.11

Because the clinical results of surgical therapy of coronoid fractures are sometimes poor,12 a novel locking plate specifically designed for coronoid process was developed in an effort to improve clinical outcome. The design goal was to utilize a large area to stabilize the coronoid process against secondary loss of reduction at the same time as ensuring that the surrounding soft tissue is saved and easy re-fixation of the capsule and ligaments is possible. The geometrical shape of the novel plate is matched to the anatomy of the coronoid process and the proximal ulna, whereas the insertion of the brachialis muscle, which has a well-described size and position,13 is preserved (Fig. 1). The shaft of the novel plate has a thickness of 1.5 mm and offers two holes for locking screws with one long hole between them. In the area of the coronoid process, the maximum thickness of the plate is 2.5 mm with three holes perpendicular to the long axis. Additionally, three small holes with a diameter of 1.4 mm in the area of the coronoid process allow for the repair of the capsule. The total length of the plate is 59 mm.

Figure 1.

Figure
1.

Osteosynthesis with the novel coronoid plate.

The present study aimed to compare the novel locking plate to the often-used nonlocking mini L-plates and cannulated screws with biomechanical tests in an anatomically fashioned synthetic bone model. Our hypothesis was that the novel plate would show equivalent mechanical properties in load-bearing capacity and endurance.

Materials and methods

Specimen preparation

The present study was performed with large left fourth-generation composite bone (Item 3426, ulna; Sawbone, Malmö, Sweden) intended for biomechanical testing. Similar to natural bone, a hard cortical shell, which is made from epoxy resin, surrounds cancellous core material made from solid rigid polyurethane foam. Because the synthetic bone is produced industrially, availability, comparability and reproducibility are superior to natural bone.14 In the first processing step, the distal third of the bone was cut with a diamond belt grinder (Exakt 310; Exakt, Norderstedt, Germany), which gave a length of 180 mm to allow for standardized mounting in the test set-up.

The novel locking coronoid plate (Coronoid TiFix-plate, 4 hole left, ref. no.2905907LT; Litos, Ahrensburg, Germany) made from commercial grade 2 pure Titanium was tested in comparison with two clinically established implants: a retrograde screw osteosynthesis with two cannulated screws (ASNIS III, 4.0 mm, Ti-6Al-4 V; Stryker Corp., Kiel, Germany) and nonlocking mini L-plates (stainless steel with 3.2-mm screws; Synthes, Freiburg, Germany). Initially, the plates were mounted on the synthetic bone with intact coronoid process. For the locking coronoid plate (Fig. 1), two 2.6-mm self-cutting locking and one nonlocking (in the sliding hole) titanium screws with lengths of 20 mm to 22 mm in the shaft were bicortically fixed and 32 mm to 36 mm in the coronoid process were unicortically fixed. The holes were drilled with a diameter of 1.9 mm.

The mini L-plates were bent to match the shape of the bone with flat-nose pliers. After drilling holes of 2.5 mm in diameter and tapping with a 3.5-mm tap, stainless steel cortical screws with a diameter of 3.5 mm and lengths of 26 mm to 32 mm were placed bicortically in the shaft and unicortically in the coronoid process. The differences in screw length were a result of minor differences in the location of tested plates. In the area of the coronoid process, partially threaded cancellous bone screws of 30–35 mm length were placed unicortically at diverging positions. The cannulated screws were placed retrograde according to common clinical practice.15 Two 1.4-mm K-wires were placed from the dorsal side into the tip of the coronoid process. Predrilling with a 1.4-mm spiral drill was required as a result of the hardness of the synthetic bone. Guided by the K-wires, holes of 2.7 mm in diameter were drilled and tapped with a thread of 4.0 mm. After countersinking, screws with lengths of 28 mm to 34 mm were unicortically placed.

After removing the implants from the synthetic bone, the coronoid process was osteotomized at a height of 17 mm (equivalent to 75% height) using a diamond belt grinder to generate reproducible instability in the joint.4,5 Afterwards, the implants were reattached to the same synthetic bone. In total, eight samples were prepared per group.

Testing procedures

The same test set-up (Fig. 2A) was used for all mechanical testing procedures. The distal part of the ulna was fixed in an imprint of acrylic resin (Technovit 4006; Heraeus Kulzer, Wehrheim, Germany) in a stainless steel pot, thus allowing for reproducible positioning at 0° extension and 5° valgus using a ball joint (Fig. 2B). The straight alignment in extension was chosen to enable a direct force transfer to the coronoid process parallel to the osteotomy line.16,17 The 5° valgus is equivalent to the mechanical axis of the arm in full extension.18 The force was applied via a shaft hinged on ball bearings. The diameter of the shaft was 18 mm to match the size of the internal diameter of the trocheal notch on the synthetic bone. The degrees of freedom perpendicular to the test axis were freed using a sliding table. Fragment movement was monitored planar and recorded with a microscopy camera (DigiMicro 2.0 Scale; DNT, Dietzenbach, Germany) during the tests. The resolution of the camera is 35 µm/pixel. Each implant/bone construct was loaded four times in the quasi-static test, whereas the first measurement was for preconditioning and was not used in analysis.

Figure 2.

Figure 2.

(a) Schematic sketch of the complete test set-up. (b) Mounting the synthetic bone in 5° valgus position.

Three mechanical tests were conducted:

  1. A nondestructive test to determine the stiffness of the implant/bone construct.

The test set-up was placed in a material testing machine (Zwick 1456; Zwick, Ulm, Germany). The test was performed with a constant deformation rate of 0.1 mm/s controlled by the software testXpert II® (Zwick/Roell, Ulm, Germany). The embedded displacement sensor and a force sensor (±2 KN, U2A; HBM, Darmstadt, Germany) were used. After a first settling cycle of loading to 30 N and unloading to 5 N, eight samples per group were tested up to 30 N, and the force–displacement curve was recorded. Using Excel 2010 (Microsoft, Redmond Corp., WA, USA), the regression line was calculated in the range between 12 N and 24 N. The slope of the regression line is the sample stiffness (N/mm).

  1. Cyclic testing to analyze the long-term endurance.

The test set-up was placed in a pneumatic fatigue tester (DHM, Clausthal-Zellerfeld, Germany). The same samples (eight per group) were used for this test. The loading profile had a frequency of 2.5 Hz with a minimal force of 20 N and a maximal force of 100 N controlled by the software SysCon easyTest (DHM). The number of cycles was set to 5000. The failure criterion was a displacement of the fracture fragment of more than 2 mm, which was monitored using the microscopy camera. The number of cycles was recorded.

  1. Destructive test to access the maximum strength of the samples.

The test set-up was placed in the material testing machine (Zwick) again. The same samples (eight plates per group and the remaining five nondislocated constructs with screws) were tested similar to test 1. The test was stopped after failure of the load displacement curve during testing. The maximum force was read from the force–displacement curve and saved in an Excel spreadsheet (Microsoft). The failure of samples was documented.

Statistical analysis

Initially, we made a sample size calculation for an unpaired t-test. The basis for this calculation was the results of a previously conducted study with a similar study design.19 We estimated a necessary sample size of eight per group (calculation parameters: significance level α = 0.05, SD = 15, effect = 21, statistical power = 0.8). The recorded data were saved in an Excel spreadsheet and analyzed in cooperation with a statistician employing SPSS, version 20 (IBM Corp., Armonk, NY, USA). The data underwent descriptive statistical analysis, comprising single factor analysis of variance with Dunnett T3 post-hoc tests. Dunnett T3 was chosen because the variances are not homogeneous. Pearson’s chi-squared test was used on categorical variables. Mean numbers of groups were rounded to integers and significances were rounded to the third decimal place.

Results

Stiffness of the implant/bone construct

The mean (SD) highest stiffness was measured for the novel plate [693 (180) N/mm] followed by the mini L-plate [646 (37) N/mm] and the cannulated screws [249 (113) N/mm] (Fig. 3). Significant differences were found between the novel plate and the mini L-plate (p = 0.024), between the novel plate and the cannulated screws (p = 0.000) and also between the mini L-plate and the cannulated screws (p = 0.000).

Figure 3.

Figure 3.

Stiffness of the implant/bone constructs (significant differences between all groups).

Cyclic testing

During the cycling testing of the novel plate and the mini L-plate, no failures occurred and also no movement of the fracture fragment was visible. Three samples of cannulated screws failed during the test: one immediately after the start of the test, one at 566 cycles and one after 1472 cycles. The remaining screws showed an average micro motion from 0.39 to 1.08 mm (Table 1). Statistical evaluation using Pearson’s chi-squared test showed a significant correlation (p = 0.032) between failure and the fixation method.

Table 1.

Results of the cyclic testing of the cannulated screws.

Sample number Average motion (mm) Failure occurred at:
1 0.39
2 0.43
3 1.05
4 1.08
5 >2 566 cycles
6 0.59
7 >2 1472 cycles
8 >2 1 cycle

Destructive testing

The maximum strength was 1333 (234) N for the novel plate, 1338 (227) N for the mini-L-plate and 459 (56) N for the cannulated screws (Fig. 4). Statistical evaluation showed no significant difference between the two plates (p = 0.999), although it did show a significant difference between the both plates and the cannulated screws (p = 0.000 each).

Figure 4.

Figure 4.

Maximum strength of the three groups in the destructive testing.

The failure of osteosynthesis was documented in all three groups:

  • For the novel plate, a loosening of the cancellous bone screws and bending of the plate was seen in six of the eight constructs, whereas the bicortical fixation in the shaft remained fixed (Fig. 5A). An impression from the articular surface of the coronoid occurred in four synthetic bones (Fig. 5B).

  • The screws of the mini L-plate remained fixed, although the coronoid process was destroyed during the test in eight of eight constructs (Fig. 6).

  • In all cases, the cannulated screws lost fixation with the fracture fragment (Fig. 7).

Figure 5.

Figure
5.

Failure of the novel coronoid plate (a) and impression of the articular surface (b) in destructive testing.

Figure 6.

Figure
6.

Failure of the mini L-plates in destructive testing.

Figure 7.

Figure
7.

Failure of the cannulated screw in destructive testing.

Discussion

Mobilization-stable osteosynthesis is essential for the treatment of fractures of the coronoid process.20 An immobilization of more than 3 weeks can lead to impairment of the range of motion.12,21,22 In the case of osteoporotic patients or comminuted fractures, implants are required to allow for stable reconstruction and retention of the coronoid process. The present study aimed to evaluate a novel locking plate with commonly used implants. All relevant mechanical parameters of the implants were tested in full extension and with an osteotomy equivalent to a fracture of type III according to Regan and Morrey,10 leading to a reproducibly unstable joint.4,5 Testing was performed in extension because the highest force transmission occurs in this motion6 and the force is transmitted parallel to the osteotomy so that the osteosynthesis construct is directly forced.

No similar biomechanical analysis of coronoid fracture fixation was found in the literature. The applicable force for the nondestructive test was initially calculated from literature on the distal humerus.23 A linear force–displacement curve was described up to 50 N using cadaver specimens. Using a force transmission ratio of 60% to the radius and 40% to the ulna,24 we calculated a force of 20 N for nondestructive testing. We increased the load in the quasi-static test to 30 N because, in pretests, the linear range of the synthetic bones was found to be higher than for the cadaver specimens. The novel plate showed a significantly higher level of stiffness than the mini-L-plate, even though the novel plate is made of the more ductile pure titanium and the mini-L-plate is made of tougher stainless steel. However, because the fracture gap after plate fixation is very small and because bicortical screws can be used in the shaft, a high stability of the bone/implant construct can be achieved with only a minor influence of the plate itself. The stiffness of the unicortically fixed, partially threaded, cannulated screws was significantly lower than that of either plate.

We used partially threaded screws to induce compression in the fracture gap.25 This operative procedure was published as a standard surgical procedure.15 The manufacturer (Stryker Corp.) recommended the insertion of the cannulated screws only to an appropriate depth; bicortical fixation was not demanded. The bicortical fixation of a screw ostheosynthesis probably has a higher mechanical load capacity compared to unicortical fixation.26 Perhaps the use of different screw types would also yield different results.2630 On the other hand, there is still the difference of a missing osteosynthesis plate and it can be assumed that other types of screw osteosynthesis also do not reach the fixation strength of a plate osteosynthesis.25 The forces during the elbow motion are difficult to quantify because of the complex interplay of muscles.31 In different biomechanical analyses of internal fixation of the distal humerus, a load of 150 N19,23 and 200 N32 was used. During normal daily activity, forces of up to 300 N act on the arm.33,34 Again, considering the force transmission ratio between the radius and ulna, we calculated a typical load of 60 N to 80 N, which was increased to 100 N to give worst-case conditions in the dynamic testing. The number of cycles was set given that, typically, 4 weeks to 5 weeks are necessary for the onset of bony bridging in the fracture gap.35,36 During the first weeks, the implant alone carries the load. Under the assumption of one arm movement per 5 minutes over 12 hours per day, the number of movements was 5040 in the relevant postoperative timespan.23 The failure criterion of 2 mm was chosen based on scientific findings of successful fracture healing.3739

During the nondestructive tests and the cyclic testing, none of the plate samples failed. No fragment dislocation ≥2 mm was visible using the microscopy camera. The stiffness of the implant/bone construct matched the nondestructive testing. Thus, it can be concluded that either plate is appropriate in the fracture treatment of the coronoid process in so far that one large fragment with good bone quality is present. Nonetheless, the applied loads under cyclic testing are able to produce failure of the osteosynthesis if the cannulated screws are used. Three of the eight samples failed during this test (Table 1). In clinical practice, routine screw osteosynthesis of coronoid fractures is commonly used with good clinical results.15,40,41

Very high force values on the humerus of up to 1900 N have been reported during the chair rising test34 and of 2500 N when carrying objects.6,33 To analyze the behaviour of the implants even under peak loads, the force during the destructive test was increased until complete mechanical failure of the samples. The novel coronoid plate and the mini-L-plate had similar strengths. The mini L-plate has a smaller contact area with the coronoid and is made of stainless steel, which is very tough, so that the bone failed in destructive testing. The larger area of the novel plate and the more ductile pure titanium can better avoid break-off of the bone fragment, although half of the constructs also showed an impression from the articular surface in destructive testing too. The loss of retention was seen in six of the eight constructs because the novel plate was bent in the curved area excluding the attachment point of the musculus brachialis. The positioning of the mini L-plate during the tests yielded optimum fixation at the ulna but would sacrifice a considerable fraction of the brachialis muscle attachment point clinically.

An advantage of the composite bones is that their variability is significantly lower than that of cadaveric specimens for all loading regimes.42,43 A limitation of the use of currently available anatomically designed synthetic bone is that the geometries and biomechanical properties are based on bones from an 890 -N (90.8 kg), 183-cm tall male.44 Because of this limitation in available sizes, it is not possible to perform testing with smaller geometries. Foam blocks mimicking osteoporotic bone are not available in the anatomic form of the human ulna. The biomechanical effects of locked osteosynthesis on osteoporotic bone have been evaluated in the past;45 we therefore aimed to determine the stability of an anatomically precountered plate on the bone for which it is designed.

Other limitations of the present study result from the testing procedure. The plates were first mounted on the bone and then removed again, and the osteotomy was performed. Subsequently, the plates were re-fixated, which could alter the holding strength, especially of the screws in the cancellous bone area. Furthermore, we performed serial testing on the constructs up to the destructive testing. The influence of these procedures is unclear and subject to debate.

A simple fracture model with only one fragment was used to simplify the testing despite comminuted fractures, which are more problematic to treat clinically. Furthermore, we examined only the longitudinal loading of the coronoid; the rotational stability was not compared.

Conclusions

Our in vitro biomechanical study suggests that plate fixation with the novel locking coronoid plate, as well as the nonlocking mini L-plate, was superior to screw fixation.

Acknowledgements

The authors would like to thank Dr Friedrich Pahlke (Seeretz, Germany) for his statistical analysis. This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.

Declaration of conflicting interests

None declared

Level of evidence

Basic science study; biomechanical study

References

  • 1.Beingessner D, Dunning C, Stacpoole R, Johnson J, King G. The effect of coronoid fractures on elbow kinematics and stability. Clin Biomech 2007; 22: 183–90. [DOI] [PubMed] [Google Scholar]
  • 2.Schneeberger AG, Sadowski MM, Jacob HA. Coronoid process and radial head as posterolateral rotatory stabilizers of the elbow. J Bone Joint Surg Am 2004; 86A: 975–82. [DOI] [PubMed] [Google Scholar]
  • 3.Deutch S, Jensen S, Tyrdal S, Olsen B, Sneppen O. Elbow joint stability following experimental osteoligamentous injury and reconstruction. J Shoulder Elbow Surg 2003; 12: 466–71. [DOI] [PubMed] [Google Scholar]
  • 4.Closkey RF, Goode JR, Kirschenbaum D, Cody RP. The role of the coronoid process in elbow stability. A biomechanical analysis of axial loading. J Bone Joint Surg Am 2000; 82: 1749–53. [DOI] [PubMed] [Google Scholar]
  • 5.Breitfuss H, Muhr G. Kronenfortsatzbrüche und Ellenbogenstabilität. Eine biomechanische und klinische Studie 1992; 226: 1–79. [Google Scholar]
  • 6.Morrey BF, Sanchez-Sotelo J. The elbow and its disorders, Philadelphia, PA: WB Saunders Co, 2009. [Google Scholar]
  • 7.van Riet RP, Morrey BF, O'Driscoll SW, Van Glabbeek F. Associated injuries complicating radial head fractures: a demographic study. Clin Orthop Rel Res 2005; 441: 351–5. [DOI] [PubMed] [Google Scholar]
  • 8.Stoneback JW, Owens BD, Sykes J, Athwal GS, Pointer L, Wolf JM. Incidence of elbow dislocations in the United States population. J Bone Joint Surg Am 2012; 94: 240–5. [DOI] [PubMed] [Google Scholar]
  • 9.Adams JE, Hoskin TL, Morrey BF, Steinmann SP. Management and outcome of 103 acute fractures of the coronoid process of the ulna. J Bone Joint Surg Br 2009; 91: 632–5. [DOI] [PubMed] [Google Scholar]
  • 10.Regan W, Morrey B. Fractures of the coronoid process of the ulna. J Bone Joint Surg Am 1989; 71: 1348–54. [PubMed] [Google Scholar]
  • 11.Lee SK, Kim HY, Kim KJ, Yang DS, Choy WS. Coronoid plate fixation of type II and III coronoid process fractures: outcome and prognostic factors. Eur J Orthop Surg Trauma 2012; 22: 213–9. [Google Scholar]
  • 12.Kiene J, Waeldchen J, Paech A, Juergens C, Schulz AP. midterm results of 58 fractures of the coronoid process of the ulna and their concomitant injuries. Open Orthop J 2013; 7: 86–93. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Neill Cage DJ, Abrams RA, Callahan JJ, Botte MJ. Soft tissue attachments of the ulnar coronoid process: an anatomic study with radiographic correlation. Clin Orthop Rel Res 1995; 320: 154–8. [PubMed] [Google Scholar]
  • 14.Grover P, Albert C, Wang M, Harris GF. Mechanical characterization of fourth generation composite humerus. Proc Instit Mech Eng H 2011; 225: 1169–76. [DOI] [PubMed] [Google Scholar]
  • 15.McKee MD, Pugh DMW, Wild LM, Schemitsch EH, King GJW. Standard surgical protocol to treat elbow dislocations with radial head and coronoid fractures. J Bone Joint Surg Am 2005; 87(Suppl 1): 22–32. [DOI] [PubMed] [Google Scholar]
  • 16.Pauwels F. Gesammelte Abhandlungen zur funktionellen Anatomie des Bewegungsapparates, Berlin: Springer-Verlag, 1965. [Google Scholar]
  • 17.Hull JR, Owen JR, Fern SE, Wayne JS, Boardman ND III. Role of the coronoid process in varus osteoarticular stability of the elbow. J Shoulder Elbow Surg 2005; 14: 441–6. [DOI] [PubMed] [Google Scholar]
  • 18.London JT. Kinematics of the elbow. J Bone Joint Surg Am 1981; 63: 529–35. [PubMed] [Google Scholar]
  • 19.Voigt C, Rank C, Waizner K, et al. Biomechanical testing of a new plate system for the distal humerus compared to two well-established implants. Int Orthop 2013; 37: 667–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Doornberg JN, Ring DC. Fracture of the anteromedial facet of the coronoid process. J Bone Joint Surg Am 2006; 88: 2216–24. [DOI] [PubMed] [Google Scholar]
  • 21.Josefsson P, Gentz C, Johnell O, Wendeberg B. Surgical versus non-surgical treatment of ligamentous injuries following dislocation of the elbow joint. A prospective randomized study. J Bone Joint Surg Am 1987; 69: 605–8. [PubMed] [Google Scholar]
  • 22.Mehlhoff T, Noble P, Bennett J, Tullos H. Simple dislocation of the elbow in the adult. Results after closed treatment. J Bone Joint Surg Am 1988; 70: 244–9. [PubMed] [Google Scholar]
  • 23.Schuster I, Korner J, Arzdorf M, Schwieger K, Diederichs G, Linke B. Mechanical comparison in cadaver specimens of three different 90-degree double-plate osteosyntheses for simulated C2-type distal humerus fractures with varying bone densities. J Orthop Trauma 2008; 22: 113–20. [DOI] [PubMed] [Google Scholar]
  • 24.Halls AA, Travill A. Transmission of pressures across the elbow joint. Anat Rec 1964; 150: 243–7. [DOI] [PubMed] [Google Scholar]
  • 25.Rüedi TP, Murphy WM, Renner M, et al. AO Prinzipien des Frakturmanagements, Stuttgart: AO Pub, 2003. [Google Scholar]
  • 26.Fowler T, Pugh K, Litsky A, Taylor B, French B. Medial malleolar fractures: a biomechanical study of fixation techniques. Orthopedics 2011; 34: e349–55. [DOI] [PubMed] [Google Scholar]
  • 27.Ricci W, Tornetta P, Borrelli J., Jr Lag screw fixation of medial malleolar fractures: a biomechanical, radiographic, and clinical comparison of unicortical partially threaded lag screws and bicortical fully threaded lag screws. J Orthop Trauma 2012; 26: 602–6. [DOI] [PubMed] [Google Scholar]
  • 28.Thompson JD, Benjamin JB, Szivek JA. Pullout strengths of cannulated and noncannulated cancellous bone screws. Clin Orthop Rel Res 1997; 341: 241–9. [PubMed] [Google Scholar]
  • 29.Asnis SE, Ernberg JJ, Bostrom MP, et al. Cancellous bone screw thread design and holding power. J Orthop Trauma 1996; 10: 462–9. [DOI] [PubMed] [Google Scholar]
  • 30.Brown GA, McCarthy T, Bourgeault CA, Callahan DJ. Mechanical performance of standard and cannulated 4.0 mm cancellous bone screws. J Orthop Res 2000; 18: 307–12. [DOI] [PubMed] [Google Scholar]
  • 31.Amis A, Dowson D, Wright V. Elbow joint force predictions for some strenuous isometric actions. J Biomech 1980; 13: 765–75. [DOI] [PubMed] [Google Scholar]
  • 32.Helfet DL, Hotchkiss RN. Internal fixation of the distal humerus: a biomechanical comparison of methods. J Orthop Trauma 1990; 4: 260–4. [DOI] [PubMed] [Google Scholar]
  • 33.Chadwick E and Nicol A. Elbow and wrist joint contact forces during occupational pick and place activities. J Biomech 2000; 33: 591–600. [DOI] [PubMed]
  • 34.Nordin M, Frankel VH. Basic biomechanics of the musculoskeletal system, Philadelphia, PA: Lippincott Williams & Wilkins, 2001. [Google Scholar]
  • 35.Claes L, Wolf S, Augat P. Mechanical modification of callus healing. Chirurg 2000; 71: 989–94. [DOI] [PubMed] [Google Scholar]
  • 36.McRae R, Esser M. Praxis der Frakturbehandlung, Amsterdam: Elsevier, 2009. [Google Scholar]
  • 37.Claes L, Augat P, Suger G, Wilke HJ. Influence of size and stability of the osteotomy gap on the success of fracture healing. J Orthop Res 1997; 15: 577–84. [DOI] [PubMed] [Google Scholar]
  • 38.Goodship A, Kenwright J. The influence of induced micromovement upon the healing of experimental tibial fractures. J Bone Joint Surg Br 1985; 67: 650–5. [DOI] [PubMed] [Google Scholar]
  • 39.Perren SM. Evolution of the internal fixation of long bone fractures: the scientific basis of biological internal fixation: choosing a new balance between stability and biology. J Bone Joint Surg Br 2002; 84: 1093–110. [DOI] [PubMed] [Google Scholar]
  • 40.Meeder PJ, Weller S, Holz U. Die operative Behandlung von Frakturen des Processus coronoideus ulnae bei komplexer Ellenbogengelenkverletzung. Op Orthop Trauma 1992; 4: 11–20. [Google Scholar]
  • 41.Ring D, Jupiter JB, Zilberfarb J. Posterior dislocation of the elbow with fractures of the radial head and coronoid. J Bone Joint Surg Am 2002; 84: 547–51. [DOI] [PubMed] [Google Scholar]
  • 42.Cristofolini L, Viceconti M. Mechanical validation of whole bone composite tibia models. J Biomech 2000; 33: 279–88. [DOI] [PubMed] [Google Scholar]
  • 43.Heiner A. Structural properties of fourth-generation composite femurs and tibias. J Biomech 2008; 41: 3282–4. [DOI] [PubMed] [Google Scholar]
  • 44.Heiner A, Brown T. Structural properties of a new design of composite replicate femurs and tibias. J Biomech 2001; 34: 773–81. [DOI] [PubMed] [Google Scholar]
  • 45.Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ. Biomechanics of locked plates and screws. J Orthop Trauma 2004; 18: 488–93. [DOI] [PubMed] [Google Scholar]

Articles from Shoulder … Elbow are provided here courtesy of SAGE Publications

RESOURCES