Table 1.
Instrument conditions | |
---|---|
Temperatures | |
Injector temperature programme | 40° C, held for 1 min |
Ramp 1 | 300° C at a rate of 600° C min−1, held for 1.9 min |
Ramp 2 | 200° C at a rate of −10° C min−1, held for 1 min |
Final | Inlet cooled to 40° C |
Oven programme | 35° C for 1.5 min |
Ramp 1 | 15° C min−1 to 300° C (hold 2 min) |
MS interface | 280° C |
Injection conditions | |
Injector | PTV-LVI |
Syringe volume Injection volume |
100 μL 50 μL |
Injection speed | 1 μL s−1 |
Flow rates | |
Carrier gas | Helium |
Injector mode | PTV solvent vent |
Purge flow to split vent | 50 mL min−1 |
Vent | 100 mL min−1 |
Inlet pressure | 0.479 psi until 0.81 min |
Gas saver | 20 mL min−1 after 2 min |
Column flow rate | 2.5 mL min−1 |
Columns | |
Guard column | Non-polar Rxi® retention gap (1 m × 0.25 mm i.d.) |
Analytical column | HP5-MS (30 m × 0.25 mm × 0.25 μm film thickness) |
Mass spectrometer conditions | |
Mode | Electron ionisation (70 eV) |
Solvent delay | 8 min |
EMV mode | Relative |
Acquisition mode | Selected ion monitoring |
Dwell time | 20 (ms per mass) |