TABLE.

Absolute Prediction Error by Intraocular Lens Calculation Formula using Manufacturer Constants

Absolute Predictor Error (Diopters)
Formula (n=42) Mean ± SD Median (*)
Haigis (n=33) 3.5 ± 2.2 3.5 (1.4, 5.1)
Hoffer Q 2.6 ± 2.0 2.1 (0.7, 4.0)
Holladay 1 1.7 ± 1.3 1.2 (0.7, 2.5)
Holladay 2 1.9 ± 1.5 1.4 (0.6, 2.9)
Olsen (n=33) 2.2 ± 1.7 1.6 (0.9, 3.3)
SRK/T 1.4 ± 1.1 1.3 (0.3, 2.1)
SRK II 2.4 ± 1.8 2.2 (0.9, 3.6)
T2 1.3 ± 1.0 1.1 (0.5, 2.0)
Universal 3.0 ± 2.2 2.8 (1.1, 4.7

SD=Standard Deviation; SRK/t=Sanders-Retzlaff-Kraff Theoretic; SRK II=Sanders-Retzlaff-Kraff II.

*

(1st Quartile, 3rd Quartile)