Table 2.

CBCT image QA and baseline values acquired at 65 kVp and 0.5 mA. CT numbers of solid water, bone, and lung equivalent materials, unifomity, noise, and CBCT image accuracy. For image resolution, the image was acquired at 50 kVp and 0.8 mA.

Tasks Baselines QA QA QA QA




Month1 Month2 Month3 Month4
CT number of solid water equivalent material (HU) 43.86 ± 6 40.12 ± 11 45.62 ± 5 44.56 ± 6 39.56 ± 5
CT number of bone equivalent material (HU) 280.68 ± 12 282.65 ± 8 278.69 ± 13 289.65 ± 12 290.16 ± 9
CT number of lung equivalent material (HU) −644.07 ± 9 −660.23 ± 1 −638.56 ± 18 −670.15 ± 11 −638.54 ± 6
Uniformity (HU) 19.54 ± 8.7 19.2 ± 1 22.3 ± 7.1 20.1 ± 9 18.75 ± 4.2
Noise (HU) 44.25 ± 14 50.89 ± 9 53.65 ± 6 38.62 ± 8 40.69 ± 11
CBCT image accuracy (mm pixel−1) 0.27 ± 0.01 0.27 ± 0.01 0.26 ± 0.01 0.27 ± 0.01 0.26 ± 0.01
Resolution (mm) 0.5 0.5 0.5 0.5 0.5