Table 1.
MRI protocol | Axial T2 W SSFSE | Sagittal T2 W FRFSE | Oblique coronal T2 W FRFSE | Oblique axial T2 W FRFSE | Axial DWI SE EPI |
---|---|---|---|---|---|
Repetition time/Echo time (m s) | 765/59 | 3560/100 | 3440/100 | 3440/100 | 5425/74.8 |
Flip angle | 90° | 90° | 90° | 90° | 90° |
Echo train length | – | 21 | 23 | 23 | – |
Section thickness (mm) | 6 | 3 | 3 | 3 | 5 |
Interslice gap (mm) | 0.6 | 0,3 | 0.3 | 0.3 | 1 |
Bandwidht (kHz) | 31.25 | 25 | 31.25 | 31.25 | 250 |
Field of view (cm) | 38 | 30 | 28 | 28 | 40 |
Matrix | 320 × 288 | 288 × 224 | 288 × 224 | 288 × 224 | 160 × 160 |
No. of averages | 0.54 | 3 | 4 | 4 | 8 |
No. of images | 30 | 30 | 30 | 30 | 24 |
Frequency direction | Right to left | Anterior to posterior | Right to left | Anterior to posterior | Right to left |
Acquisition time | 24 s | 4 min 10 s | 4 min 41 s | 4 min 49 s | 2 min 59 s |
b-value (sec/mm2) | – | – | – | – | 0−800 |
T2W = T2-weighted, SSFSE = single-shot fast spin-echo, FRFSE = fast relaxation fast spin-echo, DWI = diffusion-weighted imaging, SE = spin-echo, EPI = echoplanar imaging.