Table 4.
Factor | Uncertainty (%) | |
---|---|---|
Electrical lead heating | 1.0000 | 0.06 |
Radiant case heating | 1.0000 | 0.05 |
Reflection and emission loss | 1.0001 | 0.10 |
Convection-conduction loss | 1.0010 | 0.10 |
Non-uniformity of response | 0.9998 | 0.05 |
Infrared response | 1.0001 | 0.01 |
Overall correction | 1.0010 | 0.17 |