Table 2. Results of proficiency testing surveys for cardiac troponin† (1).
Assay | Slope of regression equations‡ | |||
---|---|---|---|---|
2004 | 2009 | 2013 | 2016 | |
Tosoh AIA | 0.34 | – | 0.33 | 0.34 |
Ortho ECi | 0.15 | 0.11 | 0.46 | 0.62 |
Beckman Acces’s | –§ | 0.44 | 0.31 | 0.34 |
Abbott Architect | – | 0.52 | 0.89 | 0.36 |
Siemens Stratus CS | 0.70 | 0.15 | 0.51 | 0.54 |
Siemens Dimension | 1.00 | 1.00 | 1.00 | 1.00 |
Siemens Centaur | 1.20 | 1.50 | 1.46 | 0.86 |
Siemens Vista | – | 2.10 | 1.06 | 1.00 |
Abbott AxSYM | 4.50 | 0.60 | – | – |
Roche E6000 | – | – | 1.59 | 0.65 |
Ratio of slopes (high/low) | 30–X | 19–X | 5.2–X | 2.9–X |
†, results from the College of American Pathologists Proficiency Testing Survey for cardiac markers; ‡, slopes report from a plot of each assay against (y) vs. the Dade Dimension (x, arbitrarily selected); §, no data available.