Skip to main content
. 1990 Jan-Feb;95(1):49–67. doi: 10.6028/jres.095.006

Table 1.

Quoted uncertainties (3σ) of current NIST measurement services for radiation thermometry, radiometry, and photometry and estimated changes in reported values resulting from a −0.25 K reassignment of the gold-point temperature

Quantity Uncertainty Change of value
Radiance temperature [4]
800 °C   ±0.5 °C −0.2 °C
1100 0.6  −0.2
1400 0.8  −0.4
1800 1.3  −0.6
2300 2.0  −0.9
Spectral radiance [3]
250 nm ±1.6% −0.8%
350 1.2  −0.6
650 0.6  −0.3
900 0.5  −0.2
1600 0.4  −0.1
2400 0.4  −0.1
Spectral irradiance [5]
250 nm ±2.3% −0.8%
350 1.4  −0.6
650 1.0  −0.3
900 1.3  −0.2
1600 1.9  −0.1
2400 6.5  −0.1
Luminous intensity [6]
2856 K ±1.0% −0.4%
Luminous flux [6]
2856 K ±1.4% −0.4%