Abstract
Permanent destruction of abnormal cardiac tissue responsible for cardiac arrhythmogenesis whilst avoiding collateral tissue injury forms the cornerstone of catheter ablation therapy. As the acceptance and performance of catheter ablation increases worldwide, limitations in current technology are becoming increasingly apparent in the treatment of complex arrhythmias such as atrial fibrillation. This review will discuss the role of new technologies aimed to improve lesion formation with the ultimate goal of improving arrhythmia-free survival of patients undergoing catheter ablation of atrial arrhythmias.
Keywords: Atrial Fibrillation, Contact Force Sensing, Ablation Biophysics, Novel Technologies
Introduction
Catheter ablation forms an essential part of the management of cardiac arrhythmias including supraventricular tachycardia (SVT), atrial flutter (AFL), atrial fibrillation (AF) and ventricular tachycardia (VT).[1-4] Catheter ablation has demonstrable superiority over anti-arrhythmic drugs in improving arrhythmia burden and quality of life in patients with atrial arrhythmias and SVT.[5-12] When the arrhythmia has a well-defined electrophysiologic mechanism and the target region is localized such as in SVT or typical AFL, arrhythmia-free survival approaching 90 to 95% can be achieved.[13] However with more complex arrhythmia mechanisms such as AF, arrhythmia recurrence is observed in 50-70% of patients.[14,15] This lower efficacy is likely due to limitations in mapping, incomplete understanding of the driving mechanisms of arrhythmia (as is the case in persistent AF), and perhaps most importantly, our inability to create transmural and durable lesions in a linear fashion. Successful catheter ablation requires not only precise localization of the arrhythmogenic substrate but complete and permanent elimination of that substrate without producing collateral injury. Despite over 30 years of intensive research focusing on novel energy sources, the reliable creation of effective and permanent lesions remains challenging. This review will examine the role of existing and emerging technologies in catheter ablation designed to improve lesion quality during catheter ablation of AF. First, however, it is important to revisit some key concepts in ablation biophysics, which help to define the barriers to better lesion creation, and to explore current methods of assessing an effective lesion.
Ablation Biophysics
Radiofrequency (RF) current is the most common and widely employed source used energy source in catheter-based ablation, mainly due to a well-understood safety profile. RF ablation transfers electromagnetic energy into thermal energy by a process called resistive heating at the surface of tissue. Lethal heating destroys deeper tissue by conduction. The RF current is delivered in unipolar fashion between the catheter tip and the indifferent electrode patch placed on skin. In bipolar RF systems, the current flows between two closely apposed small electrodes.
Some important concepts of RF ablation are worthy of review. First, a rise of tissue temperature to above 50°C is required to produce irreversible thermal injury.[16-19] and loss of cellular depolarization; with lower tissue temperatures (eg 45-50°C), there may be reversible loss of cellular excitability[19] Second, lesion size is proportional to the electrode-tissue interface temperature with higher temperature at the source resulting in deeper heating.[19] Third, increasing current flow through the tissue increases resistive heating and thereby lesion size. This third fact is relevant for RF systems that are limited by calculated power (Watts). Because power = current[2] x resistance, the lower the resistance the greater the current must be to achieve the same power. One consequence is that the size of the dispersive electrode has an impact. Increasing electrode surface area leads to lower impedance through an ablation circuit, resulting in higher current delivery for a specified power, increased tissue temperature and thus deeper conductive tissue heating. The addition of a second dispersive electrode or optimizing skin preparation between the dispersive electrode and the skin can thereby improve energy delivery when the system impedance is <100Ω.[19,20] Fourth, tissue temperature decreases predictably with increasing distance from the temperature source.[21] Approximately 90% of the power delivered to the tissue is absorbed within the first 1-1.5 mm from the electrode surface, hence only a narrow rim of tissue in close contact with the catheter (2-3 mm)experiences resistive heating. Fixed level energy application results in a mono-exponential increase in electrode-tissue interface temperature within 7-10 seconds.[22] A thermal gradient of conduction created in which the tissue further from the ablation electrode may not reach its stable maximum (thermal equilibrium) until up to 2 minutes later.[22] Because of this thermal decay, small increases in distance from the ablation target may result in large increases in RF current requirements. Larger current delivery will result in deeper heating of the tissue layers passively through conduction of a hotter surface temperature.
Fifth, lesion size is also directly proportional to the radius of the heat source. Larger electrode diameter (i.e. with larger tip electrodes), length and contact area all result in a larger source radius and larger lesion size for the same current density, which is maintained by increasing power settings when a larger electrode is used.[23] It is important to note that lesion size is proportional to electrode-tissue interface temperature in conventional ablation catheters. With cooled tip catheters it is proportional to power delivered.
Although heat transfer into tissue should be a predictable biophysical phenomenon, in vivo, other factors (including the cooling effect of regional blood flow, and blood flow through the tissue) alter the heat transfer to tissue, making it less predictable. Once tissue temperature reaches or exceeds 70°C, coagulum can form at the electrode tissue interface and a sudden rise in electrical impedance can occur, that limits energy delivery. There is evidence that protein denaturation may start at temperatures around 55°C,[22] which is probably the first step in the cascade that leads to coagulation formation. Reaching temperatures of 100°C also can results in “steam pops”, that may lead to perforation and cardiac tamponade.[24] Convective cooling is a major thermodynamic factor opposing the transfer of thermal energy to deeper tissue layers by limiting surface temperature, however it is exploited because cooling the catheter tip allows greater energy delivery without boiling or coagulum formation at the electrode tip; thus, increasing the amount of current that produces direct resistive heating increases the radius of the effective heat source. For these reasons, greater depth and increased diameter of lesions can be produced with cooled-tip catheters.[24] Electrode tip cooling can be achieved passively with large tip ablation electrode or actively with closed or open-tip irrigated catheters. Electrode tip cooling, however limits the value of temperature monitoring during irrigated catheter ablation, as there is a larger disparity between electrode and tissue temperature.[19]
Factors that influence passive cooling at the electrode-tissue interface include the magnitude of regional blood flow and stability of the electrode catheter on the tissue surface.[25] The tricuspid and mitral valve annuli have high volume of regional blood flow. When endocardium is smooth, catheter stability may be difficult. Both of these factors can decrease lesion size. In contrast, ablation in trabeculated myocardium or under valve leaflets can result in highly efficient heat transfer and risk of excessive tissue heating. Other factors such as electrode-tissue contact force (CF) can directly affects lesion size, depth and volume and this will be reviewed in detail in subsequent sections.[20] Because of the variability of these factors in vivo, applied energy, power and current are poor indicators of the extent of lesion formation during RF ablation.[19,20,22] However, as tissue is heated, there is a temperature-dependent fall in tissue impedance.[26] that is a useful surrogate for lesion formation. The availability of contact force measurements has also improved the ability to create reliable lesion size.
In summary there are multiple factors that effect RF ablation lesion size including power, ablation electrode temperature, duration of energy delivery, ablation circuit impedance, catheter-tissue CF, electrode size and myocardial blood flow. It is important to note that although larger RF ablation lesions can increase the success of ablation, they can also increase the risk of collateral tissue injury, for example, to the esophagus during pulmonary vein isolation (PVI). It is thus critical to revisit our motivation for attempting to create larger, more permanent RF lesions.
The Need For Better Lesion Creation
It is now well accepted that the pulmonary veins (PVs) play a critical role in the initiation and maintenance of AF.[27] and that complete electrical disconnection of the PVs from the left atrium (LA) is an important electrophysiological endpoint.[28-30] Although electrical disconnection can be achieved almost universally by the end of the procedure, long term efficacy of catheter ablation remains soberingly modest, with single and multiple procedure success rates of 54% and 79% recently reported for paroxysmal AF over long term follow up (≥ 3 years).[14] The time-dependent inevitability of PV reconnection and its relationship to future AF recurrence has been demonstrated in many studies showing that, prior to the use of contact force for guiding ablation lesions, up to 50-64% of PVs reconnect during an intra-procedural waiting period of up to 60 minutes post acute PV isolation.[31-33] It is well appreciated that AF recurrence post catheter ablation is usually associated with resumption of PV-LA conduction.[28,29,34] Gaps in lines that allow PV to LA reconnection allow PV triggers to also serves as a triggers for other macro-reentrant atrial arrhythmias and as the substrate for some macroreentrant atrial tachycardias after circumferential antral PVI.[28,29,31,34,35] In a histopathologic study in patients undergoing MAZE for recurrent AF after initially successful catheter ablation for AF, Kowalski et al performed intraoperative assessment of PV-LA junction conduction during the procedure and took biopsies from the PV antrum at sites of endocardial scar. They found that PVs showing electrical reconnection after catheter-based RF ablation frequently exhibited either anatomic gaps or non-transmural lesions at sites of ablation. However, non-transmural lesions in some PVs had evidence of persistent PV-LA block suggesting that fact that lesion geometry may influence PV conduction.[36] Indeed, gap geometry plays a critical influence on conduction through discontinuities in RF lesions such that bifurcated (“Y” shaped) or angled gaps (containing two right angles within the conduction path) were less likely to produce bidirectional conduction block but were more likely to produce unidirectional or rate dependent conduction block compared with straight gaps.[37]
Ineffective lesion formation is the hallmark of PV-LA conduction recovery and recurrence of arrhythmia. Despite an apparently successful lesion, conduction recovery may occur. At such a site, there may be acute loss of electrical signals and electrical inexcitability as both reversible and irreversible cell injury have occurred, and in fact, acute PVI may be achieved by a combination of reversible and irreversible circumferential tissue injury.[38] A broad anterograde wavefront of edema on the leading edge of each lesion may be present that is capable of producing conduction block. This possibility may explain why PVI can often be achieved without complete PV encirclement.[39] With time, tissue at the boundary of the lesion recovers normal conduction over the course of 1-4 weeks.[40] This allows gaps within an ablation line or boundary to remain undetected (i.e. electrically silent) during the index procedure. With time as tissue recovers, conduction occurs through gaps in the ablation lines.[41]
Aside from deficiencies in mapping and imprecise understanding of arrhythmia mechanism, it is clear that a permanent and effective energy source with efficient tissue specificity to permanently eliminate arrhythmogenic substrate, yet avoiding unnecessary collateral damage is of paramount importance in the goal to improve ablation efficacy and long term arrhythmia-free survival in patients with AF. These methods must be combined however, with means for acutely assessing creation of an effective lesion.
How Effectively Do We Identify A “Better” Lesion?
As discussed previously, RF ablation produces thermal injury. Permanent myocardial necrosis requires that a tissue temperature beyond 50° C is achieved.[16-19] However, tissue temperature is not measurable in patients and is not indicated by the intraelectrode temperature measurement provided by irrigated RF ablation catheter electrodes.
There are a limited number of intra-procedural methods for assessing completeness. Impedance at baseline and subsequent changes during ablation have been used as surrogates for cathetertissue contact and for predicting lesion efficacy in a number of experimental studies.[42-47] Higher impedance with greater electrodetissue contact is thought to be due to larger interface area between the electrode and tissue. In contrast, lower impedance recorded with lower electrode-issue contact is due to lower resistivity of blood compared to myocardium and greater interface area between the electrode and blood. However, impedance varies at different sites in the heart, and the baseline impedance does not appear to be a reliable indicator of RF ablation lesion creation.[48,49] With RF application, tissue heating results in reduced myocardial resistivity and a fall in tissue impedance.[44,46,47] Animal studies have shown that the magnitude of impedance fall early after the onset of RF ablation strongly correlates with lesion depth (correlation co-efficient squared, R2 = 0.68[45]), diameter (R2=0.66[45]) and/or volume (R=0.72[42-45,47]). Furthermore, CF influences the rate and duration of impedance decrease with “poor” contact lesions resulting in earlier plateau (≤13 seconds) compared to “good” contact during which a continuous fall in impedance may reach a later plateau (eg plateau at 40 seconds).[45]
In human studies, the extent of fall in impedance has been shown to correlate with the CF.[48,50] Kumar et al. showed that there was modest linear relationship between CF during ablation and the impedance fall (r=0.32, P<0.001). Similarly, de Bortoli et al. found a stronger association between contact force and maximum impedance fall (r= 0.54, P<0.01).[50] An impedance fall of 10 ohms at 10 seconds was shown to correspond to a contact force < 20g. However there was a large degree of overlap between different ranges of CF and impedance fall was only able to roughly differentiate low, intermediate and high CF beyond 15 seconds of RF (Figure 1).[48]
Figure 1. Relationship between maximal impedance fall and average CF or FTI. Definitions of low, intermediate and high CF were ≤20 g, 21–60 g and >60g respectively and for FTI were (≤500 g*s), 501–1000 g*s and >1000 g*s, respectively. There was a moderate correlation between maximal impedance fall, average CF (A) and FTI (B) during ablation. Note, however that there is a wide variation in CF or FTI( vertical scatter of dots) at any 1 particular impedance fall value. Maximal impedance fall was significantly greater with each increment in CF category (C) or FTI category (D). Note, however that there is a large degree of overlap in the magnitude of impedance fall between different CF or FTI categories. From Kumar et al[48] permissions obtained.
These findings were confirmed by an independent study by Reichlin and colleagues.[51] In that study, the impedance decrease in the first 20 seconds was larger with increasing average category of CF over the same time. CF <5 g, 5-10 g, 10-20 g, and <20 g was associated with a median fall in impedance at 20 seconds of 5 ohm (interquartile range [IQR].[2-7] 8 ohm.[4-11] 10 ohm.[7-16] and 14 ohm.[10-19] with one commercially available CF sensing catheter (SmartTouch, Biosense-Webster, Diamond Bar, CA, USA). Similar findings have been noted with another type of CF-sensing catheter (Tacticath, St.Jude Medical, St. Paul, MN, USA) (Figure 2).[51,52] A major limitation of impedance monitoring is that it is useful only after ablation has begun. Furthermore, the relationship between impedance fall and lesion formation is imperfect. It is controversial whether impedance drop give reliable information about lesion formation. Many factors such as body surface characteristics (e.g. chest), indifferent electrode position, and hemodynamic conditions, make the correlation between impedance and CF imperfect.[53] In addition, there is no fixed value of impedance drop that has shown to reliably predict transmural lesion formation.
Figure 2. Initial impedance decrease (within the first 20 seconds) is greater with higher average CF. Shown are the median values of absolute impedance decreases in ohms (Ω) within the first 5, 10, 15, and 20 seconds during RF performed with the SmartTouch CF-sensing catheter (A) and the TactiCath CF-sensing catheter (B) according to the achieved CF for the groups o5g, 5–10 g, 10–20 g, and >20 g. Adapted from Reichlin et al,[51] permissions obtained.

The effectiveness of a lesion also may be assessed from a change in electrogram (EGM) amplitude or morphology or increase in pacing threshold at the site[54,55] EGM changes have been described as a surrogate for lesion transmurality. A reduction in EGM amplitude of <80% is thought to indicate a transmural lesion.[56] Changes in unipolar and bipolar EGM morphology post versus pre- ablation may also detect lesion transmurality.[57] Elimination of the negative deflection on the unipolar EGM identified a transmural lesion with 100% sensitivity and specificity in an animal study.[57] Indeed, continuing RF application until the creation of a completely positive unipolar signal at a particular site was associated with improved AFfree survival compared to empiric 30 seconds of RF at each site alone during PVI in one non-randomized study.[58] It is important to note that unipolar distal EGM amplitude changes are likely to provide more useful information than bipolar recordings as the signal from the ring electrode may dominate the EGM bipolar amplitude and wavefront of activation is less important.[22]
Although changes in EGM properties have been used to guide lesion effectiveness, the magnitude of change in pacing threshold with RF ablation has been shown to be substantially greater than the magnitude of decrease in bipolar or unipolar EGM. In the atrium, loss of bipolar capture post ablation predicts formation of a uniform transmural lesions.[59] and is superior to EGM-based detection of gaps within ablation lines.[60,61] In the ventricle, change in pacing threshold correlates with and is a better marker of lesion size than change in unipolar or bipolar EGM amplitude.[54] In human studies, RF ablation with the endpoint of electrical inexcitability along a circumferential line during a PVI has been shown to markedly enhance single procedure success of AF ablation compared with the aim of achieving bidirectional PV-LA conduction block alone.[61] It may also reduce acute PV reconnection rates and dormant conduction.[62]
Real-time MRI, and MRI thermography, although not readily available, are promising methods for intra-procedural assessment of lesion completeness.[41,63,64] Cardiac MR imaging appears capable of delineating areas of permanent tissue damage caused by ablation, and of identifying gaps in atrial ablation lesion lines in animal studies.[41,63]
Important limitations exist when using methods to define a “better lesion”. The gold standard of transmural tissue necrosis by histopathology is highly relevant for animal models, but is not applicable clinically. Thus the effectiveness of any new technology in human studies is primarily assessed by acute and long term clinical outcomes, which are clearly the most clinically relevant goals. Further work is thus needed to validate simple intra-procedural parameters for assessing acute lesion efficacy that reliably translate to long-term procedural success.
Technologies For Better Lesion Creation
There has been a rapid expansion of research aimed at improving the efficacy, simplicity and speed of catheter ablation. A number of “single shot” devices such as endoscopic ablation systems, multielectrode catheters, and balloons seek to provide a simpler and quicker approach to PVI. Whether these technologies will be more effective than traditional point-by-point RF ablation has not been established, and has been reviewed elsewhere.[65-67] In this review we will discuss: (i) improvement in our ability to assess real-time catheter tissue contact with force sensing; (ii) variation of existing systems that allow creation of deep intramural lesions (bipolar ablation); (iii) catheters that allow extensive and deep myocardial lesions without significant damage to collateral structures (electroporation).
Contact Force Sensing
Two new contact force (CF) sensing catheters are now available for clinical use in the USA: the TactiCath (St. Jude Medical, St. Paul, MN, USA) and SmartTouch Thermocool (Biosense Webster, Diamond Bar, CA, USA) catheters. These allow real-time measurement of the catheter tip-to-tissue CF during catheter ablation.[43,68-72]
Tacticath is a 7 French (Fr) open irrigation catheter (6 holes at the distal tip) with a 3.5-mm tip electrode, and a force sensor incorporated into the distal part of the catheter between the second and third electrode. The force sensor has a deformable body (elastic polymer) and 3 optical fibers that measure micro-deformations correlating with force applied to the catheter tip. Infrared laser light is emitted through the proximal end of the 3 optical fibers. The light is reflected by fiber Bragg gratings on the deformable body at the distal end of the optical fibers, near the tip of the catheter. Applying CF to the tip of the catheter produces a microdeformation, causing the fiber Bragg gratings to either stretch or compress, thereby changing the wavelength of the reflected light. The change of wavelength is proportional to the CF applied to the tip. The system is able to calculate and display the vector of the CF (magnitude and angle) at a sampling rate of 100 milliseconds (ms). It has a resolution and sensitivity of about 1 gram in bench testing.[43,70] During RF delivery, the force–time integral (FTI; measured in grams x seconds, g*s) is displayed (Figure 3).
Figure 3. Tacticath CF-sensing display (From Wong et al,[77] permission obtained).

The SmartTouch Thermocool catheter is a steerable, 3.5-mm sixhole open-irrigated tip ablation catheter. The catheter tip electrode is mounted on a precision spring that permits a small amount of electrode deflection. A transmitter coil that is coupled to the tip electrode, distal to the spring, emits a location reference signal. Location sensor coils placed at the proximal end of the spring detect micromovement of the dRecently, there has been renewed interest in this technology and a novel irreversible electroporation (IRE) modality that can ablate myocardial tissue using DC energy level lower than the arcing threshold has been developed. IRE is capable of creating large myocardial lesions and interestingly, there seems to be a greater safety margin for achieving large myocardial lesions without damage to surrounding structures such as the epicardial coronary arteries. The impetus for this interest is derived from the fact that RF ablation near coronary arteries can lead to acute vessel injury, and also progressive vessel stenosis from tunica intima and media thickening.[117-120] Tissue cooling by arterial flow can also lead ineffective lesion formation and unsuccessful ablation at sites adjacent to an arterial branch.[121]
Figure 4. SmartTouch platform with the CARTO mapping system (Biosense Webster).

In pre-clinical work IRE catheters (7 French).[115,122-127] have been developed in varying shapes (circular, linear) and varying number of electrodes.[8-10] Energy is applied between the electrodes and a remote indifferent electrode using a monophasic external defibrillator charged to delivery energy in a range of 50 and 360 J.[1225] In animal work, greater applied current through IRE resulted in a linear increase in median lesion depth. No signs of thermal damage to the electrode or ablation site (when examined histologically) were seen, suggesting that the putative mechanism of injury is likely nonthermal,[115] though tissue temperature was not examined specifically in those studies. Furthermore the region of IRE-induced injury did not have a sharp “cut off ” as is usually seen grossly with RF lesions , but often irregular borders were observed.[115]
IRE, when used to deliver a single pulse, non-arching 200J application between the catheter and an indifferent electrode patch at the PV ostia of swine (maximum 4 pulses/pig), resulted in acute PV EGM attenuation, increased PV stimulation threshold, and no PV stenosis with a 3.5 mm-deep lesions histologically at 3 weeks’ follow up.[127] In a follow up study in pigs, IRE to PVs resulted in acute PV stenosis that reversed at 3 months; in comparison, RF application at equivalent sites resulted in acute PV stenosis that persisted at 3 months.[126] Epicardial IRE in pigs, caused extensive myocardial lesions whilst preserving coronary artery patency and wall integrity.[122] Lesion depth and width increased with energy of electroporation applications in swine epicardium with reproducible transmurality observed and, critically, no evidence of intimal hyperplasia of the coronary arteries located within the regions of myocardial damage.[123,124] Further studies in pigs have shown epicardial IRE delivered intentionally overlying major epicardial coronaries resulted in short-lasting (<30 minutes) luminal narrowing in the targeted area with subsequent normalization, suggesting coronary spasm.[125] No injury to organs adjacent to the pericardium was noted.[123] Importantly, there was myocardial injury to tissue surrounding the arteries, as the presumed mechanism of injury was not related to tissue heating. This finding is of particular interest as myocardial sparing around an artery is not uncommon in RF or cryoablation as a result of tissue cooling (or warming) by arterial flow.[128,129] Overall, the findings from the intriguing animal work suggests that the myocardium may be more sensitive to electroporation injury than the coronary arteries, a property that could be potentially exploited in the future in epicardial ablation. However, important issues such as controlling lesion size (wide ranges of lesion depths were noted in some studies), effect on phrenic nerve injury, the prevention of acute coronary artery vasospasm, and long term effects on the coronary lumen are important issues that need to be addressed[116] The catheter has not been tested in humans and is not available for use in the USA. It is likely to have important clinical applications if safety and efficacy can be demonstrated.
Figure 5. Graphs showing that pre-ablation CF and EGM characteristics such as EGM amplitude and width (A,B) had a modest positive and modest negative correlation respectively with increasing preablation CF; median EGM amplitude increased (C) and median EGM width decreased (D) with increasing pre-ablation CF category but note the large degree of overlap in values with different CF categories. Adapted from Kumar et al,[83] permission obtained.
Other Technologies For Assessing RF Lesion Creation
A number of other technologies have been tested that will be mentioned briefly. None are available for clinical use at this stage. Direct endocardial visualization (IRISTM Ablation Catheter, Voyage Medical, Redwood City, CA, USA) are catheters that allow direct visualization of tissue characteristics (for example dense scar versus border zone in infarcted ventricles), to monitor catheter location and stability, lesion development and lesion contiguity.[130,131] In one study, this catheter created wider lesions at lower power compared to irrigated tip ablation but with comparable pacing and sensing efficacy.[131] In an animal model, visually-guided RF ablation in the ventricle resulted in more reliable lesion creation as assessed on histopathology compared with standard open irrigated catheter guided by “traditional” markers of contact where only 78% of lesions delivered were identifiable histologically.[132] This technology appears promising in improving the efficacy of ablation using currently available RF-methods.
Figure 6. Range of CFs obtained during PVI performed by experienced operators using conventional methods of contact include EGM amplitude, tactile feedback, catheter motion on fluoroscopy and the mapping system and impedance fall. Operators were blinded to real-time CF readings. Note the wide variation in CF readings within each PV segment. For example at the left superior PV segment, the CF varied from 0-80 g. Also note that the lowest CFs was at the carina and anterior PV segment on the left and the carina of the right PVs. From Kumar et al,[87] permissions obtained.

Investigators have previously demonstrated that variable catheter tip sliding can lead to poor tissue contact, compromise lesion efficacy and create unpredictable lesion dimensions.[79] Electrogram-gated pulsed RF ablation is a variation of currently available RF technology that aims to deliver the same average power as a conventional ablation but for a shorter duration based on the onset of the local electrogram. This method aims to counteracts the effects of sliding catheter movement that results in degradation of ablation efficiency. This system was shown to create consistently deeper lesions irrespective of the degree of catheter movement as well as reaching target lesion depths significantly faster than conventional ablation in one study carried out on a thermochromic gel myocardial phantom. Further work is needed to validate this technology in vivo.[25]
Figure 7. Average CF is significantly higher during short RF ablation performed with short duration apnea compared with ventilation. From Kumar et al,[89] permissions obtained.

Acoustic radiation force impulse (ARFI) imaging is an ultrasoundbased technique that allows quantification of tissue stiffness based on mechanical displacement of tissue in response to ultrasonic impulses and monitors the tissue response using conventional ultrasound methods. The technology can be utilized with intracardiac echo (ICE) and can visualize the relative elasticity difference between ablated and unablated myocardium and can accurately assess focal RFA lesion morphology in vitro.[133,134] This technology has the potential for guiding lesion transmurality in the future.
Figure 8. Real-time recordings of CF during ventilation compared with apnea at 2 anatomically adjacent sites on the cavotricuspid isthmus. With ventilation, there is an abrupt drop in contact force with each respiratory swing (marked with asterisk), which is absent during apnea. Swing in CF during cardiac motion is similar (~15 g). Adapted from Kumar et al,[89] permissions obtained.
Conclusions
A number of important developments have been made in catheter ablation aiming to improve lesion creation whilst minimizing collateral injury and improving long term arrhythmia free survival. With each new development, one must be wary of potential for harm not uncovered from small pre-clinical or clinical studies. Randomized controlled data is valuable in validating and accepting the superiority of new technology compared to traditional methods of catheter ablation, but the safety and efficacy of each technology must be tested in a systematic fashion. Rigorous surveillance of safety is important to prevent avoidable harm to patients. Progress has been made, but further work is needed on intra-procedural methods of identifying lesion completeness that reliably predict arrhythmiafree survival in long term follow up.
Disclosures
Dr. Kumar is a recipient of the Neil Hamilton Fairley Overseas Research scholarship co-funded by the National Health and Medical Research Council and the National Heart Foundation of Australia; and the Bushell Travelling Fellowship funded by the Royal Australasian College of Physicians. Dr. Tedrow receives consulting fees/honoraria from Boston Scientific Corp. and St. Jude Medical and research funding from Biosense Webster, Inc., and St. Jude Medical. Dr. John receives consulting fees/honoraria from St. Jude Medical. Dr. Epstein receives consulting fees/honoraria from Boston scientific Corp., Medtronic, Inc., and Spectranetics Corp. Dr. Michaud receives consulting fees/honoraria from Boston Scientific Corp., Medtronic, Inc., and St. Jude Medical, and research funding from Boston Scientific Corp., and Biosense Webster, Inc. Dr. William Stevenson is co-holder of a patent for needle ablation that is consigned to Brigham and Women’s Hospital. The remaining authors have no disclosures.
References
- 1.Pedersen Christian Torp, Kay G Neal, Kalman Jonathan, Borggrefe Martin, Della-Bella Paolo, Dickfeld Timm, Dorian Paul, Huikuri Heikki, Kim Youg-Hoon, Knight Bradley, Marchlinski Francis, Ross David, Sacher Frédéric, Sapp John, Shivkumar Kalyanam, Soejima Kyoko, Tada Hiroshi, Alexander Mark E, Triedman John K, Yamada Takumi, Kirchhof Paulus, Lip Gregory Y H, Kuck Karl Heinz, Mont Lluís, Haines David, Indik Jukia, Dimarco John, Exner Derek, Iesaka Yoshito, Savelieva Irene. EHRA/HRS/APHRS expert consensus on ventricular arrhythmias. Heart Rhythm. 2014 Oct;11 (10):e166–96. doi: 10.1016/j.hrthm.2014.07.024. [DOI] [PubMed] [Google Scholar]
- 2.January Craig T, Wann L Samuel, Alpert Joseph S, Calkins Hugh, Cigarroa Joaquin E, Cleveland Joseph C, Conti Jamie B, Ellinor Patrick T, Ezekowitz Michael D, Field Michael E, Murray Katherine T, Sacco Ralph L, Stevenson William G, Tchou Patrick J, Tracy Cynthia M, Yancy Clyde W. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the Heart Rhythm Society. Circulation. 2014 Dec 02;130 (23):2071–104. doi: 10.1161/CIR.0000000000000040. [DOI] [PubMed] [Google Scholar]
- 3.Cohen Mitchell I, Triedman John K, Cannon Bryan C, Davis Andrew M, Drago Fabrizio, Janousek Jan, Klein George J, Law Ian H, Morady Fred J, Paul Thomas, Perry James C, Sanatani Shubhayan, Tanel Ronn E. PACES/HRS expert consensus statement on the management of the asymptomatic young patient with a Wolff-Parkinson-White (WPW, ventricular preexcitation) electrocardiographic pattern: developed in partnership between the Pediatric and Congenital Electrophysiology Society (PACES) and the Heart Rhythm Society (HRS). Endorsed by the governing bodies of PACES, HRS, the American College of Cardiology Foundation (ACCF), the American Heart Association (AHA), the American Academy of Pediatrics (AAP), and the Canadian Heart Rhythm Society (CHRS). Heart Rhythm. 2012 Jun;9 (6):1006–24. doi: 10.1016/j.hrthm.2012.03.050. [DOI] [PubMed] [Google Scholar]
- 4.Blomström-Lundqvist Carina, Scheinman Melvin M, Aliot Etienne M, Alpert Joseph S, Calkins Hugh, Camm A John, Campbell W Barton, Haines David E, Kuck Karl H, Lerman Bruce B, Miller D Douglas, Shaeffer Charlie Willard, Stevenson William G, Tomaselli Gordon F, Antman Elliott M, Smith Sidney C, Alpert Joseph S, Faxon David P, Fuster Valentin, Gibbons Raymond J, Gregoratos Gabriel, Hiratzka Loren F, Hunt Sharon Ann, Jacobs Alice K, Russell Richard O, Priori Silvia G, Blanc Jean-Jacques, Budaj Andzrej, Burgos Enrique Fernandez, Cowie Martin, Deckers Jaap Willem, Garcia Maria Angeles Alonso, Klein Werner W, Lekakis John, Lindahl Bertil, Mazzotta Gianfranco, Morais João Carlos Araujo, Oto Ali, Smiseth Otto, Trappe Hans-Joachim. ACC/AHA/ESC guidelines for the management of patients with supraventricular arrhythmias--executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Patients With Supraventricular Arrhythmias). Circulation. 2003 Oct 14;108 (15):1871–909. doi: 10.1161/01.CIR.0000091380.04100.84. [DOI] [PubMed] [Google Scholar]
- 5.Mallidi Jaya, Nadkarni Girish N, Berger Ronald D, Calkins Hugh, Nazarian Saman. Meta-analysis of catheter ablation as an adjunct to medical therapy for treatment of ventricular tachycardia in patients with structural heart disease. Heart Rhythm. 2011 Apr;8 (4):503–10. doi: 10.1016/j.hrthm.2010.12.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Piccini Jonathan P, Lopes Renato D, Kong Melissa H, Hasselblad Vic, Jackson Kevin, Al-Khatib Sana M. Pulmonary vein isolation for the maintenance of sinus rhythm in patients with atrial fibrillation: a meta-analysis of randomized, controlled trials. Circ Arrhythm Electrophysiol. 2009 Dec;2 (6):626–33. doi: 10.1161/CIRCEP.109.856633. [DOI] [PubMed] [Google Scholar]
- 7.Calkins Hugh, Reynolds Matthew R, Spector Peter, Sondhi Manu, Xu Yingxin, Martin Amber, Williams Catherine J, Sledge Isabella. Treatment of atrial fibrillation with antiarrhythmic drugs or radiofrequency ablation: two systematic literature reviews and meta-analyses. Circ Arrhythm Electrophysiol. 2009 Aug;2 (4):349–61. doi: 10.1161/CIRCEP.108.824789. [DOI] [PubMed] [Google Scholar]
- 8.Natale A, Newby K H, Pisanó E, Leonelli F, Fanelli R, Potenza D, Beheiry S, Tomassoni G. Prospective randomized comparison of antiarrhythmic therapy versus first-line radiofrequency ablation in patients with atrial flutter. J. Am. Coll. Cardiol. 2000 Jun;35 (7):1898–904. doi: 10.1016/s0735-1097(00)00635-5. [DOI] [PubMed] [Google Scholar]
- 9.Goldberg Andrea S, Bathina Murali N, Mickelsen Steven, Nawman Rosella, West Gail, Kusumoto Fred M. Long-term outcomes on quality-of-life and health care costs in patients with supraventricular tachycardia (radiofrequency catheter ablation versus medical therapy). Am. J. Cardiol. 2002 May 01;89 (9):1120–3. doi: 10.1016/s0002-9149(02)02285-3. [DOI] [PubMed] [Google Scholar]
- 10.Bubien R S, Knotts-Dolson S M, Plumb V J, Kay G N. Effect of radiofrequency catheter ablation on health-related quality of life and activities of daily living in patients with recurrent arrhythmias. Circulation. 1996 Oct 01;94 (7):1585–91. doi: 10.1161/01.cir.94.7.1585. [DOI] [PubMed] [Google Scholar]
- 11.Kuck Karl-Heinz, Schaumann Anselm, Eckardt Lars, Willems Stephan, Ventura Rodolfo, Delacrétaz Etienne, Pitschner Heinz-Friedrich, Kautzner Josef, Schumacher Burghard, Hansen Peter S. Catheter ablation of stable ventricular tachycardia before defibrillator implantation in patients with coronary heart disease (VTACH): a multicentre randomised controlled trial. Lancet. 2010 Jan 02;375 (9708):31–40. doi: 10.1016/S0140-6736(09)61755-4. [DOI] [PubMed] [Google Scholar]
- 12.Reddy Vivek Y, Reynolds Matthew R, Neuzil Petr, Richardson Allison W, Taborsky Milos, Jongnarangsin Krit, Kralovec Stepan, Sediva Lucie, Ruskin Jeremy N, Josephson Mark E. Prophylactic catheter ablation for the prevention of defibrillator therapy. N. Engl. J. Med. 2007 Dec 27;357 (26):2657–65. doi: 10.1056/NEJMoa065457. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Spector Peter, Reynolds Matthew R, Calkins Hugh, Sondhi Manu, Xu Yingxin, Martin Amber, Williams Catherine J, Sledge Isabella. Meta-analysis of ablation of atrial flutter and supraventricular tachycardia. Am. J. Cardiol. 2009 Sep 01;104 (5):671–7. doi: 10.1016/j.amjcard.2009.04.040. [DOI] [PubMed] [Google Scholar]
- 14.Ganesan Anand N, Shipp Nicholas J, Brooks Anthony G, Kuklik Pawel, Lau Dennis H, Lim Han S, Sullivan Thomas, Roberts-Thomson Kurt C, Sanders Prashanthan. Long-term outcomes of catheter ablation of atrial fibrillation: a systematic review and meta-analysis. J Am Heart Assoc. 2013 Mar 18;2 (2) doi: 10.1161/JAHA.112.004549. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Stevenson William G, Wilber David J, Natale Andrea, Jackman Warren M, Marchlinski Francis E, Talbert Timothy, Gonzalez Mario D, Worley Seth J, Daoud Emile G, Hwang Chun, Schuger Claudio, Bump Thomas E, Jazayeri Mohammad, Tomassoni Gery F, Kopelman Harry A, Soejima Kyoko, Nakagawa Hiroshi. Irrigated radiofrequency catheter ablation guided by electroanatomic mapping for recurrent ventricular tachycardia after myocardial infarction: the multicenter thermocool ventricular tachycardia ablation trial. Circulation. 2008 Dec 16;118 (25):2773–82. doi: 10.1161/CIRCULATIONAHA.108.788604. [DOI] [PubMed] [Google Scholar]
- 16.Wood Mark, Goldberg Scott, Lau Melissa, Goel Aneesh, Alexander Daniel, Han Frederick, Feinstein Shawn. Direct measurement of the lethal isotherm for radiofrequency ablation of myocardial tissue. Circ Arrhythm Electrophysiol. 2011 Jun;4 (3):373–8. doi: 10.1161/CIRCEP.110.961169. [DOI] [PubMed] [Google Scholar]
- 17.Haines David E. Letter by Haines regarding article, "Direct measurement of the lethal isotherm for radiofrequency ablation of myocardial tissue". Circ Arrhythm Electrophysiol. 2011 Oct;4 (5) doi: 10.1161/CIRCEP.111.965459. [DOI] [PubMed] [Google Scholar]
- 18.Nath S, Lynch C, Whayne J G, Haines D E. Cellular electrophysiological effects of hyperthermia on isolated guinea pig papillary muscle. Implications for catheter ablation. Circulation. 1993 Oct;88 (4 Pt 1):1826–31. doi: 10.1161/01.cir.88.4.1826. [DOI] [PubMed] [Google Scholar]
- 19.Haines David. Biophysics of ablation: application to technology. J. Cardiovasc. Electrophysiol. 2004 Oct;15 (10 Suppl):S2–S11. doi: 10.1046/j.1540-8167.2004.15102.x. [DOI] [PubMed] [Google Scholar]
- 20.Narayan Sanjiv M, Krummen David E, Shivkumar Kalyanam, Clopton Paul, Rappel Wouter-Jan, Miller John M. Treatment of atrial fibrillation by the ablation of localized sources: CONFIRM (Conventional Ablation for Atrial Fibrillation With or Without Focal Impulse and Rotor Modulation) trial. J. Am. Coll. Cardiol. 2012 Aug 14;60 (7):628–36. doi: 10.1016/j.jacc.2012.05.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Simmers T A, de Bakker J M, Coronel R, Wittkampf F H, van Capelle F J, Janse M J, Hauer R N. Effects of intracavitary blood flow and electrode-target distance on radiofrequency power required for transient conduction block in a Langendorff-perfused canine model. J. Am. Coll. Cardiol. 1998 Jan;31 (1):231–5. doi: 10.1016/s0735-1097(97)00435-x. [DOI] [PubMed] [Google Scholar]
- 22.Wittkampf Fred H M, Nakagawa Hiroshi. RF catheter ablation: Lessons on lesions. Pacing Clin Electrophysiol. 2006 Nov;29 (11):1285–97. doi: 10.1111/j.1540-8159.2006.00533.x. [DOI] [PubMed] [Google Scholar]
- 23.Tsai C F, Tai C T, Yu W C, Chen Y J, Hsieh M H, Chiang C E, Ding Y A, Chang M S, Chen S A. Is 8-mm more effective than 4-mm tip electrode catheter for ablation of typical atrial flutter? Circulation. 1999 Aug 17;100 (7):768–71. doi: 10.1161/01.cir.100.7.768. [DOI] [PubMed] [Google Scholar]
- 24.Nakagawa H, Yamanashi W S, Pitha J V, Arruda M, Wang X, Ohtomo K, Beckman K J, McClelland J H, Lazzara R, Jackman W M. Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation. 1995 Apr 15;91 (8):2264–73. doi: 10.1161/01.cir.91.8.2264. [DOI] [PubMed] [Google Scholar]
- 25.Chik William W B, Barry Michael Anthony, Pouliopoulos Jim, Byth Karen, Midekin Christine, Bhaskaran Abhishek, Sivagangabalan Gopal, Thomas Stuart P, Ross David L, McEwan Alistair, Kovoor Pramesh, Thiagalingam Aravinda. Electrogram-gated radiofrequency ablations with duty cycle power delivery negate effects of ablation catheter motion. Circ Arrhythm Electrophysiol. 2014 Oct;7 (5):920–8. doi: 10.1161/CIRCEP.113.001112. [DOI] [PubMed] [Google Scholar]
- 26.Ko W C, Huang S K, Lin J L, Shau W Y, Lai L P, Chen P H. New method for predicting efficiency of heating by measuring bioimpedance during radiofrequency catheter ablation in humans. J. Cardiovasc. Electrophysiol. 2001 Jul;12 (7):819–23. doi: 10.1046/j.1540-8167.2001.00819.x. [DOI] [PubMed] [Google Scholar]
- 27.Haïssaguerre M, Jaïs P, Shah D C, Takahashi A, Hocini M, Quiniou G, Garrigue S, Le Mouroux A, Le Métayer P, Clémenty J. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N. Engl. J. Med. 1998 Sep 03;339 (10):659–66. doi: 10.1056/NEJM199809033391003. [DOI] [PubMed] [Google Scholar]
- 28.Verma Atul, Kilicaslan Fethi, Pisano Ennio, Marrouche Nassir F, Fanelli Raffaele, Brachmann Johannes, Geunther Jens, Potenza Domenico, Martin David O, Cummings Jennifer, Burkhardt J David, Saliba Walid, Schweikert Robert A, Natale Andrea. Response of atrial fibrillation to pulmonary vein antrum isolation is directly related to resumption and delay of pulmonary vein conduction. Circulation. 2005 Aug 02;112 (5):627–35. doi: 10.1161/CIRCULATIONAHA.104.533190. [DOI] [PubMed] [Google Scholar]
- 29.Ouyang Feifan, Antz Matthias, Ernst Sabine, Hachiya Hitoshi, Mavrakis Hercules, Deger Florian T, Schaumann Anselm, Chun Julian, Falk Peter, Hennig Detlef, Liu Xingpeng, Bänsch Dietmar, Kuck Karl-Heinz. Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique. Circulation. 2005 Jan 18;111 (2):127–35. doi: 10.1161/01.CIR.0000151289.73085.36. [DOI] [PubMed] [Google Scholar]
- 30.Calkins Hugh, Kuck Karl Heinz, Cappato Riccardo, Brugada Josep, Camm A John, Chen Shih-Ann, Crijns Harry J G, Damiano Ralph J, Davies D Wyn, DiMarco John, Edgerton James, Ellenbogen Kenneth, Ezekowitz Michael D, Haines David E, Haissaguerre Michel, Hindricks Gerhard, Iesaka Yoshito, Jackman Warren, Jalife José, Jais Pierre, Kalman Jonathan, Keane David, Kim Young-Hoon, Kirchhof Paulus, Klein George, Kottkamp Hans, Kumagai Koichiro, Lindsay Bruce D, Mansour Moussa, Marchlinski Francis E, McCarthy Patrick M, Mont J Lluis, Morady Fred, Nademanee Koonlawee, Nakagawa Hiroshi, Natale Andrea, Nattel Stanley, Packer Douglas L, Pappone Carlo, Prystowsky Eric, Raviele Antonio, Reddy Vivek, Ruskin Jeremy N, Shemin Richard J, Tsao Hsuan-Ming, Wilber David. 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design: a report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation. Developed in partnership with the European Heart Rhythm Association (EHRA), a registered branch of the European Society of Cardiology (ESC) and the European Cardiac Arrhythmia Society (ECAS); and in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), the Asia Pacific Heart Rhythm Society (APHRS), and the Society of Thoracic Surgeons (STS). Endorsed by the governing bodies of the American College of Cardiology Foundation, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, the Asia Pacific Heart Rhythm Society, and the Heart Rhythm Society. Heart Rhythm. 2012 Apr;9 (4):632–696.e21. doi: 10.1016/j.hrthm.2011.12.016. [DOI] [PubMed] [Google Scholar]
- 31.Cheema Aamir, Dong Jun, Dalal Darshan, Marine Joseph E, Henrikson Charles A, Spragg David, Cheng Alan, Nazarian Saman, Bilchick Kenneth, Sinha Sunil, Scherr Daniel, Almasry Ibrahim, Halperin Henry, Berger Ronald, Calkins Hugh. Incidence and time course of early recovery of pulmonary vein conduction after catheter ablation of atrial fibrillation. J. Cardiovasc. Electrophysiol. 2007 Apr;18 (4):387–91. doi: 10.1111/j.1540-8167.2007.00760.x. [DOI] [PubMed] [Google Scholar]
- 32.Wang Xin-hua, Liu Xu, Sun Yu-min, Gu Jia-ning, Shi Hai-feng, Zhou Li, Hu Wei. Early identification and treatment of PV re-connections: role of observation time and impact on clinical results of atrial fibrillation ablation. Europace. 2007 Jul;9 (7):481–6. doi: 10.1093/europace/eum101. [DOI] [PubMed] [Google Scholar]
- 33.Kumar Saurabh, Michaud Gregory F. Unipolar electrogram morphology to assess lesion formation during catheter ablation of atrial fibrillation: successful translation into clinical practice. Circ Arrhythm Electrophysiol. 2013 Dec;6 (6):1050–2. doi: 10.1161/CIRCEP.113.001145. [DOI] [PubMed] [Google Scholar]
- 34.Nanthakumar Kumaraswamy, Plumb Vance J, Epstein Andrew E, Veenhuyzen George D, Link Dale, Kay G Neal. Resumption of electrical conduction in previously isolated pulmonary veins: rationale for a different strategy? Circulation. 2004 Mar 16;109 (10):1226–9. doi: 10.1161/01.CIR.0000121423.78120.49. [DOI] [PubMed] [Google Scholar]
- 35.Cappato Riccardo, Negroni Silvia, Pecora Domenico, Bentivegna Stefano, Lupo Pier Paolo, Carolei Adriana, Esposito Cristina, Furlanello Francesco, De Ambroggi Luigi. Prospective assessment of late conduction recurrence across radiofrequency lesions producing electrical disconnection at the pulmonary vein ostium in patients with atrial fibrillation. Circulation. 2003 Sep 30;108 (13):1599–604. doi: 10.1161/01.CIR.0000091081.19465.F1. [DOI] [PubMed] [Google Scholar]
- 36.Kowalski Marcin, Grimes Margaret M, Perez Francisco J, Kenigsberg David N, Koneru Jayanthi, Kasirajan Vigneshwar, Wood Mark A, Ellenbogen Kenneth A. Histopathologic characterization of chronic radiofrequency ablation lesions for pulmonary vein isolation. J. Am. Coll. Cardiol. 2012 Mar 06;59 (10):930–8. doi: 10.1016/j.jacc.2011.09.076. [DOI] [PubMed] [Google Scholar]
- 37.Pérez Francisco J, Wood Mark A, Schubert Christine M. Effects of gap geometry on conduction through discontinuous radiofrequency lesions. Circulation. 2006 Apr 11;113 (14):1723–9. doi: 10.1161/CIRCULATIONAHA.105.596593. [DOI] [PubMed] [Google Scholar]
- 38.Arujuna Aruna, Karim Rashed, Caulfield Dennis, Knowles Benjamin, Rhode Kawal, Schaeffter Tobias, Kato Bernet, Rinaldi C Aldo, Cooklin Michael, Razavi Reza, O'Neill Mark D, Gill Jaswinder. Acute pulmonary vein isolation is achieved by a combination of reversible and irreversible atrial injury after catheter ablation: evidence from magnetic resonance imaging. Circ Arrhythm Electrophysiol. 2012 Aug 01;5 (4):691–700. doi: 10.1161/CIRCEP.111.966523. [DOI] [PubMed] [Google Scholar]
- 39.Miller Marc A, d'Avila Andre, Dukkipati Srinivas R, Koruth Jacob S, Viles-Gonzalez Juan, Napolitano Craig, Eggert Charles, Fischer Avi, Gomes Joseph A, Reddy Vivek Y. Acute electrical isolation is a necessary but insufficient endpoint for achieving durable PV isolation: the importance of closing the visual gap. Europace. 2012 May;14 (5):653–60. doi: 10.1093/europace/eus048. [DOI] [PubMed] [Google Scholar]
- 40.Wood Mark A, Fuller Ithiel A. Acute and chronic electrophysiologic changes surrounding radiofrequency lesions. J. Cardiovasc. Electrophysiol. 2002 Jan;13 (1):56–61. doi: 10.1046/j.1540-8167.2002.00056.x. [DOI] [PubMed] [Google Scholar]
- 41.Ranjan Ravi, Kato Ritsushi, Zviman Menekhem M, Dickfeld Timm M, Roguin Ariel, Berger Ronald D, Tomaselli Gordon F, Halperin Henry R. Gaps in the ablation line as a potential cause of recovery from electrical isolation and their visualization using MRI. Circ Arrhythm Electrophysiol. 2011 Jun;4 (3):279–86. doi: 10.1161/CIRCEP.110.960567. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Thiagalingam Aravinda, D'Avila Andre, Foley Lori, Guerrero J Luis, Lambert Hendrik, Leo Giovanni, Ruskin Jeremy N, Reddy Vivek Y. Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. J. Cardiovasc. Electrophysiol. 2010 Jul;21 (7):806–11. doi: 10.1111/j.1540-8167.2009.01693.x. [DOI] [PubMed] [Google Scholar]
- 43.Yokoyama Katsuaki, Nakagawa Hiroshi, Shah Dipen C, Lambert Hendrik, Leo Giovanni, Aeby Nicolas, Ikeda Atsushi, Pitha Jan V, Sharma Tushar, Lazzara Ralph, Jackman Warren M. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol. 2008 Dec;1 (5):354–62. doi: 10.1161/CIRCEP.108.803650. [DOI] [PubMed] [Google Scholar]
- 44.Dumas Iii John H, Himel Iv Herman D, Kiser Andy C, Quint Stephen R, Knisley Stephen B. Myocardial electrical impedance as a predictor of the quality of RF-induced linear lesions. Physiol Meas. 2008 Oct;29 (10):1195–207. doi: 10.1088/0967-3334/29/10/004. [DOI] [PubMed] [Google Scholar]
- 45.Avitall B, Mughal K, Hare J, Helms R, Krum D. The effects of electrode-tissue contact on radiofrequency lesion generation. Pacing Clin Electrophysiol. 1997 Dec;20 (12 Pt 1):2899–910. doi: 10.1111/j.1540-8159.1997.tb05458.x. [DOI] [PubMed] [Google Scholar]
- 46.Zheng X, Walcott G P, Hall J A, Rollins D L, Smith W M, Kay G N, Ideker R E. Electrode impedance: an indicator of electrode-tissue contact and lesion dimensions during linear ablation. J Interv Card Electrophysiol. 2000 Dec;4 (4):645–54. doi: 10.1023/a:1026586119600. [DOI] [PubMed] [Google Scholar]
- 47.Stagegaard Niels, Petersen Helen Høgh, Chen Xu, Svendsen Jesper Hastrup. Indication of the radiofrequency induced lesion size by pre-ablation measurements. Europace. 2005 Nov;7 (6):525–34. doi: 10.1016/j.eupc.2005.05.016. [DOI] [PubMed] [Google Scholar]
- 48.Kumar Saurabh, Haqqani H M, Chan Martin, Lee Justin, Yudi Matias, Wong Michael C G, Morton Joseph B, Ling Liang-Han, Robinson Timothy, Heck Patrick M, Kelland Nicholas F, Halloran Karen, Spence Steven J, Kistler Peter M, Kalman Jonathan M. Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans. Heart Rhythm. 2013 Jul;10 (7):962–9. doi: 10.1016/j.hrthm.2013.03.022. [DOI] [PubMed] [Google Scholar]
- 49.Nakagawa Hiroshi, Kautzner Josef, Natale Andrea, Peichl Petr, Cihak Robert, Wichterle Dan, Ikeda Atsushi, Santangeli Pasquale, Di Biase Luigi, Jackman Warren M. Locations of high contact force during left atrial mapping in atrial fibrillation patients: electrogram amplitude and impedance are poor predictors of electrode-tissue contact force for ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2013 Aug;6 (4):746–53. doi: 10.1161/CIRCEP.113.978320. [DOI] [PubMed] [Google Scholar]
- 50.De Bortoli Alessandro, Sun Li-Zhi, Solheim Eivind, Hoff Per Ivar, Schuster Peter, Ohm Ole-Jørgen, Chen Jian. Ablation effect indicated by impedance fall is correlated with contact force level during ablation for atrial fibrillation. J. Cardiovasc. Electrophysiol. 2013 Nov;24 (11):1210–5. doi: 10.1111/jce.12215. [DOI] [PubMed] [Google Scholar]
- 51.Reichlin Tobias, Knecht Sven, Lane Christopher, Kühne Michael, Nof Eyal, Chopra Nagesh, Tadros Thomas M, Reddy Vivek Y, Schaer Beat, John Roy M, Osswald Stefan, Stevenson William G, Sticherling Christian, Michaud Gregory F. Initial impedance decrease as an indicator of good catheter contact: insights from radiofrequency ablation with force sensing catheters. Heart Rhythm. 2014 Feb;11 (2):194–201. doi: 10.1016/j.hrthm.2013.10.048. [DOI] [PubMed] [Google Scholar]
- 52.Kumar Saurabh, Haqqani H M, Chan Martin, Lee Justin, Yudi Matias, Wong Michael C G, Morton Joseph B, Ling Liang-Han, Robinson Timothy, Heck Patrick M, Kelland Nicholas F, Halloran Karen, Spence Steven J, Kistler Peter M, Kalman Jonathan M. Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans. Heart Rhythm. 2013 Jul;10 (7):962–9. doi: 10.1016/j.hrthm.2013.03.022. [DOI] [PubMed] [Google Scholar]
- 53.Makimoto Hisaki, Lin Tina, Rillig Andreas, Metzner Andreas, Wohlmuth Peter, Arya Anita, Antz Matthias, Mathew Shibu, Deiss Sebastian, Wissner Erik, Rausch Peter, Bardyszewski Aleksander, Kamioka Masashi, Li Xuping, Kuck Karl-Heinz, Ouyang Feifan, Tilz Roland Richard. In vivo contact force analysis and correlation with tissue impedance during left atrial mapping and catheter ablation of atrial fibrillation. Circ Arrhythm Electrophysiol. 2014 Feb;7 (1):46–54. doi: 10.1161/CIRCEP.113.000556. [DOI] [PubMed] [Google Scholar]
- 54.Sapp John L, Soejima Kyoko, Cooper Joshua M, Epstein Laurence M, Stevenson William G. Ablation lesion size correlates with pacing threshold: a physiological basis for use of pacing to assess ablation lesions. Pacing Clin Electrophysiol. 2004 Jul;27 (7):933–7. doi: 10.1111/j.1540-8159.2004.00561.x. [DOI] [PubMed] [Google Scholar]
- 55.Delacretaz Etienne, Soejima Kyoko, Brunckhorst Corinna B, Maisel William H, Friedman Peter L, Stevenson William G. Assessment of radiofrequency ablation effect from unipolar pacing threshold. Pacing Clin Electrophysiol. 2003 Oct;26 (10):1993–6. doi: 10.1046/j.1460-9592.2003.00307.x. [DOI] [PubMed] [Google Scholar]
- 56.Gepstein L, Hayam G, Shpun S, Cohen D, Ben-Haim S A. Atrial linear ablations in pigs. Chronic effects on atrial electrophysiology and pathology. Circulation. 1999 Jul 27;100 (4):419–26. doi: 10.1161/01.cir.100.4.419. [DOI] [PubMed] [Google Scholar]
- 57.Otomo Kiyoshi, Uno Kikuya, Fujiwara Hideomi, Isobe Mitsuaki, Iesaka Yoshito. Local unipolar and bipolar electrogram criteria for evaluating the transmurality of atrial ablation lesions at different catheter orientations relative to the endocardial surface. Heart Rhythm. 2010 Sep;7 (9):1291–300. doi: 10.1016/j.hrthm.2010.06.014. [DOI] [PubMed] [Google Scholar]
- 58.Bortone Agustín, Appetiti Anthony, Bouzeman Abdeslam, Maupas Eric, Ciobotaru Vlad, Boulenc Jean-Marc, Pujadas-Berthault Pénélope, Rioux Philippe. Unipolar signal modification as a guide for lesion creation during radiofrequency application in the left atrium: prospective study in humans in the setting of paroxysmal atrial fibrillation catheter ablation. Circ Arrhythm Electrophysiol. 2013 Dec;6 (6):1095–102. doi: 10.1161/CIRCEP.113.000749. [DOI] [PubMed] [Google Scholar]
- 59.Kosmidou Ioanna, Houde-Walter Haley, Foley Lori, Michaud Gregory. Loss of pace capture after radiofrequency application predicts the formation of uniform transmural lesions. Europace. 2013 Apr;15 (4):601–6. doi: 10.1093/europace/eus381. [DOI] [PubMed] [Google Scholar]
- 60.Steven Daniel, Reddy Vivek Y, Inada Keiichi, Roberts-Thomson Kurt C, Seiler Jens, Stevenson William G, Michaud Gregory F. Loss of pace capture on the ablation line: a new marker for complete radiofrequency lesions to achieve pulmonary vein isolation. Heart Rhythm. 2010 Mar;7 (3):323–30. doi: 10.1016/j.hrthm.2009.11.011. [DOI] [PubMed] [Google Scholar]
- 61.Steven Daniel, Sultan Arian, Reddy Vivek, Luker Jakob, Altenburg Manuel, Hoffmann Boris, Rostock Thomas, Servatius Helge, Stevenson William G, Willems Stephan, Michaud Gregory F. Benefit of pulmonary vein isolation guided by loss of pace capture on the ablation line: results from a prospective 2-center randomized trial. J. Am. Coll. Cardiol. 2013 Jul 02;62 (1):44–50. doi: 10.1016/j.jacc.2013.03.059. [DOI] [PubMed] [Google Scholar]
- 62.Andrade Jason G, Pollak Scott J, Monir George, Khairy Paul, Dubuc Marc, Roy Denis, Talajic Mario, Deyell Marc, Rivard Léna, Thibault Bernard, Guerra Peter G, Nattel Stanley, Macle Laurent. Pulmonary vein isolation using a pace-capture-guided versus an adenosine-guided approach: effect on dormant conduction and long-term freedom from recurrent atrial fibrillation--a prospective study. Circ Arrhythm Electrophysiol. 2013 Dec;6 (6):1103–8. doi: 10.1161/CIRCEP.113.000454. [DOI] [PubMed] [Google Scholar]
- 63.Ranjan Ravi, Kholmovski Eugene G, Blauer Joshua, Vijayakumar Sathya, Volland Nelly A, Salama Mohamed E, Parker Dennis L, MacLeod Rob, Marrouche Nassir F. Identification and acute targeting of gaps in atrial ablation lesion sets using a real-time magnetic resonance imaging system. Circ Arrhythm Electrophysiol. 2012 Dec;5 (6):1130–5. doi: 10.1161/CIRCEP.112.973164. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Kolandaivelu Aravindan, Zviman Menekhem M, Castro Valeria, Lardo Albert C, Berger Ronald D, Halperin Henry R. Noninvasive assessment of tissue heating during cardiac radiofrequency ablation using MRI thermography. Circ Arrhythm Electrophysiol. 2010 Oct;3 (5):521–9. doi: 10.1161/CIRCEP.110.942433. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Bordignon Stefano, Chun Kyoung-Ryul Julian, Gunawardene Melanie, Schulte-Hahn Britta, Nowak Bernd, Fuernkranz Alexander, Schmidt Boris. Endoscopic ablation systems. Expert Rev Med Devices. 2013 Mar;10 (2):177–83. doi: 10.1586/erd.12.86. [DOI] [PubMed] [Google Scholar]
- 66.Andrikopoulos George, Tzeis Stylianos, Vardas Panos E. Invasive therapy for atrial fibrillation: recent developments in ablation, navigation and mapping technology. Heart. 2011 Feb;97 (3):237–43. doi: 10.1136/hrt.2009.190017. [DOI] [PubMed] [Google Scholar]
- 67.Dewire Jane, Calkins Hugh. Update on atrial fibrillation catheter ablation technologies and techniques. Nat Rev Cardiol. 2013 Oct;10 (10):599–612. doi: 10.1038/nrcardio.2013.121. [DOI] [PubMed] [Google Scholar]
- 68.Natale Andrea, Reddy Vivek Y, Monir George, Wilber David J, Lindsay Bruce D, McElderry H Thomas, Kantipudi Charan, Mansour Moussa C, Melby Daniel P, Packer Douglas L, Nakagawa Hiroshi, Zhang Baohui, Stagg Robert B, Boo Lee Ming, Marchlinski Francis E. Paroxysmal AF catheter ablation with a contact force sensing catheter: results of the prospective, multicenter SMART-AF trial. J. Am. Coll. Cardiol. 2014 Aug 19;64 (7):647–56. doi: 10.1016/j.jacc.2014.04.072. [DOI] [PubMed] [Google Scholar]
- 69.Perna Francesco, Heist E Kevin, Danik Stephan B, Barrett Conor D, Ruskin Jeremy N, Mansour Moussa. Assessment of catheter tip contact force resulting in cardiac perforation in swine atria using force sensing technology. Circ Arrhythm Electrophysiol. 2011 Apr;4 (2):218–24. doi: 10.1161/CIRCEP.110.959429. [DOI] [PubMed] [Google Scholar]
- 70.Kuck Karl-Heinz, Reddy Vivek Y, Schmidt Boris, Natale Andrea, Neuzil Petr, Saoudi Nadir, Kautzner Josef, Herrera Claudia, Hindricks Gerhard, Jaïs Pierre, Nakagawa Hiroshi, Lambert Hendrik, Shah Dipen C. A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm. 2012 Jan;9 (1):18–23. doi: 10.1016/j.hrthm.2011.08.021. [DOI] [PubMed] [Google Scholar]
- 71.Reddy Vivek Y, Shah Dipen, Kautzner Josef, Schmidt Boris, Saoudi Nadir, Herrera Claudia, Jaïs Pierre, Hindricks Gerhard, Peichl Petr, Yulzari Aude, Lambert Hendrik, Neuzil Petr, Natale Andrea, Kuck Karl-Heinz. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm. 2012 Nov;9 (11):1789–95. doi: 10.1016/j.hrthm.2012.07.016. [DOI] [PubMed] [Google Scholar]
- 72.Neuzil Petr, Reddy Vivek Y, Kautzner Josef, Petru Jan, Wichterle Dan, Shah Dipen, Lambert Hendrik, Yulzari Aude, Wissner Erik, Kuck Karl-Heinz. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study. Circ Arrhythm Electrophysiol. 2013 Apr;6 (2):327–33. doi: 10.1161/CIRCEP.113.000374. [DOI] [PubMed] [Google Scholar]
- 73.Martinek Martin, Lemes Christine, Sigmund Elisabeth, Derndorfer Michael, Aichinger Josef, Winter Siegmund, Nesser Hans-Joachim, Pürerfellner Helmut. Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. Pacing Clin Electrophysiol. 2012 Nov;35 (11):1312–8. doi: 10.1111/j.1540-8159.2012.03503.x. [DOI] [PubMed] [Google Scholar]
- 74.Shah Dipen C, Lambert Hendrik, Nakagawa Hiroshi, Langenkamp Arne, Aeby Nicolas, Leo Giovanni. Area under the real-time contact force curve (force-time integral) predicts radiofrequency lesion size in an in vitro contractile model. J. Cardiovasc. Electrophysiol. 2010 Sep;21 (9):1038–43. doi: 10.1111/j.1540-8167.2010.01750.x. [DOI] [PubMed] [Google Scholar]
- 75.Weiss Christian, Antz Matthias, Eick Olaf, Eshagzaiy Kablai, Meinertz Thomas, Willems Stephan. Radiofrequency catheter ablation using cooled electrodes: impact of irrigation flow rate and catheter contact pressure on lesion dimensions. Pacing Clin Electrophysiol. 2002 Apr;25 (4 Pt 1):463–9. doi: 10.1046/j.1460-9592.2002.00463.x. [DOI] [PubMed] [Google Scholar]
- 76.Haines DE. Determinants of lesion size during radiofrequency catheter ablation: The role of electrode-tissue contact pressure and duration of energy delivery. Journal of cardiovascular electrophysiology. 1991;2:509–515. [Google Scholar]
- 77.Wong Michael C G, Edwards Glenn, Spence Steven J, Kalman Jonathan M, Kumar Saurabh, Joseph Stephen A, Morton Joseph B. Characterization of catheter-tissue contact force during epicardial radiofrequency ablation in an ovine model. Circ Arrhythm Electrophysiol. 2013 Dec;6 (6):1222–8. doi: 10.1161/CIRCEP.113.000184. [DOI] [PubMed] [Google Scholar]
- 78.Okumura Yasuo, Johnson Susan B, Bunch T Jared, Henz Benhur D, O'Brien Christine J, Packer Douglas L. A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. J. Cardiovasc. Electrophysiol. 2008 Jun;19 (6):632–40. doi: 10.1111/j.1540-8167.2008.01135.x. [DOI] [PubMed] [Google Scholar]
- 79.Kalman J M, Fitzpatrick A P, Olgin J E, Chin M C, Lee R J, Scheinman M M, Lesh M D. Biophysical characteristics of radiofrequency lesion formation in vivo: dynamics of catheter tip-tissue contact evaluated by intracardiac echocardiography. Am. Heart J. 1997 Jan;133 (1):8–18. doi: 10.1016/s0002-8703(97)70242-4. [DOI] [PubMed] [Google Scholar]
- 80.Strickberger S A, Vorperian V R, Man K C, Williamson B D, Kalbfleisch S J, Hasse C, Morady F, Langberg J J. Relation between impedance and endocardial contact during radiofrequency catheter ablation. Am. Heart J. 1994 Aug;128 (2):226–9. doi: 10.1016/0002-8703(94)90472-3. [DOI] [PubMed] [Google Scholar]
- 81.Koa-Wing Michael, Kojodjojo Pipin, Malcolme-Lawes Louisa C, Salukhe Tushar V, Linton Nick W F, Grogan Aaron P, Bergman Dale, Lim Phang Boon, Whinnett Zachary I, McCarthy Karen, Ho Siew Yen, O'Neill Mark D, Peters Nicholas S, Davies D Wyn, Kanagaratnam Prapa. Robotically assisted ablation produces more rapid and greater signal attenuation than manual ablation. J. Cardiovasc. Electrophysiol. 2009 Dec;20 (12):1398–404. doi: 10.1111/j.1540-8167.2009.01590.x. [DOI] [PubMed] [Google Scholar]
- 82.Azegami K, Satake S, Okishige K, Sasano T, Ohira H, Yamashita K. Monitoring the local electrogram at the ablation site during radiofrequency application for common atrial flutter. Jpn. Circ. J. 1998 Aug;62 (8):559–64. doi: 10.1253/jcj.62.559. [DOI] [PubMed] [Google Scholar]
- 83.Kumar Saurabh, Chan Martin, Lee Justin, Wong Michael C G, Yudi Matias, Morton Joseph B, Spence Steven J, Halloran Karen, Kistler Peter M, Kalman Jonathan M. Catheter-tissue contact force determines atrial electrogram characteristics before and lesion efficacy after antral pulmonary vein isolation in humans. J. Cardiovasc. Electrophysiol. 2014 Feb;25 (2):122–9. doi: 10.1111/jce.12293. [DOI] [PubMed] [Google Scholar]
- 84.Sasaki Naoko, Okumura Yasuo, Watanabe Ichiro, Sonoda Kazumasa, Kogawa Rikitake, Takahashi Keiko, Iso Kazuki, Nakahara Shiro, Maruyama Ayako, Takemura Saya, Hirayama Atsushi. Relations between contact force, bipolar voltage amplitude, and mapping point distance from the left atrial surfaces of 3D ultrasound- and merged 3D CT-derived images: Implication for atrial fibrillation mapping and ablation. Heart Rhythm. 2015 Jan;12 (1):36–43. doi: 10.1016/j.hrthm.2014.09.007. [DOI] [PubMed] [Google Scholar]
- 85.Sacher Frederic, Wright Matthew, Derval Nicolas, Denis Arnaud, Ramoul Khaled, Roten Laurent, Pascale Patrizzio, Bordachar Pierre, Ritter Philippe, Hocini Meleze, Dos Santos Pierre, Haissaguerre Michel, Jais Pierre. Endocardial versus epicardial ventricular radiofrequency ablation: utility of in vivo contact force assessment. Circ Arrhythm Electrophysiol. 2013 Feb;6 (1):144–50. doi: 10.1161/CIRCEP.111.974501. [DOI] [PubMed] [Google Scholar]
- 86.Tilz Roland Richard, Makimoto Hisaki, Lin Tina, Rillig Andreas, Metzner Andreas, Mathew Shibu, Deiss Sebastian, Wissner Erik, Rausch Peter, Kamioka Masashi, Heeger Christian, Kuck Karl-Heinz, Ouyang Feifan. In vivo left-ventricular contact force analysis: comparison of antegrade transseptal with retrograde transaortic mapping strategies and correlation of impedance and electrical amplitude with contact force. Europace. 2014 Sep;16 (9):1387–95. doi: 10.1093/europace/eut430. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Kumar Saurabh, Morton Joseph B, Lee Justin, Halloran Karen, Spence Steven J, Gorelik Alexandra, Hepworth Graham, Kistler Peter M, Kalman Jonathan M. Prospective characterization of catheter-tissue contact force at different anatomic sites during antral pulmonary vein isolation. Circ Arrhythm Electrophysiol. 2012 Dec;5 (6):1124–9. doi: 10.1161/CIRCEP.112.972208. [DOI] [PubMed] [Google Scholar]
- 88.Kistler Peter M, Ho Siew Yen, Rajappan Kim, Morper Michael, Harris Stuart, Abrams Dominic, Sporton Simon C, Schilling Richard J. Electrophysiologic and anatomic characterization of sites resistant to electrical isolation during circumferential pulmonary vein ablation for atrial fibrillation: a prospective study. J. Cardiovasc. Electrophysiol. 2007 Dec;18 (12):1282–8. doi: 10.1111/j.1540-8167.2007.00981.x. [DOI] [PubMed] [Google Scholar]
- 89.Kumar Saurabh, Morton Joseph B, Halloran Karen, Spence Steven J, Lee Geoffrey, Wong Michael C G, Kistler Peter M, Kalman Jonathan M. Effect of respiration on catheter-tissue contact force during ablation of atrial arrhythmias. Heart Rhythm. 2012 Jul;9 (7):1041–1047.e1. doi: 10.1016/j.hrthm.2012.02.015. [DOI] [PubMed] [Google Scholar]
- 90.Haldar Shouvik, Jarman Julian W E, Panikker Sandeep, Jones David G, Salukhe Tushar, Gupta Dhiraj, Wynn Gareth, Hussain Wajid, Markides Vias, Wong Tom. Contact force sensing technology identifies sites of inadequate contact and reduces acute pulmonary vein reconnection: a prospective case control study. Int. J. Cardiol. 2013 Sep 30;168 (2):1160–6. doi: 10.1016/j.ijcard.2012.11.072. [DOI] [PubMed] [Google Scholar]
- 91.Evelyn Chris R, Wade Susan M, Wang Qin, Wu Mei, Iñiguez-Lluhí Jorge A, Merajver Sofia D, Neubig Richard R. CCG-1423: a small-molecule inhibitor of RhoA transcriptional signaling. Mol. Cancer Ther. 2007 Aug;6 (8):2249–60. doi: 10.1158/1535-7163.MCT-06-0782. [DOI] [PubMed] [Google Scholar]
- 92.Squara Fabien, Latcu Decebal Gabriel, Massaad Youssef, Mahjoub Marouane, Bun Sok-Sithikun, Saoudi Nadir. Contact force and force-time integral in atrial radiofrequency ablation predict transmurality of lesions. Europace. 2014 May;16 (5):660–7. doi: 10.1093/europace/euu068. [DOI] [PubMed] [Google Scholar]
- 93.Matía Francés Roberto, Hernández Madrid Antonio, Delgado Antonia, Carrizo Laura, Pindado Carlos, Moro Serrano Concepción, Zamorano Gómez José L. Characterization of the impact of catheter-tissue contact force in lesion formation during cavo-tricuspid isthmus ablation in an experimental swine model. Europace. 2014 Nov;16 (11):1679–83. doi: 10.1093/europace/eut351. [DOI] [PubMed] [Google Scholar]
- 94.Sotomi Yohei, Kikkawa Takayuki, Inoue Koichi, Tanaka Koji, Toyoshima Yuko, Oka Takafumi, Tanaka Nobuaki, Nozato Yoichi, Orihara Yoshiyuki, Iwakura Katsuomi, Sakata Yasushi, Fujii Kenshi. Regional difference of optimal contact force to prevent acute pulmonary vein reconnection during radiofrequency catheter ablation for atrial fibrillation. J. Cardiovasc. Electrophysiol. 2014 Sep;25 (9):941–7. doi: 10.1111/jce.12443. [DOI] [PubMed] [Google Scholar]
- 95.Andrade Jason G, Monir George, Pollak Scott J, Khairy Paul, Dubuc Marc, Roy Denis, Talajic Mario, Deyell Marc, Rivard Léna, Thibault Bernard, Guerra Peter G, Nattel Stanley, Macle Laurent. Pulmonary vein isolation using "contact force" ablation: the effect on dormant conduction and long-term freedom from recurrent atrial fibrillation--a prospective study. Heart Rhythm. 2014 Nov;11 (11):1919–24. doi: 10.1016/j.hrthm.2014.07.033. [DOI] [PubMed] [Google Scholar]
- 96.Park Chan-Il, Lehrmann Heiko, Keyl Cornelius, Weber Reinhold, Schiebeling Jochen, Allgeier Juergen, Schurr Patrick, Shah Ashok, Neumann Franz-Josef, Arentz Thomas, Jadidi Amir S. Mechanisms of pulmonary vein reconnection after radiofrequency ablation of atrial fibrillation: the deterministic role of contact force and interlesion distance. J. Cardiovasc. Electrophysiol. 2014 Jul;25 (7):701–8. doi: 10.1111/jce.12396. [DOI] [PubMed] [Google Scholar]
- 97.Sohns Christian, Karim Rashed, Harrison James, Arujuna Aruna, Linton Nick, Sennett Richard, Lambert Hendrik, Leo Giovanni, Williams Steven, Razavi Reza, Wright Matt, Schaeffter Tobias, O'Neill Mark, Rhode Kawal. Quantitative magnetic resonance imaging analysis of the relationship between contact force and left atrial scar formation after catheter ablation of atrial fibrillation. J. Cardiovasc. Electrophysiol. 2014 Feb;25 (2):138–45. doi: 10.1111/jce.12298. [DOI] [PubMed] [Google Scholar]
- 98.Marijon Eloi, Fazaa Samia, Narayanan Kumar, Guy-Moyat Benoit, Bouzeman Abdeslam, Providencia Rui, Treguer Frederic, Combes Nicolas, Bortone Agustin, Boveda Serge, Combes Stephane, Albenque Jean-Paul. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: procedural and 1-year results. J. Cardiovasc. Electrophysiol. 2014 Feb;25 (2):130–7. doi: 10.1111/jce.12303. [DOI] [PubMed] [Google Scholar]
- 99.Providência Rui, Marijon Eloi, Combes Stéphane, Bouzeman Abdeslam, Jourda François, Khoueiry Ziad, Cardin Christelle, Combes Nicolas, Boveda Serge, Albenque Jean-Paul. Higher contact-force values associated with better mid-term outcome of paroxysmal atrial fibrillation ablation using the SmartTouch™ catheter. Europace. 2015 Jan;17 (1):56–63. doi: 10.1093/europace/euu218. [DOI] [PubMed] [Google Scholar]
- 100.Sciarra Luigi, Golia Paolo, Natalizia Andrea, De Ruvo Ermenegildo, Dottori Serena, Scarà Antonio, Borrelli Alessio, De Luca Lucia, Rebecchi Marco, Fagagnini Alessandro, Bandini Alberto, Guarracini Fabrizio, Galvani Marcello, Calò Leonardo. Which is the best catheter to perform atrial fibrillation ablation? A comparison between standard ThermoCool, SmartTouch, and Surround Flow catheters. J Interv Card Electrophysiol. 2014 Apr;39 (3):193–200. doi: 10.1007/s10840-014-9874-2. [DOI] [PubMed] [Google Scholar]
- 101.Kimura Masaomi, Sasaki Shingo, Owada Shingen, Horiuchi Daisuke, Sasaki Kenichi, Itoh Taihei, Ishida Yuji, Kinjo Takahiko, Tomita Hirofumi, Okumura Ken. Comparison of lesion formation between contact force-guided and non-guided circumferential pulmonary vein isolation: a prospective, randomized study. Heart Rhythm. 2014 Jun;11 (6):984–91. doi: 10.1016/j.hrthm.2014.03.019. [DOI] [PubMed] [Google Scholar]
- 102.Ho S Y, Cabrera J A, Tran V H, Farré J, Anderson R H, Sánchez-Quintana D. Architecture of the pulmonary veins: relevance to radiofrequency ablation. Heart. 2001 Sep;86 (3):265–70. doi: 10.1136/heart.86.3.265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103.Hutchinson Mathew D, Garcia Fermin C, Mandel Jeff E, Elkassabany Nabil, Zado Erica S, Riley Michael P, Cooper Joshua M, Bala Rupa, Frankel David S, Lin David, Supple Gregory E, Dixit Sanjay, Gerstenfeld Edward P, Callans David J, Marchlinski Francis E. Efforts to enhance catheter stability improve atrial fibrillation ablation outcome. Heart Rhythm. 2013 Mar;10 (3):347–53. doi: 10.1016/j.hrthm.2012.10.044. [DOI] [PubMed] [Google Scholar]
- 104.Anter Elad, Tschabrunn Cory M, Contreras-Valdes Fernando M, Buxton Alfred E, Josephson Mark E. Radiofrequency ablation annotation algorithm reduces the incidence of linear gaps and reconnection after pulmonary vein isolation. Heart Rhythm. 2014 May;11 (5):783–90. doi: 10.1016/j.hrthm.2014.02.022. [DOI] [PubMed] [Google Scholar]
- 105.Reichlin Tobias, Michaud Gregory F. Our approach to maximizing the durability of pulmonary vein isolation during a paroxysmal atrial fibrillation ablation procedure. J. Cardiovasc. Electrophysiol. 2012 Nov;23 (11):1272–6. doi: 10.1111/j.1540-8167.2012.02414.x. [DOI] [PubMed] [Google Scholar]
- 106.Datino Tomás, Macle Laurent, Qi Xiao-Yan, Maguy Ange, Comtois Philippe, Chartier Denis, Guerra Peter G, Arenal Angel, Fernández-Avilés Francisco, Nattel Stanley. Mechanisms by which adenosine restores conduction in dormant canine pulmonary veins. Circulation. 2010 Mar 02;121 (8):963–72. doi: 10.1161/CIRCULATIONAHA.109.893107. [DOI] [PubMed] [Google Scholar]
- 107.Yamane Teiichi, Matsuo Seiichiro, Date Taro, Lellouche Nicolas, Hioki Mika, Narui Ryosuke, Ito Keiichi, Tanigawa Shin-ichi, Yamashita Seigo, Tokuda Michifumi, Yoshida Hiroshi, Inada Keiichi, Shibayama Kenri, Miyanaga Satoru, Miyazaki Hidekazu, Abe Kunihiko, Sugimoto Ken-ichi, Yoshimura Michihiro. Repeated provocation of time- and ATP-induced early pulmonary vein reconnections after pulmonary vein isolation: eliminating paroxysmal atrial fibrillation in a single procedure. Circ Arrhythm Electrophysiol. 2011 Oct;4 (5):601–8. doi: 10.1161/CIRCEP.110.960138. [DOI] [PubMed] [Google Scholar]
- 108.McLellan Alex J A, Kumar Saurabh, Smith Catherine, Morton Joseph B, Kalman Jonathan M, Kistler Peter M. The role of adenosine following pulmonary vein isolation in patients undergoing catheter ablation for atrial fibrillation: a systematic review. J. Cardiovasc. Electrophysiol. 2013 Jul;24 (7):742–51. doi: 10.1111/jce.12121. [DOI] [PubMed] [Google Scholar]
- 109.Macle Laurent, Khairy Paul, Weerasooriya Rukshen, Novak Paul, Verma Atul, Willems Stephan, Arentz Thomas, Deisenhofer Isabel, Veenhuyzen George, Scavée Christophe, Jaïs Pierre, Puererfellner Helmut, Levesque Sylvie, Andrade Jason G, Rivard Lena, Guerra Peter G, Dubuc Marc, Thibault Bernard, Talajic Mario, Roy Denis, Nattel Stanley. Adenosine-guided pulmonary vein isolation for the treatment of paroxysmal atrial fibrillation: an international, multicentre, randomised superiority trial. Lancet. 2015 Aug 15;386 (9994):672–9. doi: 10.1016/S0140-6736(15)60026-5. [DOI] [PubMed] [Google Scholar]
- 110.Koruth Jacob S, Dukkipati Srinivas, Miller Marc A, Neuzil Petr, d'Avila Andre, Reddy Vivek Y. Bipolar irrigated radiofrequency ablation: a therapeutic option for refractory intramural atrial and ventricular tachycardia circuits. Heart Rhythm. 2012 Dec;9 (12):1932–41. doi: 10.1016/j.hrthm.2012.08.001. [DOI] [PubMed] [Google Scholar]
- 111.Sivagangabalan Gopal, Barry Michael A, Huang Kaimin, Lu Juntang, Pouliopoulos Jim, Thomas Stuart P, Ross David L, Thiagalingam Aravinda, Kovoor Pramesh. Bipolar ablation of the interventricular septum is more efficient at creating a transmural line than sequential unipolar ablation. Pacing Clin Electrophysiol. 2010 Jan;33 (1):16–26. doi: 10.1111/j.1540-8159.2009.02602.x. [DOI] [PubMed] [Google Scholar]
- 112.Nagashima Koichi, Watanabe Ichiro, Okumura Yasuo, Ohkubo Kimie, Kofune Masayoshi, Ohya Toshiyuki, Kasamaki Yuji, Hirayama Atsushi. Lesion formation by ventricular septal ablation with irrigated electrodes: comparison of bipolar and sequential unipolar ablation. Circ. J. 2011;75 (3):565–70. doi: 10.1253/circj.cj-10-0870. [DOI] [PubMed] [Google Scholar]
- 113.Nagashima Koichi, Watanabe Ichiro, Okumura Yasuo, Sonoda Kazumasa, Kofune Masayoshi, Mano Hiroaki, Ohkubo Kimie, Nakai Toshiko, Kunimoto Satoshi, Kasamaki Yuji, Hirayama Atsushi. Epicardial ablation with irrigated electrodes: – effect of bipolar vs. unipolar ablation on lesion formation –. Circ. J. 2012;76 (2):322–7. doi: 10.1253/circj.cj-11-0984. [DOI] [PubMed] [Google Scholar]
- 114.Gizurarson Sigfus, Spears Danna, Sivagangabalan Gopal, Farid Talha, Ha Andrew C T, Massé Stéphane, Kusha Marjan, Chauhan Vijay S, Nair Krishnakumar, Harris Louise, Downar Eugene, Nanthakumar Kumaraswamy. Bipolar ablation for deep intra-myocardial circuits: human ex vivo development and in vivo experience. Europace. 2014 Nov;16 (11):1684–8. doi: 10.1093/europace/euu001. [DOI] [PubMed] [Google Scholar]
- 115.Wittkampf Fred H M, van Driel Vincent J, van Wessel Harry, Neven Kars G E J, Gründeman Paul F, Vink Aryan, Loh Peter, Doevendans Pieter A. Myocardial lesion depth with circular electroporation ablation. Circ Arrhythm Electrophysiol. 2012 Jun 01;5 (3):581–6. doi: 10.1161/CIRCEP.111.970079. [DOI] [PubMed] [Google Scholar]
- 116.Nakagawa Hiroshi, Jackman Warren M. Electroporation (revival of direct current ablation): new approach for increasing epicardial ablation safety in close proximity to a coronary artery. Circ Arrhythm Electrophysiol. 2014 Oct;7 (5):779–80. doi: 10.1161/CIRCEP.114.002255. [DOI] [PubMed] [Google Scholar]
- 117.Viles-Gonzalez Juan F, de Castro Miranda Reynaldo, Scanavacca Mauricio, Sosa Eduardo, d'Avila Andre. Acute and chronic effects of epicardial radiofrequency applications delivered on epicardial coronary arteries. Circ Arrhythm Electrophysiol. 2011 Aug;4 (4):526–31. doi: 10.1161/CIRCEP.110.961508. [DOI] [PubMed] [Google Scholar]
- 118.Paul T, Bökenkamp R, Mahnert B, Trappe H J. Coronary artery involvement early and late after radiofrequency current application in young pigs. Am. Heart J. 1997 Apr;133 (4):436–40. doi: 10.1016/s0002-8703(97)70185-6. [DOI] [PubMed] [Google Scholar]
- 119.Aoyama Hiroshi, Nakagawa Hiroshi, Pitha Jan V, Khammar George S, Chandrasekaran Krishnaswamy, Matsudaira Kagari, Yagi Tetsuo, Yokoyama Katsuaki, Lazzara Ralph, Jackman Warren M. Comparison of cryothermia and radiofrequency current in safety and efficacy of catheter ablation within the canine coronary sinus close to the left circumflex coronary artery. J. Cardiovasc. Electrophysiol. 2005 Nov;16 (11):1218–26. doi: 10.1111/j.1540-8167.2005.50126.x. [DOI] [PubMed] [Google Scholar]
- 120.Roberts-Thomson Kurt C, Steven Daniel, Seiler Jens, Inada Keiichi, Koplan Bruce A, Tedrow Usha B, Epstein Laurence M, Stevenson William G. Coronary artery injury due to catheter ablation in adults: presentations and outcomes. Circulation. 2009 Oct 13;120 (15):1465–73. doi: 10.1161/CIRCULATIONAHA.109.870790. [DOI] [PubMed] [Google Scholar]
- 121.Pardo Meo Januário, Scanavacca Maurício, Sosa Eduardo, Correia Aristides, Hachul Denise, Darrieux Francisco, Lara Sissy, Hardy Carina, Jatene Fabio, Jatene Marcelo. Atrial coronary arteries in areas involved in atrial fibrillation catheter ablation. Circ Arrhythm Electrophysiol. 2010 Dec;3 (6):600–5. doi: 10.1161/CIRCEP.110.951525. [DOI] [PubMed] [Google Scholar]
- 122.du Pré Bastiaan C, van Driel Vincent J, van Wessel Harry, Loh Peter, Doevendans Pieter A, Goldschmeding Roel, Wittkampf Fred H, Vink Aryan. Minimal coronary artery damage by myocardial electroporation ablation. Europace. 2013 Jan;15 (1):144–9. doi: 10.1093/europace/eus171. [DOI] [PubMed] [Google Scholar]
- 123.Neven Kars, van Driel Vincent, van Wessel Harry, van Es René, Doevendans Pieter A, Wittkampf Fred. Myocardial lesion size after epicardial electroporation catheter ablation after subxiphoid puncture. Circ Arrhythm Electrophysiol. 2014 Aug;7 (4):728–33. doi: 10.1161/CIRCEP.114.001659. [DOI] [PubMed] [Google Scholar]
- 124.Neven Kars, van Driel Vincent, van Wessel Harry, van Es René, Doevendans Pieter A, Wittkampf Fred. Epicardial linear electroporation ablation and lesion size. Heart Rhythm. 2014 Aug;11 (8):1465–70. doi: 10.1016/j.hrthm.2014.04.031. [DOI] [PubMed] [Google Scholar]
- 125.Neven Kars, van Driel Vincent, van Wessel Harry, van Es René, du Pré Bastiaan, Doevendans Pieter A, Wittkampf Fred. Safety and feasibility of closed chest epicardial catheter ablation using electroporation. Circ Arrhythm Electrophysiol. 2014 Oct;7 (5):913–9. doi: 10.1161/CIRCEP.114.001607. [DOI] [PubMed] [Google Scholar]
- 126.van Driel Vincent J H M, Neven Kars G E J, van Wessel Harry, du Pré Bastiaan C, Vink Aryan, Doevendans Pieter A F M, Wittkampf Fred H M. Pulmonary vein stenosis after catheter ablation: electroporation versus radiofrequency. Circ Arrhythm Electrophysiol. 2014 Aug;7 (4):734–8. doi: 10.1161/CIRCEP.113.001111. [DOI] [PubMed] [Google Scholar]
- 127.Wittkampf Fred H, van Driel Vincent J, van Wessel Harry, Vink Aryan, Hof Irene E, Gründeman Paul F, Hauer Richard N W, Loh Peter. Feasibility of electroporation for the creation of pulmonary vein ostial lesions. J. Cardiovasc. Electrophysiol. 2011 Mar;22 (3):302–9. doi: 10.1111/j.1540-8167.2010.01863.x. [DOI] [PubMed] [Google Scholar]
- 128.Fuller Ithiel A, Wood Mark A. Intramural coronary vasculature prevents transmural radiofrequency lesion formation: implications for linear ablation. Circulation. 2003 Apr 08;107 (13):1797–803. doi: 10.1161/01.CIR.0000058705.97823.F4. [DOI] [PubMed] [Google Scholar]
- 129.D'Avila André, Gutierrez Paulo, Scanavacca Mauricio, Reddy Vivek, Lustgarten Daniel L, Sosa Eduardo, Ramires José Antonio F. Effects of radiofrequency pulses delivered in the vicinity of the coronary arteries: implications for nonsurgical transthoracic epicardial catheter ablation to treat ventricular tachycardia. Pacing Clin Electrophysiol. 2002 Oct;25 (10):1488–95. doi: 10.1046/j.1460-9592.2002.01488.x. [DOI] [PubMed] [Google Scholar]
- 130.Betensky Brian P, Jauregui Miguel, Campos Bieito, Michele John, Marchlinski Francis E, Oley Leslie, Wylie Bryan, Robinson David, Gerstenfeld Edward P. Use of a novel endoscopic catheter for direct visualization and ablation in an ovine model of chronic myocardial infarction. Circulation. 2012 Oct 23;126 (17):2065–72. doi: 10.1161/CIRCULATIONAHA.112.112540. [DOI] [PubMed] [Google Scholar]
- 131.Chik William W B, Barry M A, Malchano Zach, Wylie Bryan, Pouliopoulos Jim, Huang Kaimin, Lu Juntang, Thavapalachandran Sujitha, Robinson David, Saadat Vahid, Thomas Stuart P, Ross David L, Kovoor Pramesh, Thiagalingam Aravinda. In vivo evaluation of virtual electrode mapping and ablation utilizing a direct endocardial visualization ablation catheter. J. Cardiovasc. Electrophysiol. 2012 Jan;23 (1):88–95. doi: 10.1111/j.1540-8167.2011.02169.x. [DOI] [PubMed] [Google Scholar]
- 132.Sacher Frederic, Derval Nicolas, Jadidi Amir, Scherr Daniel, Hocini Meleze, Haissaguerre Michel, Dos Santos Pierre, Jais Pierre. Comparison of ventricular radiofrequency lesions in sheep using standard irrigated tip catheter versus catheter ablation enabling direct visualization. J. Cardiovasc. Electrophysiol. 2012 Aug;23 (8):869–73. doi: 10.1111/j.1540-8167.2012.02338.x. [DOI] [PubMed] [Google Scholar]
- 133.Fahey Brian J, Nightingale Kathryn R, McAleavey Stephen A, Palmeri Mark L, Wolf Patrick D, Trahey Gregg E. Acoustic radiation force impulse imaging of myocardial radiofrequency ablation: initial in vivo results. IEEE Trans Ultrason Ferroelectr Freq Control. 2005 Apr;52 (4):631–41. doi: 10.1109/tuffc.2005.1428046. [DOI] [PubMed] [Google Scholar]
- 134.Eyerly Stephanie A, Hsu Stephen J, Agashe Shruti H, Trahey Gregg E, Li Yang, Wolf Patrick D. An in vitro assessment of acoustic radiation force impulse imaging for visualizing cardiac radiofrequency ablation lesions. J. Cardiovasc. Electrophysiol. 2010 May;21 (5):557–63. doi: 10.1111/j.1540-8167.2009.01664.x. [DOI] [PMC free article] [PubMed] [Google Scholar]