Table 1.

Summary of overall figures of merit produced from comparing CAMERA-reconstructed (exponentially) subsampled 1H-15N HSQC spectra of ubiquitin with their true uniformly sampled spectrum. Matched peak counts are from a total of 121 reference peaks.

Meth. Matched Gained rint dH / ppm dN / ppm
PR 30% 117±1.6 2±1.2 0.996±0.014 0.001±0.0003 0.011±0.0016
10% 95±2.5 1±0.9 0.994±0.011 0.002±0.0003 0.009±0.0022
5% 89±3.7 3±2.5 0.961±0.024 0.003±0.0004 0.009±0.0023

SR 30% 119 1 0.998 0.002 0.011
10% 85 79 0.970 0.002 0.007
5% 53 151 0.956 0.001 0.005

PR+Jit 30% 118±0.9 1±1.0 0.994±0.054 0.002±0.0003 0.011±0.0016
10% 96±1.3 1±0.8 0.989±0.064 0.002±0.0003 0.010±0.0019
5% 88±1.4 3±1.8 0.954±0.041 0.003±0.0003 0.008±0.0014

SR+Jit 30% 118 2 0.973 0.002 0.011
10% 96 0 0.997 0.002 0.010
5% 88 5 0.960 0.003 0.009