Table 1.
Pulse Width | Single 90° pulse with GARP carbon decoupling |
Relaxation delay | 60 s |
Acquisition time | 4 s |
Number of Scans | 64 |
Spinning status | Non-spinning |
Sample temperature | 25°C |
Data points acquired | 64 K |
Dummy scans | 4 |
Preacquisition delay | 59.57 μsec |
Probe adjustments | Tuning and matching were performed on each sample |