Table 1.

Technique factors and parameters set with a variable gantry speed. All images were acquired with 125 kVp x-ray tube setting

Gantry speed (°/s) Field of view (cm) Matrix size Pixel size (mm) Slice thickness (mm) Tube voltage (kVp) mAs Projections Δt (min)
1.0 46.5 512 × 512 0.7 2.0 125 5716 5400 6.0
2.0 46.5 512 × 512 0.7 2.0 125 2856 2700 3.0
3.0 46.5 512 × 512 0.7 2.0 125 1903 1800 2.0
4.0 46.5 512 × 512 0.7 2.0 125 1427 1350 1.5
5.0 46.5 512 × 512 0.7 2.0 125 1140 1080 1.2
6.0 46.5 512 × 512 0.7 2.0 125 949 900 1.0