Table 1. Neurophysiological parameters for mapping cranial motor nerves nuclei.

Stimulation
 Cathode: monopolar hand held probe
 Anode: monopolar needle in cervical muscles
 Wave form: square wave, single pulse
 Duration stimuli: 0.2 ms
 Frequency: 1.0–4.0 Hz
 Intensity: 2 mA for screening, then decrease the intensity to detect threshold
Recordings
 Type of electrodes: hook wires
 Filter: 20-3,000 Hz
 Epoch length: 20 ms
 Amplification: 10,000 times