Table 3. Studies comparing conventional biventricular pacing with multisite pacing through a quadripolar lead.
NYHA, New York Heart Association functional class
First author, year | n | Comparison | Study findings |
---|---|---|---|
Thibault, 2013 (88) | 19 | Distal electrode versus 4 multisite pacing configurations | Invasive dP/dt better with multisite pacing in 72% (mostly through combined proximal and distal electrode stimulation) |
Rinaldi, 2013 (89, 91) | 41/40 | Distal electrode versus 8 multisite pacing configurations | 1. Dyssynchrony by tissue Doppler echocardiography reduced in 63% with multisite pacing 2. Radial strain >20% higher in 63% with multisite pacing |
Pappone, 2014 (90) | 44 | Distal/proximal electrode versus 7 multisite pacing configurations | Acute improvement in hemodynamic parameters and significant decrease in NYHA with more pronounced reverse remodeling after 3 months in multisite pacing group |