CV of Peak Area |
Check signal stability |
%CV of peak area of EIC at 100% |
<15% |
Instable signal, likely an issue at the electrospray |
Protein Sequence Coverage |
Check method settings for identification capability |
Determined from search algorithm based on the confidence level chosen |
Near complete is ideal |
Acquisition setting, intrinsic sensitivity, LC separation |
Average Mass Error |
Report overall mass accuracy |
Average of absolute ppm error of all identified peptides with high confidence (or known peptides) |
From system specification <5 ppm or lower (High resolution) |
Temperature change, invalid calibration, incorrect method settings (e.g., lock mass, AGC) |
Mass Resolution |
Check instrument resolution |
Full width at half maximum mass peak resolution at 100% concentration |
Within instrument specification (instrument and operation mode dependent) |
Invalid instrument tuning/qualification, instrument malfunctioning |
Ion Source Settings |
Minimize non-native species generated in source |
Monitor the relative abundance of a fragile peptide in the standard |
Maintain sensitivity while minimize excess activation of fragile species |
Source tuning, source acceleration voltage too high |
S/N of EIC |
Defining LOD |
S/N of target peptide EIC |
> 3 |
Bad instrument tuning, bad EIC peak shape, bad spraying condition, intrinsic sensitivity limit of system |
MS2 Identification Score |
Check MS2 quality at different concentrations |
Target peptide score/cutoff threshold for intra and inter-scan |
≥1 |
Acquisition setting, intrinsic sensitivity, LC separation |
Mass Accuracy/Resolution Change |
Evaluate confidence of mass measurement at low concentrations |
Compare mass error/resolution at low and 100% concentration |
Ideally minimal change at low concentrations |
Instrument specific and can be normal to see mass error/resolution change at different signal intensities. Need to adjust data analysis procedure accordingly. |
Accuracy of Relative Quantitation |
Check accuracy and dynamic range of quantitation |
Percent accuracy of the experimental result relative to the expected result |
80% –120% |
Bad EIC extraction method, co-eluting species, instrument intrinsic limit |
CV of Relative Quantitation Result |
Check precision of quantitation |
%CV of peak area ratios of a target peptide pair |
<15% unless only intended for semi-quantitation |
Low sensitivity/bad peak shape at the concentration, incorrect data processing, interfering co-eluting species |
Differential Peak Sampling |
Check potential error in quantitation due to stochastic sampling across EIC peak |
Ratio of relative quantitation results between the proposed method and the MS1 only method |
0.8∼1.2 Use the MS1 only method for quantitation if exceeds the range |
Improper MS2 settings such as the selection threshold, MS1 and MS2 scan rate, etc. |