Table 2.
A Comparison of the Three Bleach Correction Strategies Used in this Study to Correct for the Continuous Bleaching in FRAP
Correction Strategy | Assumptions | Requirements on data | Suitable experimental systems |
---|---|---|---|
External Exponential | Complete recovery of fluorescence | FRAP recovery recorded until a high degree of convergence | In vitro molecular structures (e.g., SLBs) |
Internal Exponential | Monoexponential bleaching | FRAP recovery recorded until a very high degree of convergence needed for reliable fitting with Eq. 3 | In vitro molecular structures (e.g., SLBs) |
Reference Area | The FRAP region and the reference area have almost identical bleaching profile | Intensity trace from a reference area (optimally recorded simultaneously with the FRAP recovery) | Cells and in vitro systems |