Skip to main content
. 2016 Jun 14;8(6):e642. doi: 10.7759/cureus.642

Figure 1. N-Localizer and Its Interaction with the Tomographic Section.

Figure 1

Inline graphic Side view of the N-localizer. The tomographic section intersects the N-localizer at two vertical rods and one diagonal rod. Inline graphic Tomographic image. The intersection of the tomographic section with the N-localizer produces two circles and one ellipse. The relative spacing between the centers of the ellipse and the two circles varies according to the height at which the tomographic section intersects the diagonal rod. Measuring this spacing permits calculation of the position of the tomographic section with respect to the N-localizer [2].