Table 1. The measured absolute blood SO2 from three different PA modes.

Blood sample Blood gas analyzer SO2 (%) Two-wavelength single-ended PA with error % Two-wavelength differential PA
with error %a
Three-wavelength differential PA
with error %
1 93 93 (ref.) 93 (ref.) – 93 (ref.) 93 (ref.)
2 86.5 85.74 (0.87) 86.75 (0.29) – 87.22 (0.83) 85.63 (1.00)
3 82.3 79.73 (3.13) 79.55 (3.34) – 80.57 (2.10) 78.86 (4.18)
4 72.8 71.07 (2.38) 71.06 (2.39) – 72.72 (0.11) 70.14 (3.65)
5 62.2 64.86 (4.28) 60.94 (2.03) – 63.36 (1.86) 62.2 (ref.)
a

CHb values of 93% SO2 and 62.2% SO2 blood samples (1.923 mM and 2.070 mM, respectively) were used to obtain the range of the estimated absolute SO2 that consider the possible local CHb variation.