The feature-to-vulnerable-background contrasts (CF) for CN, FN, and iodinated-blood nodule of the investigated CT systems with FBP-Ramp and FBP-Hanning.

CT System Feature-to-vulnerable-background contrast (CF)a
Calcified nodule (CN) Fatty-mixed nodule (FN) Iodinated-blood nodule (IB)
FBP-Ramp FPB-Hanning FBP-Ramp FBP-Hanning FBP-Ramp FBP-Hanning
CT-80kVp 1.3e+0±1.0e2 1.3e+0±0.5e2 9.8e1±9.4e3 1.0e+0±0.4e+2 2.3e+0±1.3e2 2.2e+0±0.6e2
CT-120kVp 1.2e+0±1.3e2 1.2e+0±0.5e2 1.0e+0±1.2e2 1.0e+0±0.5e2 2.0e+0±1.3e2 1.9e+0±0.5e2
CT-140kVp 1.2e+0±1.1e2 1.2e+0±0.5e2 1.0e+0±1.2e2 1.0e+0±0.4e2 1.9e+0±1.4e2 1.9e+0±0.5e2
DECT-DS j=1 1.3e+0±1.0e2 1.3e+0±0.5e2 9.8e1±1.2e2 1.0e+0±0.5e2 2.3e+0±1.3e2 2.2e+0±0.7e2
j=2 1.8e+0±4.5e2 1.8e+0±2.0e2 1.3e+0±3.5e2 1.0e+0±1.6e2 3.1e+0±6.1e2 3.1e+0±2.8e2
DECT-SW j=1 1.3e+0±1.0e2 1.3e+0±0.5e2 9.8e1±1.2e2 1.0e+0±0.5e2 2.3e+0±1.3e2 2.2e+0±0.7e2
j=2 1.5e+0±3.0e2 1.5e+0±1.4e2 1.1e+0±2.7e2 1.1e+0±1.1e2 2.6e+0±6.1e2 2.6e+0±1.9e2
MECT-PC j=1 1.6e+0±2.4e2 1.6e+0±1.1e2 9.0e1±1.8e2 1.0e+0±0.9e2 1.6e+0±1.9e2 1.7e+0±0.8e2
j=2 1.3e+0±1.8e2 1.3e+0±0.8e2 9.6e1±1.6e2 1.0e+0±0.8e2 3.1e+0±2.7e2 3.0e+0±1.3e2
j=3 1.2e+0±2.1e2 1.2e+0±0.9e2 1.0e+0±1.9e2 1.0e+0±0.7e2 1.9e+0±2.2e2 1.9e+0±1.0e2
j=4 1.1e+0±3.0e2 1.1e+0±1.1e2 1.0e+0±3.6e2 1.0e+0±1.3e2 1.6e+0±3.5e2 1.6e+0±1.6e2
j=5 1.0e+0±3.1e2 1.1e+0±1.3e2 1.0e+0±3.1e2 1.0e+0±1.3e2 1.3e+0±3.0e2 1.3e+0±1.4e2
a

Each CF was in the format as mean±standard deviation (std), where std was computed over the contrasts of 50 noise realizations.