Table 1.

GC/MS conditions used for the analysis of organotin compounds.

Instrument conditions
Temperatures
Injector temperature programme 40° C, held for 1 min
Ramp 1 300° C at a rate of 600° C min−1, held for 1.9 min
Ramp 2 200° C at a rate of −10° C min−1, held for 1 min
Final Inlet cooled to 40° C
Oven programme 35° C for 1.5 min
Ramp 1 15° C min−1 to 300° C (hold 2 min)
MS interface 280° C



Injection conditions
Injector PTV-LVI
Syringe volume
Injection volume
100 μL
50 μL
Injection speed 1 μL s−1



Flow rates
Carrier gas Helium
Injector mode PTV solvent vent
Purge flow to split vent 50 mL min−1
Vent 100 mL min−1
Inlet pressure 0.479 psi until 0.81 min
Gas saver 20 mL min−1 after 2 min
Column flow rate 2.5 mL min−1



Columns
Guard column Non-polar Rxi® retention gap (1 m × 0.25 mm i.d.)
Analytical column HP5-MS (30 m × 0.25 mm × 0.25 μm film thickness)



Mass spectrometer conditions
Mode Electron ionisation (70 eV)
Solvent delay 8 min
EMV mode Relative
Acquisition mode Selected ion monitoring
Dwell time 20 (ms per mass)