Table 7. Comparison of Diagnostic Accuracy of ABCS-K with That of NCSP 2002–2006, ACR BI-RADS and BCSC.
ABCS-K 2005-2010 | NCSP 2002-2006* | Desirable Goal of BI-RADS† | Benchmarks of BCSC‡ | |
---|---|---|---|---|
Recall rate (%) | 19.1 | 7.0 | < 10 | 10.6 |
CDR per 1000 examinations | 2.69 | 0.64 | 2-10 | 4.7 |
PPV (%) | 1.4 | 0.64 | 5-10 | 4.4 |
Sensitivity (%) | 86.5 | 34.8 | > 85 | 79.0 |
Specificity (%) | 81.1 | 89.1 | > 90 | 89.8 |
*Not published data, †Desirable goal of the ACR BI-RADS 4th ed. (6), ‡BCSC (12). ABCS-K = Alliance for Breast Cancer Screening in Korea, ACR BI-RADS = American College of Radiology Breast Imaging-Reporting and Data System, BCSC = Breast Cancer Screening Consortium, CDR = cancer detection rate, NCSP = National Cancer Screening Program, PPV = positive predictive value