Skip to main content
. 2016 Apr 11;(110):53249. doi: 10.3791/53249
Theoretical glucose value Covered with silicone cap mat Test of evaporation (uncovered)
Meana Precisiona Accuracya Meana Precisiona Accuracya Evaporation
mg/L mg/L CV % Bias (%error) mg/L CV % Bias (%error) %error 
45.0 45.2 0.69 0.44 46.0 0.66 2.05 1.60
18.0 17.7 0.80 -1.68 18.0 0.72 -0.01 1.69
9.0 8.74 1.20 -2.98 8.92 0.81 -0.95 2.09
4.5 4.50 1.26 -0.04 4.58 1.57 1.76 1.80
1.8 1.85 0.74 2.90 2.01 2.82 11.6 8.48
0.9 1.03 1.43 14.1 1.16 3.52 28.3 12.4
a (n = 4)