Fig. 8.

Fig. 8

Variation of measurement accuracy with the wavelength of the excitation lasers. The length of the range gate was set such that it was greater than the tube diameter so the velocity measurement is integrated over the tube cross section. The solid line represents the percentage of light (right axis) that penetrates into the centre of the tube (diameter: 390 µm), calculated from the absorption coefficient spectrum of oxyhaemoglobin (see Appendix, Fig. 11). The data points correspond to the accuracy (left axis) of velocity measurements made for a blood suspension (25% relative to whole blood) flowing at 15 mm/s. For each wavelength, the accuracy (1-fractional error) was calculated for the mean of five velocity measurements, and the vertical error bars correspond to the standard error.