Supplementary Table VI.

Intraday and interday precision and accuracy of HPLC methods used for determination of statins in DMEM culture medium samples. Both intraday and interday assay variability values were within acceptance criteria for all tested compounds at the 4 QC levels

Statins Quality control levels
Intraday (n = 6)
Interday (n = 6)
Concentration (μmol/L) Concentration (μg/mL) Precision (RSD%) Accuracy (RE%) Precision (RSD%) Accuracy (RE%)
Simvastatin lactone LLOQ (0.156) 0.065 2.06 7.36 17.28 −3.25
LQC (5) 2.09 10.13 6.83 11.82 3.8
MQC (40) 16.74 2.36 −3.34 9.35 11.15
HQC (120) 50.23 6.71 −2.71 10.13 0.41
Simvastatin hydroxy acid LLOQ (0.313) 0.142 3.99 −18.51 11.48 −15.85
LQC (5) 2.26 11.97 6.67 12.7 1.97
MQC (40) 18.14 11.47 −12.34 9.44 −14.98
HQC (120) 54.43 2.68 −5.3 3.54 −4.16
Pravastatin lactone LLOQ (0.234) 0.095 8.14 17.37 8.87 18.54
LQC (10) 4.06 7.87 6.59 4.07 12.59
MQC (100) 40.65 1.6 −6.71 1.5 −4.67
HQC (200) 81.30 3.63 −5.66 4.2 −2.74
Pravastatin hydroxy acid LLOQ (0.469) 0.209 7.28 −3.05 18.19 −8.12
LQC (10) 4.46 3.42 −4.19 1.64 −7.99
MQC (100) 44.65 1.54 2.01 1.19 1.29
HQC (200) 89.30 7.13 −5.56 9.75 −3.37

Abbreviations:DMEM, Dulbecco's modified eagle medium; HQC, high concentration quality control; LLOQ, lower limit of quantification; LQC, low concentration quality control; MQC, medium concentration quality control.

Relative standard deviation (RSD%) = (standard deviation [SD]/mean of obtained concentration) × 100.

Relative error (RE%) = ([measured concentration−nominal concentration]/nominal concentration) × 100.