Skip to main content
. 2015 Jul 28;7(6):1104–1117. doi: 10.1080/19420862.2015.1074364

Table 2.

Summary of the proposed metrics for system suitability test, with proposed acceptance criteria and possible troubleshooting targets. The thresholds of 15% for CV and +/− 20% for relative quantitation are adapted from current FDA recommendations34

Metric Purpose Definition Recommended Threshold Possible Cause for Failure
CV of Peak Area Check signal stability %CV of peak area of EIC at 100% <15% Instable signal, likely an issue at the electrospray
Protein Sequence Coverage Check method settings for identification capability Determined from search algorithm based on the confidence level chosen Near complete is ideal Acquisition setting, intrinsic sensitivity, LC separation
Average Mass Error Report overall mass accuracy Average of absolute ppm error of all identified peptides with high confidence (or known peptides) From system specification <5 ppm or lower (High resolution) Temperature change, invalid calibration, incorrect method settings (e.g., lock mass, AGC)
Mass Resolution Check instrument resolution Full width at half maximum mass peak resolution at 100% concentration Within instrument specification (instrument and operation mode dependent) Invalid instrument tuning/qualification, instrument malfunctioning
Ion Source Settings Minimize non-native species generated in source Monitor the relative abundance of a fragile peptide in the standard Maintain sensitivity while minimize excess activation of fragile species Source tuning, source acceleration voltage too high
S/N of EIC Defining LOD S/N of target peptide EIC > 3 Bad instrument tuning, bad EIC peak shape, bad spraying condition, intrinsic sensitivity limit of system
MS2 Identification Score Check MS2 quality at different concentrations Target peptide score/cutoff threshold for intra and inter-scan ≥1 Acquisition setting, intrinsic sensitivity, LC separation
Mass Accuracy/Resolution Change Evaluate confidence of mass measurement at low concentrations Compare mass error/resolution at low and 100% concentration Ideally minimal change at low concentrations Instrument specific and can be normal to see mass error/resolution change at different signal intensities. Need to adjust data analysis procedure accordingly.
Accuracy of Relative Quantitation Check accuracy and dynamic range of quantitation Percent accuracy of the experimental result relative to the expected result 80% –120% Bad EIC extraction method, co-eluting species, instrument intrinsic limit
CV of Relative Quantitation Result Check precision of quantitation %CV of peak area ratios of a target peptide pair <15% unless only intended for semi-quantitation Low sensitivity/bad peak shape at the concentration, incorrect data processing, interfering co-eluting species
Differential Peak Sampling Check potential error in quantitation due to stochastic sampling across EIC peak Ratio of relative quantitation results between the proposed method and the MS1 only method 0.8∼1.2 Use the MS1 only method for quantitation if exceeds the range Improper MS2 settings such as the selection threshold, MS1 and MS2 scan rate, etc.