Skip to main content
. 1990 Jan-Feb;95(1):49–67. doi: 10.6028/jres.095.006

Table 9.

Estimated uncertainties (3σ) of gold-point measurement (see sections cited in parentheses for details)

Quality Uncertainty
Composite random uncertainty (6.6) 0.30%
Least-squares fit (6.7) 0.17%
Blackbody emissivity (4.2) 0.01%
Polarization (4.3) 0.02%
Far-wing bandwidth correction (4.4) 0.04%
Detector responsivity [silicon (4.5) or ECR (4.6)] 0.20%
Detector linearity (4.8) 0.10%
Diffraction (5.2.4) 0.02%
Slit-scattering function correction fσ (6.1) 0.06%
Apparatus function correction fH (6.2) 0.03%
Sphere non-uniformity correction <LS> (6.3) 0.09%
Geometric extent, GD (6.4) 0.15%
Total radiance uncertainty (added in quadrature) 0.45%
Gold-point temperature uncertainty [from eq (10), at 600 nm] 0.34 K