Skip to main content
. 2015 Sep 10;17(8):909–917. doi: 10.1093/ehjci/jev215

Table 2.

Inter-study reproducibility of aortic parameters (n = 25)

Aortic parameters Exam 1 Exam 2 P-valuea Mean difference ICC SEM SDC Sample sizes
Max. Asc. area (cm2) 10.53 ± 1.91 10.61 ± 1.89 0.663 −0.07 ± 0.83 0.91** 0.25 0.70 7
Min. Asc. area (cm2) 9.47 ± 1.72 9.59 ± 1.48 0.452 −0.12 ± 0.77 0.89** 0.26 0.72 8
Max. Desc. area (cm2) 5.56 ± 1.24 5.57 ± 1.25 0.911 −0.01 ± 0.25 0.98** 0.03 0.09 4
Min. Desc. area (cm2) 5.14 ± 1.19 5.14 ± 1.16 0.891 −0.01 ± 0.26 0.98** 0.04 0.11 4
Transit time (ms) 16.52 ± 5.00 16.80 ± 5.44 0.721 −0.28 ± 3.87 0.73** 1.99 5.53 45
Transit distance (mm) 148.25 ± 27.59 146.90 ± 26.44 0.827 1.34 ± 30.43 0.38* 23.92 66.32 36
Asc. aortic strain (%) 11.53 ± 6.44 10.55 ± 6.64 0.443 0.99 ± 6.32 0.53* 4.29 11.89 260
Desc. aortic strain (%) 8.65 ± 5.30 8.35 ± 5.26 0.706 0.30 ± 3.90 0.74** 2.01 5.56 168
PWV (m/s) 9.92 ± 4.18 9.94 ± 4.55 0.968 −0.02 ± 3.03 0.77** 1.46 4.04 76
VD (mm) 27.91 ± 12.52 28.52 ± 13.21 0.845 −0.61 ± 15.42 0.30 12.91 35.78 237

Values are mean ± SD. SD, standard deviation; ICC, intraclass correlation coefficient; SEM, standard error of measurement; SDC, smallest detectable change; Max., maximum; Min., minimum; Asc., ascending; Desc., descending; PWV, pulse wave velocity; VD, vertical distance between plane and aortic arch. Sample sizes were calculated by a power of 80%, an α error of 0.05 and a SD of the mean difference.

aPaired t-test.

*P < 0.05.

**P < 0.001.