In Vivo Corneal Biomechanical Properties with Corneal Visualization Scheimpflug Technology in Chinese Population

Ying Wu 1, Lei Tian 2,*, Yi-fei Huang 1,*
PMCID: PMC4963556  PMID: 27493965

Abstract

Purpose. To determine the repeatability of recalculated corneal visualization Scheimpflug technology (CorVis ST) parameters and to study the variation of biomechanical properties and their association with demographic and ocular characteristics. Methods. A total of 783 healthy subjects were included in this study. Comprehensive ophthalmological examinations were conducted. The repeatability of the recalculated biomechanical parameters with 90 subjects was assessed by the coefficient of variation (CV) and intraclass correlation coefficient (ICC). Univariate and multivariate linear regression models were used to identify demographic and ocular factors. Results. The repeatability of the central corneal thickness (CCT), deformation amplitude (DA), and first/second applanation time (A1/A2-time) exhibited excellent repeatability (CV% ≤ 3.312% and ICC ≥ 0.929 for all measurements). The velocity in/out (V in/out), highest concavity- (HC-) radius, peak distance (PD), and DA showed a normal distribution. Univariate linear regression showed a statistically significant correlation between V in, V out, DA, PD, and HC-radius and IOP, CCT, and corneal volume, respectively. Multivariate analysis showed that IOP and CCT were negatively correlated with V in, DA, and PD, while there was a positive correlation between V out and HC-radius. Conclusion. The ICCs of the recalculated parameters, CCT, DA, A1-time, and A2-time, exhibited excellent repeatability. IOP, CCT, and corneal volume significantly influenced the biomechanical properties of the eye.

1. Introduction

The cornea acts as the outermost lens of the eye and aids in the focus and transmission of light. It provides mechanical stability to the eye and acts as a protective layer to the interior of the eye. Eye disorders, surgeries, and injuries may alter the shape, thickness, or biomechanical properties of the corneas, resulting in serious changes in visual performance of the eye [13]. Clinically, it is very extremely important to understand the biomechanical responses of cornea toward intraocular pressure (IOP), refractive surgery, and corneal pathology. To study the biomechanics of cornea, various methods have been devised [46]. One such method is the in vivo measurement of the biomechanical responses of the cornea using an ocular response analyzer (ORA; Reichert, Buffalo, NY, USA), one of the earliest commercially available devices [7]. Several studies have been published with diverse and new data regarding corneal hysteresis (CH) and corneal resistance factor (CRF) in healthy and pathologically challenged eyes [8, 9]. However, ORA, which reports nonstandard biomechanical terms that are different from the classical knowledge of corneal biomechanics, does not provide information associated with direct corneal deformation and cannot exhibit the corneal dynamic deformation process in real-time [10].

Recently, to overcome the lack of more direct measurement techniques of corneal deformation in vivo, a number of new approaches have been attempted, although these techniques are still in their early stages. A new device, corneal visualization Scheimpflug technology (CorVis ST; Oculus Optikgeräte GmbH, Wetzlar, Germany), designed to measure corneal deformation characteristics with a high-speed Scheimpflug camera, has been introduced. Measurement of corneal biomechanical properties using CorVis ST has received increasing attention recently [11, 12]. Biomechanical parameters may be associated with diseases such as keratoconus, stromal keratopathy, and glaucoma, which may be measured to evaluate the effect of corneal refractive surgery and corneal cross-linking [1317]. Leung et al. [18] reported that the influence of deformation amplitude (DA) on the measured error of Goldmann applanation tonometry is more in comparison to the central corneal thickness (CCT). Therefore, independent studies should focus on determining “normal” values in different populations, so that the new technology identifies wider acceptance and extensive use at clinical levels. The present study aimed to justify the repeatability of recalculated CorVis ST parameters and the variation of biomechanical properties provided by CorVis ST in a healthy Chinese population and their association with demographic and ocular characteristics.

2. Methods

2.1. Subject Recruitment

A total of 783 healthy Chinese subjects, aged 13–89 years, were recruited between July 2013 and August 2015 at the Chinese General Hospital of the People's Liberation Army (PLA, Beijing, China). This cross-sectional study was approved by the office of Research Ethics Committee of the Chinese PLA General Hospital in accordance with the tenets of the Declaration of Helsinki. Written informed consent was provided by all the subjects prior to entering the study.

2.2. Ocular Examinations

All the participants underwent a comprehensive ophthalmological examination and a standardized interview procedure. The ocular examinations included a detailed assessment of visual acuity, slit-lamp microscopy and fundus examination, corneal tomography (Pentacam; Oculus Optikgeräte GmbH, Wetzlar, Germany), corneal biomechanics, and IOP measurement (CorVis ST). All the assessments were performed at the same time during a single visit, to decrease the effect of diurnal variations.

The exclusion criteria for the subjects were as follows: any previous corneal or ocular surgery, any ocular pathology or systemic diseases known to affect the eye, or chronic use of topical medications.

2.3. CorVis ST Measurement

The CorVis ST instrument is a noncontact tonometer and imaging device that provides additional information regarding the reaction of cornea to a defined air pulse. It is mainly used to measure IOP and CCT. An ultrahigh-speed Scheimpflug camera (which can record 4,330 frames/sec) was used to capture the corneal deformation with a corneal horizontal range of 8.5 mm. The video clip containing 140 digital frames corresponded to a recording time of ~30 msec (Figure 1). In brief, the process was as follows: first, an image of the cornea was recorded by the ultrahigh-speed Scheimpflug camera, prior to the air puff in the natural convex shape. Subsequently, the cornea was made to move inward and flatten to the first applanation (A1), by precisely applying a metered air pulse to the cornea. The cornea continued to move inward until it reached a position of highest concavity (HC). As cornea is viscoelastic, it rebounded from this concavity to the second applanation (A2) and subsequently to its normal convex curvature. The IOP was determined according to Imbert-Fick's law based on A1-time [19]. The pachymetric values were detected when the cornea was in its natural shape and when there was no air pulse [19, 20]. Abbreviations shows the 10 biomechanical parameters measured by CorVis ST. The latest version of the software by Oculus (version 6.07r24) was used to recalculate all the CorVis ST measurements, which assisted in obtaining more precise data associated with the described parameters. For example, HC-radius presented an improvement from a three-point fit calculation obtained from previous software versions (version 1.0r30) to a new parabolic fit computation [21]. The quality-specification section on the output map was used to check the quality of the examination. A reading with an “OK” was considered acceptable; otherwise, the measurements were repeated.

Figure 1.

Figure 1

Deformation process observed using CorVis ST. (a) The recording starts with the cornea in a natural convex shape. (b) A precisely metered air pulse forces the cornea to move inward through an applanation (1st applanation). (c) The cornea then continues to move inward until it reaches the highest concavity. (d) The cornea rebounds to its normal convex shape. During this phase, the cornea again passes through an applanation (2nd applanation).

The CorVis ST is similar to the dynamic bidirectional applanator in that it uses an air puff to deform the cornea and records the process by a Scheimpflug camera. From the real-time series of images, deformation characteristics can be extracted along with the potential to quantify corneal elastic parameters. For example, the slower movement in V in and V out is associated with a stiffer cornea, since it takes greater force to deform a stiffer material and reach the same velocity. For the parameter of DA, it has been hypothesized that this may correspond to the stiffness of the cornea since a softer cornea with a lower elastic modulus would have greater deformation under the same load. Therefore, softer corneas would produce deep, narrow deformations, and stiffer corneas would produce wide, shallow deformations [22]. The A2-time may represent the time-dependent portion of the viscoelastic response and can be considered as an indication of the overall viscoelasticity [23].

Previous studies have reported that the IOP measurements using CorVis ST and Goldmann applanation tonometry showed satisfactory levels of agreement [24, 25]. Therefore, in the present study, the IOP reading provided by CorVis ST was used for data analysis.

2.4. Statistical Analysis

Two software programs, SPSS version 17.0 (SPSS, Inc., Chicago, IL, USA) and MedCalc 13.0 (MedCalc Software, Ostend, Belgium), were used to conduct the statistical analyses. Kolmogorov-Smirnov test was used to estimate the normality of distribution of the measured variables. Welch's modified Student's two-sample t-test and the Wilcoxon rank-sum test were used to determine differences in biomechanical parameters among gender groups. Linear regression analysis was performed to assess the effect of each variable on V in, V out, DA, PD, and HC-radius and multivariate linear regression models were further conducted with V in, V out, DA, PD, and HC-radius as the dependent variables and age, gender, IOP, CCT, mean keratometry (K m), and anterior chamber depth (ACD) as covariates. A P value < 0.05 was considered to indicate a statistically significant difference. Repeatability analysis was carried out on the recalculated biomechanical parameters (obtained using a new version of the software). The analysis was conducted by measuring three measurements accomplished by a single operator, on 90 eyes of the study samples. Precision, repeatability, coefficient of variation, and intraclass correlation coefficients (ICCs) were the factors calculated for the parameter repeatability analysis [26]. ICC was interpreted as follows: <0.75, poor to moderate repeatability; 0.75–0.90, good measurement repeatability; and >0.90, excellent repeatability for clinical measures [27].

3. Results

3.1. Repeatability

The repeatability of the recalculated biomechanical parameters was measured through 3 measurements (taken in a gap of 3 min on the same day) and was conducted on 90 eyes of the study samples. Table 1 shows the precision, repeatability, coefficient of variation, and ICC of the parameters. CCT, DA, A1-time, and A2-time showed excellent repeatability, with a CV% of 0.925%, 3.312%, 1.306%, and 0.657% and ICC of 0.983, 0.942, 0.929, and 0.929, respectively. PD, IOP, and V out showed good repeatability, with a CV% of 1.701%, 6.245%, and −9.834% and ICC of 0.896, 0.894, and 0.768, respectively. By contrast, ICCs of the other parameters were <0.75 with a moderate or poor level.

Table 1.

Repeatability of the recalculated corneal visualization Scheimpflug technology parameters for three measurementsa.

Parameters Precision Repeatability CV% ICC 95% CI for ICC
First applanation time (msec) 0.187 0.264 1.306 0.929 0.901~0.950
First applanation length (mm) 0.113 0.160 3.221 0.223 0.094~0.362
Velocity in (m/sec) 0.018 0.026 6.353 0.695 0.601~0.776
Second applanation time (msec) 0.284 0.401 0.657 0.929 0.901~0.951
Second applanation length (mm) 0.540 0.764 16.153 0.230 0.101~0.369
Velocity out (m/sec) 0.076 0.107 −9.834 0.768 0.690~0.832
Highest concavity time (msec) 0.684 0.966 2.093 0.302 0.171~0.438
Deformation amplitude (mm) 0.069 0.098 3.312 0.942 0.920~0.960
Peak distance (mm) 0.168 0.238 1.701 0.896 0.857~0.927
Highest concavity radius (mm) 1.136 1.606 8.320 0.605 0.495~0.704
Intraocular pressure (mmHg) 1.696 2.397 6.245 0.894 0.854~0.925
Central corneal thickness (μm) 9.642 13.627 0.925 0.983 0.976~0.988

a n = 90. CV: coefficient of variation; ICC: intraclass correlation coefficient; 95% CI: 95% confidence interval of the mean.

3.2. Characteristics of Healthy Subjects

A total of 783 healthy Chinese subjects (402 females, 51.34%; 381 males, 48.66%), with a mean age of 34.93 ± 17.65 years, were recruited in this study. The mean IOP and astigmatism of eyes were 14.14 ± 2.19 mmHg and 1.08 ± 0.7 diopters, respectively. The mean CCT was 541.14 ± 32.67 μm, mean keratometry was 43.63 ± 1.45 diopters, and mean corneal volume was 60.1 ± 3.69 mm3. The mean anterior chamber angle, depth, and volume were 37.03 ± 7.58 degrees, 3.08 ± 0.45 mm, and 175.14 ± 43.98 mm3, respectively. Kolmogorov-Smirnov test confirmed the normality of the biomechanical parameter distributions. The parameters, such as V in, V out, HC-radius, PD, and DA, showed a normal distribution (Figure 2). Table 2 showed that there was no significant difference between male and female subjects with respect to the 10 biomechanical parameters observed in the study (P > 0.05).

Figure 2.

Figure 2

Distribution of corneal biomechanical parameters in normal Chinese population. (a) Velocity in; (b) velocity out; (c) deformation amplitude; (d) peak distance; (e) highest concavity-radius.

Table 2.

Biomechanical parameters derived from corneal visualization Scheimpflug technology by gender and mean ± standard deviation (range).

Parameters All (n = 783) Female (n = 402) Male (n = 381) P value
First applanation time (ms) 7.33 ± 0.32 (6.63~8.39) 7.32 ± 0.29 (6.7~8.39) 7.35 ± 0.34 (6.63~8.38) 0.667
First applanation length (mm) 1.79 ± 0.07 (1.38~1.9) 1.79 ± 0.05 (1.43~1.9) 1.78 ± 0.08 (1.38~1.9) 0.940
Velocity in (m/s) 0.15 ± 0.02 (0.11~0.2) 0.15 ± 0.01 (0.11~0.19) 0.15 ± 0.02 (0.11~0.2) 0.370
Second applanation time (ms) 22.09 ± 0.45 (20.81~23.52) 22.07 ± 0.44 (20.81~23.52) 22.11 ± 0.45 (21.07~23.39) 0.218
Second applanation length (mm) 1.68 ± 0.32 (0.84~2.29) 1.69 ± 0.32 (0.89~2.15) 1.67 ± 0.32 (0.84~2.29) 0.360
Velocity out (m/s) −0.39 ± 0.07 (−0.68~−0.17) −0.39 ± 0.07 (−0.61~−0.17) −0.39 ± 0.07 (−0.68~−0.19) 0.935
Highest concavity time (ms) 16.72 ± 0.49 (14.78~18.71) 16.7 ± 0.5 (14.78~18.25) 16.74 ± 0.49 (15.25~18.71) 0.258
Deformation amplitude (mm) 1.07 ± 0.09 (0.84~1.29) 1.06 ± 0.08 (0.85~1.27) 1.07 ± 0.09 (0.84~1.29) 0.257
Peak distance (mm) 5.01 ± 0.24 (4.18~5.64) 5.01 ± 0.23 (4.4~5.59) 5.02 ± 0.25 (4.18~5.64) 0.657
Highest concavity radius (mm) 6.93 ± 0.79 (4.67~9.43) 6.89 ± 0.8 (4.93~9.43) 6.96 ± 0.78 (4.67~9.39) 0.177

Student's two-sample t-test.

Wilcoxon rank-sum test.

3.3. Determinants of Corneal Biomechanical Parameters

Table 3 showed the investigative results of the univariate linear regression analysis and multivariate models. V in, V out, DA, PD, and HC-radius showed no significant relationship with age and gender, whereas all of them varied in terms of IOP, CCT, and corneal volume.

Table 3.

Linear regression analyses on velocity in, velocity out, deformation amplitude, peak distance, and highest concavity radius.

Parameter  Velocity in (m/s)  Velocity out (m/s)  Deformation amplitude   (mm)  Peak distance (mm)  Highest concavity radius   (mm)
 Univariate  model  Multivariate  model  Univariate  model  Multivariate  model  Univariate  model  Multivariate  model  Univariate  model  Multivariate  model  Univariate  model  Multivariate  model
β value P value β value P value β value P value β value P value β value P value β value P value β value P value β value P value β value P value β value P value
Age −0.016 0.662 0.005 0.881 −0.426 0.67 −0.029 0.421 −0.021 0.561
Gender, female 0.032 0.37 −0.003 0.935 −0.041 0.257 −0.016 0.657 −0.048 0.177
Intraocular pressure −0.332 <0.001 −0.318 <0.001 0.536 <0.001 0.46 <0.001 −0.521 <0.001 −0.503 <0.001 −0.578 <0.001 −0.504 <0.001 0.378 <0.001 0.298 <0.001
Central corneal thickness −0.268 <0.001 −0.148 <0.001 0.353 <0.001 0.197 <0.001 −0.276 <0.001 −0.115 <0.001 −0.244 <0.001 −0.088 0.002 0.344 <0.001 0.248 <0.001
Mean keratometry 0.237 <0.001 0.267 <0.001 0.011 0.754 0.035 0.328 0.08 0.009 −0.0313 <0.001 −0.239 <0.001 −0.02 0.578
Corneal volume −0.107 0.003 0.253 <0.001 −0.25 <0.001 −0.289 <0.001 0.219 <0.001
Anterior chamber depth −0.018 0.623 −0.196 <0.001 −0.141 <0.001 −0.049 0.17 −0.095 0.002 0.245 <0.001 0.167 <0.001 −0.096 0.007
Anterior chamber volume −0.039 0.275 −0.225 <0.001 −0.04 0.265 0.297 <0.001 −0.13 <0.001

Adjusted for age, gender, intraocular pressure, central corneal thickness, mean keratometry, and anterior chamber depth; adjusted R 2 = 0.205 for velocity in, 0.341 for velocity out, 0.298 for deformation amplitude, 0.425 for peak distance, and 0.196 for highest concavity radius.

4. Discussion

The present study was one of the first studies that explored the correlation between gender variations and corneal biomechanical properties (provided by CorVis ST) and also their association with several demographic and ocular factors in a Chinese population. As knowledge regarding the corneal biomechanical properties and the influence of demographic and ocular characteristics on the corneal deformation response aid in predicting and diagnosing certain biomechanics-related ocular diseases, such as keratoconus during clinical practice, it is believed that study of these factors is important [11, 28, 29]. CorVis ST displays information on corneal deformation in real-time and allows a direct description of the mechanical behavior of the cornea [30]. The biomechanical parameters provided by CorVis ST may be associated with diseases such as keratoconus [11], glaucoma [15], and diabetes mellitus [31].

This study has shown that CCT, DA, A1-time, and A2-time exhibit excellent repeatability, followed by PD, IOP, and V out, which exhibited good repeatability levels. The biomechanical parameters, V in, V out, DA, PD, and HC-radius, showed a normal distribution in the healthy Chinese population. No significant differences were identified between male and female subjects in terms of the 10 biomechanical parameters evaluated. Univariate linear regression showed that V in, V out, DA, PD, and HC-radius showed statistically significant correlations with IOP, CCT, and corneal volume, respectively. It was found that IOP and CCT were still negatively correlated with V in and also with DA and PD, even following adjustment for age, gender, IOP, CCT, K m, and ACD, whereas they were positively associated with V out and HC-radius.

In a repeatability study of CorVis ST measurements, Hon and Lam [30] reported that CCT was the most repeatable corneal parameter measured by this device, followed by DA and A1-time, which was consistent with the results of Ali et al. [28]. In the present study, CCT, DA, A1-time, and A2-time showed excellent repeatability for the recalculated parameters. Previous studies had limitations of the software version, due to which the repeatability of PD and HC-radius had not been tested [18, 20]. In the present study, the ICCs were 0.896 and 0.605 for the recalculated PD and HC-radius, respectively. With the use of the new software version, the repeatability of the recalculated biomechanical parameters showed an improvement.

In the present study sample of Chinese subjects, the mean V in, V out, DA, PD, and HC-radius were 0.15 ± 0.02 m/s, −0.39 ± 0.07 m/s, 1.07 ± 0.09 mm, 5.01 ± 0.24 mm, and 6.93 ± 0.79 mm, respectively. Similar mean values of DA have been reported in studies that investigated diverse ethnicities (DA, 1.07 ± 0.10 mm) [28] and Brazilian subjects (DA, 1.05 ± 0.08 mm) [32].

In the present study, no significant correlations were identified between age and biomechanical parameters measured by CorVis ST. By contrast, certain previous studies have reported that structural changes in human corneal stroma are age-related, as they present an increase in stromal fibril diameter and interfibrillar cross-linking, contributing to increased stiffness [33, 34]. In addition, Elsheikh et al. [35] reported a similar correlation between age and corneal structure. Their study experimentally demonstrated that the cornea considerably stiffened with age, with a linearly related and increased Young's modulus of elasticity. As stiffness of the cornea increases with age, an older cornea would probably yield a lower DA and higher HC-radius. The findings of the present study appear to contradict these evidences, as this study showed no correlations between age and corneal biomechanical parameters. Nemeth et al. [20] reported that the 10 specific CorVis ST parameters showed no significant association with age. Similar findings have been reported by Hon and Lam [30], which suggested that the biomechanical parameters were not correlated with age and only HC-time showed a weak positive correlation (r = 0.18, P = 0.04) in a group of healthy subjects from a Brazilian population. Upon reviewing the aforementioned parameters, it was found that the CorVis ST instrument may not be sensitive enough to detect the change in biomechanical properties that is associated with age difference.

Hon and Lam [30] found that the biomechanical parameters were not associated with gender and there was no significant difference in the parameters between female and male subjects.

A positive correlation of K m was identified with V in, and a negative correlation was found with PD in univariate linear regression models; however, K m had no significant correlation with V out, DA, and HC-radius. These results were confirmed by earlier studies, as they also reported that DA was not correlated with corneal curvature [18, 29].

Until now, there have been no studies on the association of corneal volume, ACD, and ACV with biomechanical parameters measured using CorVis ST. However, it was found that corneal volume was negatively correlated with V in, DA, and PD but positively correlated with V out and HC-radius. Several studies have shown that, compared to normal controls, the values for corneal volume and CCT are significantly lower in keratoconus eyes [36, 37]. Mannion et al. [38] identified that when there is loss of corneal tissue, particularly in the central and paracentral area, a significant decrease in corneal volume in keratoconus is indicated. The decreases in corneal stiffness and reduction in corneal volume may be associated with each other. ACD and ACV were found to have a negative correlation with V out and HC-radius and a positive correlation with PD. However, further studies are warranted to understand the role of anterior chamber parameters and their association with the corneal deformation parameters in an improved manner.

Several previous studies have reported that CCT is negatively correlated with DA [18, 29, 39]. The present study also showed similar results. In a study conducted by Hon and Lam [30], it was reported that CCT was negatively correlated with V in and positively correlated with V out and HC-radius. This type of correlation may be due to the fact that the stiffness and elasticity are directly proportional to the corneal thickness, showing that as the corneal thickness increases, the other two mentioned parameters also increase. The effectiveness of corneal collagen fibers, the main contributors to corneal stiffness, may be less in those subjects having less CCT.

IOP is an important factor that affects the value of the biomechanical parameters. A clinical prospective observational case-control study reported that IOP has significant correlation with V in, V out, DA, and HC-radius, which shows that it has an important influence on corneal biomechanical behavior [39]. Furthermore, an ex vivo study using porcine cornea [40] and certain contact lens corneal models also verified the strong influence of IOP on corneal deformation [41]. In the present study, IOP presented a significant negative correlation with V in, DA, and PD and significant positive correlation with V out and HC-radius, which indicated that greater IOP causes a higher pressure threshold for the cornea to move, leading to lower V in. In addition, DA is lower at higher IOP; hence the cornea rebounds sooner, leading to greater V out in the recovery phase. Elsheikh et al. [42] measured the stiffness of 37 corneas from human donors using inflation tests and demonstrated that there was a positive linear association between Young's modulus and IOP. This result indicates that the cornea is less likely to deform when IOP is high. Therefore, IOP should be taken into account while conducting comparisons between the study populations as it has an important role in corneal biomechanical behavior by influencing the corneal deformation response.

The coefficients of determination (the R 2 value) in multiple linear regressions of V in, V out, DA, PD, and HC-radius were 0.205, 0.341, 0.298, 0.425, and 0.196, respectively. This implies that the variations in the IOP and anterior segment parameters explain 20–40% of the variance of the corneal biomechanical properties and the innate corneal properties explain the remaining 60–80% variance.

However, the present study has a few limitations. First, this was an observational cross-sectional study, which may limit causal inferences; and second, as this study focused only on Chinese population, it is not known whether these results can be extrapolated to other ethnicities.

In conclusion, to the best of our knowledge, this was the first study to profile variations in corneal biomechanical properties measured by CorVis ST in a large, unselected Chinese population. IOP, CCT, and corneal volume significantly influenced the biomechanical properties of the eye. These results may be relevant while investigating the role of altered corneal biomechanics in ocular diseases, such as corneal degeneration and glaucoma. As the CorVis ST is a relatively new technology, more studies should be conducted on the applicability and capabilities of this imaging technique for characterizing corneal biomechanics.

Acknowledgments

This research was supported by grants from the National Natural Science Foundation of China (81271052), the priming scientific research foundation for the junior researcher in Beijing Tongren Hospital, Capital Medical University (2015-YJJ-ZZL-008), and Beijing Key Laboratory of Ophthalmology and Visual Science (2016YKSJ02).

Abbreviations

A1-time:

First applanation time, time from the initiation of the air puff until the first applanation

A1-length:

First applanation length, length of the flattened cornea at the first applanation

Vin:

Velocity in, corneal velocity during the first applanation

HC-time:

Highest concavity time, time from the start until highest concavity of the cornea is reached

HC-radius:

Highest concavity-radius, radius of curvature at the time of highest concavity

PD:

Highest concavity peak distance, distance of the two surrounding “knees” at highest concavity

DA:

Highest concavity deformation amplitude, deformation amplitude from start to highest concavity at the corneal apex

A2-time:

Second applanation time, time from the initiation of the air puff until the second applanation

A2-length:

Second applanation length, length of the flattened cornea at the second applanation

Vout:

Velocity out, corneal velocity during the second applanation.

Disclosure

All authors concur with the submission.

Competing Interests

The authors declare that there is no conflict of interests regarding the publication of this paper.

References

  • 1.Congdon N. G., Broman A. T., Bandeen-Roche K., Grover D., Quigley H. A. Central corneal thickness and corneal hysteresis associated with glaucoma damage. American Journal of Ophthalmology. 2006;141(5):868–875. doi: 10.1016/j.ajo.2005.12.007. [DOI] [PubMed] [Google Scholar]
  • 2.Roy A. S., Shetty R., Kummelil M. K. Keratoconus: a biomechanical perspective on loss of corneal stiffness. Indian Journal of Ophthalmology. 2013;61(8):392–393. doi: 10.4103/0301-4738.116057. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Ryan D. S., Coe C. D., Howard R. S., Edwards J. D., Bower K. S. Corneal biomechanics following epi-LASIK. Journal of Refractive Surgery. 2011;27(6):458–464. doi: 10.3928/1081597x-20110112-01. [DOI] [PubMed] [Google Scholar]
  • 4.Beshtawi I. M., Akhtar R., Hillarby M. C., et al. Biomechanical properties of human corneas following low- and high-intensity collagen cross-linking determined with scanning acoustic microscopy. Investigative Ophthalmology and Visual Science. 2013;54(8):5273–5280. doi: 10.1167/iovs.13-12576. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Hollman K. W., Shtein R. M., Tripathy S., Kim K. Using an ultrasound elasticity microscope to map three-dimensional strain in a porcine cornea. Ultrasound in Medicine and Biology. 2013;39(8):1451–1459. doi: 10.1016/j.ultrasmedbio.2013.02.465. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Scarcelli G., Kling S., Quijano E., Pineda R., Marcos S., Yun S. H. Brillouin microscopy of collagen crosslinking: noncontact depth-dependent analysis of corneal elastic modulus. Investigative Ophthalmology and Visual Science. 2013;54(2):1418–1425. doi: 10.1167/iovs.12-11387. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Luce D. A. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. Journal of Cataract and Refractive Surgery. 2005;31(1):156–162. doi: 10.1016/j.jcrs.2004.10.044. [DOI] [PubMed] [Google Scholar]
  • 8.Prata T. S., Lima V. C., Guedes L. M., et al. Association between corneal biomechanical properties and optic nerve head morphology in newly diagnosed glaucoma patients. Clinical and Experimental Ophthalmology. 2012;40(7):682–688. doi: 10.1111/j.1442-9071.2012.02790.x. [DOI] [PubMed] [Google Scholar]
  • 9.Shah S., Laiquzzaman M., Bhojwani R., Mantry S., Cunliffe I. Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes. Investigative Ophthalmology and Visual Science. 2007;48(7):3026–3031. doi: 10.1167/iovs.04-0694. [DOI] [PubMed] [Google Scholar]
  • 10.McMonnies C. W. Assessing corneal hysteresis using the ocular response analyzer. Optometry and Vision Science. 2012;89(3):E343–E349. doi: 10.1097/OPX.0b013e3182417223. [DOI] [PubMed] [Google Scholar]
  • 11.Tian L., Huang Y.-F., Wang L.-Q., et al. Corneal biomechanical assessment using corneal visualization scheimpflug technology in keratoconic and normal eyes. Journal of Ophthalmology. 2014;2014:8. doi: 10.1155/2014/147516.147516 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Yu A., Zhao W., Savini G., et al. Evaluation of central corneal thickness using corneal dynamic scheimpflug analyzer corvis ST and comparison with pentacam rotating scheimpflug system and ultrasound pachymetry in normal eyes. Journal of Ophthalmology. 2015;2015:8. doi: 10.1155/2015/767012.767012 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Tian L., Ko M. W. L., Wang L.-K., et al. Assessment of ocular biomechanics using dynamic ultra high-speed scheimpflug imaging in keratoconic and normal eyes. Journal of Refractive Surgery. 2014;31(11):785–791. doi: 10.3928/1081597X-20140930-01. [DOI] [PubMed] [Google Scholar]
  • 14.Faria-Correia F., Ramos I., Valbon B., Luz A., Roberts C. J., Ambrósio R., Jr. Scheimpflug-based tomography and biomechanical assessment in pressure-induced stromal keratopathy. Journal of Refractive Surgery. 2013;29(5):356–358. doi: 10.3928/1081597x-20130129-03. [DOI] [PubMed] [Google Scholar]
  • 15.Tian L., Wang D., Wu Y., et al. Corneal biomechanical characteristics measured by the CorVis Scheimpflug technology in eyes with primary open-angle glaucoma and normal eyes. Acta Ophthalmologica. 2015 doi: 10.1111/aos.12672. [DOI] [PubMed] [Google Scholar]
  • 16.Pedersen I. B., Bak-Nielsen S., Vestergaard A. H., Ivarsen A., Hjortdal J. Corneal biomechanical properties after LASIK, ReLEx flex, and ReLEx smile by Scheimpflug-based dynamic tonometry. Graefe's Archive for Clinical and Experimental Ophthalmology. 2014;252(8):1329–1335. doi: 10.1007/s00417-014-2667-6. [DOI] [PubMed] [Google Scholar]
  • 17.Tomita M., Mita M., Huseynova T. Accelerated versus conventional corneal collagen crosslinking. Journal of Cataract and Refractive Surgery. 2014;40(6):1013–1020. doi: 10.1016/j.jcrs.2013.12.012. [DOI] [PubMed] [Google Scholar]
  • 18.Leung C. K.-S., Ye C., Weinreb R. N. An ultra-high-speed Scheimpflug camera for evaluation of corneal deformation response and its impact on IOP measurement. Investigative Ophthalmology and Visual Science. 2013;54(4):2885–2892. doi: 10.1167/iovs.12-11563. [DOI] [PubMed] [Google Scholar]
  • 19.Correia F. F., Ramos I., Roberts C. J., Steinmueller A., Krug M., Ambrósio R., Jr. Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed scheimpflug imaging. Arquivos Brasileiros de Oftalmologia. 2013;76(5):278–281. doi: 10.1590/s0004-27492013000500005. [DOI] [PubMed] [Google Scholar]
  • 20.Nemeth G., Hassan Z., Csutak A., Szalai E., Berta A., Modis L., Jr. Repeatability of ocular biomechanical data measurements with a scheimpflug-based noncontact device on normal corneas. Journal of Refractive Surgery. 2013;29(8):558–563. doi: 10.3928/1081597x-20130719-06. [DOI] [PubMed] [Google Scholar]
  • 21.Bak-Nielsen S., Pedersen I. B., Ivarsen A., Hjortdal J. Dynamic scheimpflug-based assessment of keratoconus and the effects of corneal cross-linking. Journal of Refractive Surgery. 2014;30(6):408–414. doi: 10.3928/1081597X-20140513-02. [DOI] [PubMed] [Google Scholar]
  • 22.Roberts C. J. Concepts and misconceptions in corneal biomechanics. Journal of Cataract and Refractive Surgery. 2014;40(6):862–869. doi: 10.1016/j.jcrs.2014.04.019. [DOI] [PubMed] [Google Scholar]
  • 23.Boptom H. R. V., Mrcophth D. V. P. Biomechanical properties of the keratoconic cornea: a review. Clinical and Experimental Optometry. 2015;98(1):31–38. doi: 10.1111/cxo.12211. [DOI] [PubMed] [Google Scholar]
  • 24.Hong J., Xu J., Wei A., et al. A new tonometer-the corvis ST tonometer: clinical comparison with noncontact and goldmann applanation tonometers. Investigative Ophthalmology & Visual Science. 2013;54(1):659–665. doi: 10.1167/iovs.12-10984. [DOI] [PubMed] [Google Scholar]
  • 25.Reznicek L., Muth D., Kampik A., Neubauer A. S., Hirneiss C. Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma. British Journal of Ophthalmology. 2013;97(11):1410–1414. doi: 10.1136/bjophthalmol-2013-303400. [DOI] [PubMed] [Google Scholar]
  • 26.Pryseley A., Mintiens K., Knapen K., Van der Stede Y., Molenberghs G. Estimating precision, repeatability, and reproducibility from Gaussian and non-Gaussian data: a mixed models approach. Journal of Applied Statistics. 2010;37(9-10):1729–1747. doi: 10.1080/02664760903150706. [DOI] [Google Scholar]
  • 27.Portney L. G., Watkins M. P. Foundations of Clinical Research. Stamford, Conn, USA: Appleton & Lange; 1993. [Google Scholar]
  • 28.Ali N. Q., Patel D. V., McGhee C. N. J. Biomechanical responses of healthy and keratoconic corneas measured using a noncontact scheimpflug-based tonometer. Investigative Ophthalmology & Visual Science. 2014;55(6):3651–3659. doi: 10.1167/iovs.13-13715. [DOI] [PubMed] [Google Scholar]
  • 29.Roberts C. J., Dupps W. J., Jr. Biomechanics of corneal ectasia and biomechanical treatments. Journal of Cataract and Refractive Surgery. 2014;40(6):991–998. doi: 10.1016/j.jcrs.2014.04.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Hon Y., Lam A. K. C. Corneal deformation measurement using Scheimpflug noncontact tonometry. Optometry & Vision Science. 2013;90(1):e1–e8. doi: 10.1097/opx.0b013e318279eb87. [DOI] [PubMed] [Google Scholar]
  • 31.Pérez-Rico C., Gutiérrez-Ortíz C., González-Mesa A., Zandueta A. M., Moreno-Salgueiro A., Germain F. Effect of diabetes mellitus on Corvis ST measurement process. Acta Ophthalmologica. 2015;93(3):e193–e198. doi: 10.1111/aos.12530. [DOI] [PubMed] [Google Scholar]
  • 32.Valbon B. F., Ambrośio R., Jr., Fontes B. M., Luz A., Roberts C. J., Alves M. R. Ocular biomechanical metrics by CorVis ST in healthy Brazilian patients. Journal of Refractive Surgery. 2014;30(7):468–473. doi: 10.3928/1081597x-20140521-01. [DOI] [PubMed] [Google Scholar]
  • 33.Daxer A., Misof K., Grabner B., Ettl A., Fratzl P. Collagen fibrils in the human corneal stroma: structure and aging. Investigative Ophthalmology & Visual Science. 1998;39(3):644–647. [PubMed] [Google Scholar]
  • 34.Malik N. S., Moss S. J., Ahmed N., Furth A. J., Wall R. S., Meek K. M. Ageing of the human corneal stroma: structural and biochemical changes. Biochimica et Biophysica Acta—Molecular Basis of Disease. 1992;1138(3):222–228. doi: 10.1016/0925-4439(92)90041-k. [DOI] [PubMed] [Google Scholar]
  • 35.Elsheikh A., Wang D., Brown M., Rama P., Campanelli M., Pye D. Assessment of corneal biomechanical properties and their variation with age. Current Eye Research. 2007;32(1):11–19. doi: 10.1080/02713680601077145. [DOI] [PubMed] [Google Scholar]
  • 36.Hosseini S. M. A., Abolbashari F., Niyazmand H., Sedaghat M. R. Efficacy of corneal tomography parameters and biomechanical characteristic in keratoconus detection. Contact Lens and Anterior Eye. 2014;37(1):26–30. doi: 10.1016/j.clae.2013.07.007. [DOI] [PubMed] [Google Scholar]
  • 37.Emre S., Doganay S., Yologlu S. Evaluation of anterior segment parameters in keratoconic eyes measured with the Pentacam system. Journal of Cataract and Refractive Surgery. 2007;33(10):1708–1712. doi: 10.1016/j.jcrs.2007.06.020. [DOI] [PubMed] [Google Scholar]
  • 38.Mannion L. S., Tromans C., O'Donnell C. Reduction in corneal volume with severity of keratoconus. Current Eye Research. 2011;36(6):522–527. doi: 10.3109/02713683.2011.553306. [DOI] [PubMed] [Google Scholar]
  • 39.Huseynova T., Waring G. O., IV, Roberts C., Krueger R. R., Tomita M. Corneal biomechanics as a function of intraocular pressure and pachymetry by dynamic infrared signal and scheimpflug imaging analysis in normal eyes. American Journal of Ophthalmology. 2014;157(4):885–893. doi: 10.1016/j.ajo.2013.12.024. [DOI] [PubMed] [Google Scholar]
  • 40.Kling S., Marcos S. Contributing factors to corneal deformation in air puff measurements. Investigative Ophthalmology and Visual Science. 2013;54(7):5078–5085. doi: 10.1167/iovs.13-12509. [DOI] [PubMed] [Google Scholar]
  • 41.Correia F. F., Ramos I., Roberts C. J., Steinmueller A., Krug M., Ambrósio R. Impact of chamber pressure and material properties on the deformation response of corneal models measured by dynamic ultra-high-speed scheimpflug imaging. Arquivos Brasileiros de Oftalmologia. 2013;76(5):278–281. doi: 10.1590/S0004-27492013000500005. [DOI] [PubMed] [Google Scholar]
  • 42.Elsheikh A., Alhasso D., Rama P. Biomechanical properties of human and porcine corneas. Experimental Eye Research. 2008;86(5):783–790. doi: 10.1016/j.exer.2008.02.006. [DOI] [PubMed] [Google Scholar]