TABLE I.

Calculated maximum beam current for a multisource configuration, as a function of dwell time and number of focal spots during a 5 s scan, with one x-ray source always turned on at a given moment of time (protocol considers going through all sources, before returning to any given source).

Beam current/spot (mA) Number of focal spots
Dwell time (μs) 4 20 30 40 50 100 200
10 67 246 316 369 411 530 620
20 64 216 269 307 335 410 461
50 60 175 208 229 245 282 306
100 56 143 165 178 188 209 222
200 51 115 128 136 141 153 160