. 2016 Jun 26;2016:4956589. doi: 10.1155/2016/4956589

Table 1.

Optimization of HPLC-UV conditions for three analytes.

Chromatographic factor Compared parameter Optimized conditions Main advantages
Column Phenomenex Hydro_RP (4.6 mm × 250 mm, 4 μm), Kinetex-C18
(4.6 mm × 100 mm, 2.6 μm)
Kinetex-C18 Improved column efficiency, speed, separation, and sensitivity

λ UV 203~210 nm for CA and AU, 235~240 nm for GPA 210 nm for CA and AU; 240 nm for GPA No difference in 203~210 nm; less miscellaneous peaks and stabler baseline at λ 240 nm

Mobile phase Methanol-H2O, ACN-H2O ACN-H2O Better peak shape and separation, stabler baseline
Adding or not adding of formic acid in H2O (0.1% formic acid) Better peak shape and separation for GPA; no obvious influence on CA and AU
Linear gradient or isocratic Isocratic Smooth baseline, better separation, less miscellaneous peaks

ACN : H2O (0.1% FA) 1 : 99, 2 : 98, 3 : 97, 5 : 95, 7 : 93, 10 : 90 5 : 95 Shorter elution time; however, CA overlapped with solvent peak when ACN was >5%; longer elution time for AU and GPA when ACN was less than 5%