Table 2.

Changes in reported values and uncertainties (3σ) of NIST radiance temperature calibrations

Value Change of value Quoted uncertainty (3σ)
TnistT68 1968 1990
  800 °C  −0.16 °C  ±0.5 °C  +0.5 °C
1100 −0.26 0.6 0.6
1400 −0.39 0.8 0.7
1800 −0.60 1.3 1.2
2300 −0.93 2.0 1.8