Table 2.

Method performance.

Vitamin D metabolite LLoQ nmol/L Intra-assay CV% (nmol/L)a Inter-assay CV% (nmol/L)b Bias % (nmol/L)b
25OHD3 7.0 3.34 (40.6) 5.3 (38.4) 1.4 (37.9)
0.79 (95.2) 4.3 (91.7) 4.8 (87.5)


 

 

 

 


25OHD2 5.0 4.02 (40.4) 5.9 (46.8) 11.2 (42.1)
1.09 (93.6) 3.8 (113.1) 4.7 (108.0)


 

 

 

 


3-epi-25OHD3 4.0 2.07 (36.2) 6.0 (39.3) 14.9 (34.2)
2.32 (61.0) 3.7 (67.5) 23.8 (54.5)

LLoQ: Lower limit of quantification.

a

5 sequential injections of each concentration of the MassCheck® controls were performed. Values in parentheses are the means.

b

1 aliquot of each concentration of the MassCheck® controls was analyzed over a period of 14 consecutive months [n=90]. For the inter-assay repeatability, values in parentheses are those obtained by our method. For the bias, values in parentheses are those assigned by the manufacturer.