Spin No.
|
Speed (rpm)
|
Valve No.
|
Time (sec)
|
Operation
|
1 |
3,600 |
|
60 |
blood separation |
2 |
2,400 |
1 |
3 |
open valve to transfer the plasma into the chamber containing 8 µl of detecting antibodies conjugated with HRP |
3 |
15 Hz, 15° |
|
5 |
mix plasma and detecting antibodies |
4 |
2,400 |
2 |
3 |
open valve to transfer the mixture into binding reaction chamber |
5 |
60 Hz, 2° |
|
1,200 |
mix capture antibodies on TiO2 NF mat and plasma-detecting antibodies mixture |
6 |
2,400 |
3 |
10 |
open valve to remove the mixture |
7 |
2,400 |
4 |
4 |
open valve to transfer washing buffer |
8 |
30 Hz, 30° |
|
120 |
mix washing buffer and wash TiO2 NF mat |
9 |
2,400 |
|
4 |
transfer washing buffer |
10 |
30 Hz, 30° |
|
120 |
mix washing buffer and wash TiO2 NF mat |
11 |
2,400 |
|
4 |
transfer washing buffer |
12 |
30 Hz, 30° |
|
120 |
mix washing buffer and wash TiO2 NF mat |
13 |
2,400 |
|
4 |
transfer washing buffer |
14 |
30 Hz, 30° |
|
120 |
mix washing buffer and wash TiO2 NF mat |
15 |
2,400 |
|
20 |
remove the remaining washing buffer |
16 |
|
5 |
|
close valve |
17 |
2,400 |
6 |
10 |
open valve and transfer the chemiluminescent substrate |
18 |
30 Hz, 2° |
|
60 |
mix substrate and the immunoreagents on TiO2 NF mat |
19 |
2,400 |
7 |
10 |
open valve and transfer the reacted substrate to the detection chamber |
Total time |
~ 30 min |
|