Published in final edited form as: Appl Opt. 2015 Dec 10;54(35):10376–10396. doi: 10.1364/AO.54.010376

Table 5.

Responsivities for the TB wavelength ranges of VIIRS bands M1 through M7. The responsivities for the laboratory measurements come from the SpMA and the SIS100 The comparisons show the differences from SIRCUS.

VIIRS Band SpMA/SIS Responsivity1 SIRCUS Responsivity1 Responsivity Difference (%)
M1 19.966 19.993 −0.14
M2 24.505 23.600 3.83
M3 27.568 27.053 1.90
M4 38.427 36.888 4.17
M5 52.283 51.129 2.26
M6 86.195 85.410 0.92
M7 111.892 110.639 1.13
1

Units: DN/(W m−2 sr−1 μm−1)